Skip to content

Exactech 122-65-20Recalls & reported problems

Exactech, Inc. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885862021120), received between Jun 18, 2024 and Jun 18, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K993082); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 18, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A RIGHT HIP REPLACEMENT AND THEN APPROXIMATELY 8 YEARS AFTER THE INITIAL PROCEDURE THE PATIENT HAD A RIGHT HIP REVISION. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. REVISION OP REPORT - DIAGNOSIS: FAILED RIGHT HIP. THE PATIENT WAS REVISED TO COMPETITOR'S DEVICES. THE PATIENT HAD A VERY LARGE OSTEOLYTIC DEFECT ABOVE THE CUP THAT REQUIRED CREATING A BONY WINDOW, CURETTAGE, RONGEURING, AND BONE GRAFTING AND THEN REPLACING THE BONY WINDOW. THE SURGEON REMOVED SOME GREEN NONABSORBABLE SUTURE IN THE CAPSULE. ASPIRATED THROUGH THE CAPSULE AND OBTAINED ABOUT 3 ML OF CLEAR STRAW-COLORED FLUID. THERE WAS ABUNDANT EVIDENCE OF VILLONODULAR ADVERSE LOCAL TISSUE REACTION FOUND. THERE WAS SOME ADVERSE LOCAL TISSUE REACTION AROUND THE STEM WITH A SMALL AMOUNT OF OSTEOLYSIS BENEATH THE CALCAR AND AROUND THE STEM. THE STEM WAS WELL FIXED AND SOLID. THE LINER HAD WEAR SUPERIORLY. A STERILE BANDAGE WAS APPLIED. THERE IS NO SURGICAL DISPOSITION PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 19561242 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K993082. One submission often covers many models, so check the product and lot details in each record.

  • Z-1734-2022 · Aug 11, 2022 · Open, Classified
    GXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ACUMATCH 0 DEGREE LINER 28MM 46MM, REF 130-28-05; b. ACUMATCH 0 DEGREE LINER 28MM 48MM, REF 13
  • Z-1031-2017 · Nov 7, 2016 · Terminated
    AcuMatch A-Series Connexion GXL Enhanced Polyethylene Acetabular Liner, 15, +5 mm Lateralized, 36 mm I.D., Size J.

Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page