Novation 130-32-54Recalls & reported problems
Exactech, Inc. · Class 2 · Implant
All models of Novation (184)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10885862022189), received between Oct 10, 2024 and May 5, 2025 — classified as 2 injuries.
The most frequently reported specific device problem is “Naturally Worn” (2 reports).
2 FDA recall records name this device's identifier, 2 still open.
14 other recalls cover devices cleared under the same FDA submission (K070479); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 11, 2022Class IIOpen, Classified
Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.
Z-1729-2022 · Root cause: Process control · Recall details · FDA record ↗
- Jun 29, 2021Class IIOpen, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
Z-2117-2021 · Root cause: Unknown/Undetermined by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Naturally Worn2
Patient problems
- Osteolysis2
- Joint Dislocation1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 5, 2025Naturally Worn; Naturally Worn
IT WAS REPORTED THAT A MALE PATIENT, WHO HAD A HIP PROCEDURE, UNDERWENT A REVISION PROCEDURE APPROXIMATELY 10 YEARS 4 MONTHS POST THE INITIAL PROCEDURE. AS PART OF THE MANUFACTURER'S RECALL, THE PATIENT PRESENTED FOR A FOLLOW-UP IN 2023. X-RAYS AND CT SCANS REVEALED MODERATE DECENTRATION OF THE PROSTHETIC HEAD AND OSTEOLYSIS IN THE ACETABULUM, INDICATING INLAY WEAR. DUE TO THE PATIENT'S COMORBIDITIES AND THE MINIMAL SIGNS OF WEAR, A WATCHFUL WAITING APPROACH WITH REGULAR FOLLOW-UP VISITS WAS AGREED UPON. WHEN SYMPTOMS SUDDENLY WORSENED SIGNIFICANTLY AT THE END OF 2024, A FURTHER FOLLOW-UP VISIT WAS CONDUCTED, WHICH REVEALED COMPLETE DISLOCATION OF THE CUP. THE REVISION SURGERY WAS PERFORMED, INCLUDING REPLACEMENT OF THE ENTIRE CUP (IMPLANTATION OF A REVISION CUP WITH ILIAC PEG FROM A COMPETITOR. NO FURTHER INFORMATION.
Patient problems: Joint Dislocation; Osteolysis
MAUDE report 21965193 ↗InjuryOct 10, 2024Naturally Worn; Naturally Worn
AS REPORTED, THE PATIENT UNDERWENT HIP PROSTHESIS REPLACEMENT SURGERY APPROXIMATELY 4 YEARS 11 MONTHS AFTER INITIAL IMPLANT. IN 2019, THE PATIENT HAD A HIP PROSTHESIS CUP IMPLANTED. AS PART OF THE MANUFACTURER'S RECALL, THE PATIENT CAME IN FOR A CHECK-UP. THE X-RAY AND CT SCAN SHOWED A SLIGHT DECENTRATION OF THE PROSTHESIS HEAD AND OSTEOLYSIS IN THE ACETABULUM AS A SIGN OF INLAY WEAR. THIS WAS VERIFIED WHEN THE INLAY WAS CHANGED TO A VIT E-HARDENED, SPECIALLY APPROVED INLAY (NOVATION XLE, NEUTRAL LINER, GROUP 4, 36 MM I.D., REF 140-36-54, SN (B)(6)). AS PART OF THE REPLACEMENT OPERATION, IN ADDITION TO THE INLAY EXCHANGE, THE FIRM CUP INTEGRITY WAS DETERMINED, AS WELL AS THE CURETTAGE AND FILLING OF THE CYSTS IN THE CUP BED USING HYDROXYAPATITE + CALCIUM (CERAMENT BONE VOID FILLER) AND THE PROSTHESIS HEAD WAS CHANGED.
Patient problems: Osteolysis
MAUDE report 20418042 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K070479. One submission often covers many models, so check the product and lot details in each record.
- Z-2122-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-36-XX, Acumatch GXL 15 Degree Liner, 36mm, and Novation GXL Liner, Lipped A
- Z-2121-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-28-XX, Acumatch GXL 15 Degree Liner, 28mm; Novation GXL Liner, G0, 28mm ID;
- Z-2131-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-28-XX, Acumatch GXL 15 Degree, +5 Lat Liner, 28mm; Novation GXL 10 Deg Line
- Z-2120-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-22-XX, Novation GXL Liner, G00, 22mm ID; and Novation GXL Liner, Lipped Ant
- Z-2116-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-28-XX, Acumatch GXL 0 Degree Liner, 28mm; Novation GXL Liner, Neutral, 28m
- Z-2123-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-40-XX, Novation GXL Liner, Lipped Ant, 40mm; Novation GXL Liner, Lipped, 40
- Z-2128-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-36-XX, Novation GXL Liner, +5mm Lateralized, 36mm. Orthopedic hip implant
- Z-2118-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-36-XX, Acumatch GXL 0 Degree Liner, 36mm; and Novation GXL Liner, Neutral,
- Z-2132-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-32-XX, Novation GXL Liner, 10 Deg Face, 32mm. Exactech Connexion GXL a
- Z-2126-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-28-XX, Novation GXL Liner, +5mm Lateralized, 28mm; and Novation GXL Liner,
- Z-2127-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-32-XX, Novation GXL Liner, +5mm Lateralized, 32mm. Orthopedic hip implant
- Z-2133-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-32-XX, Acumatch GXL 15 Degree Liner, 32mm; Novation GXL Liner, Lipped Ant,
- Z-0787-2013 · Nov 5, 2012 · TerminatedExactech NOVATION CROWN CUP, NEUTRAL LINER, Use with Crown Cup Only, GXL UHMWPE, SN 0123456, REF 130-32-53. Exactech Hip Systems are indicated for use in surgery for hip replaceme
- Z-0788-2013 · Nov 5, 2012 · TerminatedExactech NOVATION CROWN CUP +5mm LATERALIZED LINER, Use with Crown Cup Only, GXL UHMWPE, SN 0123456, REF 136-32-53. Exactech Hip Systems are indicated for use in surgery for hip r
Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page