Skip to content

Acumatch A Series 132-28-27Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Acumatch A Series (94)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10885862023094), received between Aug 17, 2021 and Apr 11, 2022 — classified as 2 injuries.

The most frequently reported specific device problem is “Naturally Worn” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

10 other recalls cover devices cleared under the same FDA submission (K051556); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Aug 11, 2022Class IIOpen, Classified

    Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.

    Z-1728-2022 · Root cause: Process control · Recall details · FDA record ↗

  2. Jun 29, 2021Class IIOpen, Classified

    Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.

    Z-2121-2021 · Root cause: Unknown/Undetermined by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn2

Patient problems

  • Failure of Implant2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 11, 2022Naturally Worn; Naturally Worn

    AS REPORTED, APPROXIMATELY SEVENTEEN YEARS POST INITIAL RIGHT THA, THIS (B)(6) MALE PATIENT PRESENTED WITH APPARENT HIP LINER THAT WEAR LEADING TO INSTABILITY. PATIENT HAS A SUCCESSFUL REVISION SURGERY. PATIENT HIP LINER AND FEMORAL HEAD WAS REMOVED AND REPLACED. THERE WAS NO BREAKAGE OF DEVICE OR SURGICAL DELAY/PROLONGATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. PHOTOS AND X-RAYS RECEIVED. THE DEVICE WILL BE RETURN.

    Patient problems: Failure of Implant

    MAUDE report 14078463 ↗
  • InjuryAug 17, 2021Naturally Worn; Naturally Worn

    AS REPORTED, THE (B)(6) Y/O FEMALE PATIENT PRESENTED TO THE SURGEON WITH PAIN IN THE LEFT HIP DUE TO POLY WEAR. THE 15* 28MM ACUMATCH GXL LINER WAS REVISED TO A 15* 32MM XLE LINER. THERE WAS NO SURGICAL DELAY. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. THE DEVICE WILL BE RETURNED.

    Patient problems: Failure of Implant

    MAUDE report 12331616 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K051556. One submission often covers many models, so check the product and lot details in each record.

  • Z-2117-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-32-XX, Acumatch GXL 0 Degree Liner, 32mm; and Novation GXL Liner, Neutral,
  • Z-2131-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-28-XX, Acumatch GXL 15 Degree, +5 Lat Liner, 28mm; Novation GXL 10 Deg Line
  • Z-2112-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 104-28-XX, MCS +5GXL Liner 5/15 DEG and MCS GXL Liner 5/15 DEG. Orthopedic hip
  • Z-2118-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-36-XX, Acumatch GXL 0 Degree Liner, 36mm; and Novation GXL Liner, Neutral,
  • Z-2114-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 104-36-XX, MCS +5GXL Liner 5/15 DEG and MCS GXL Liner 5/15 DEG. Orthopedic hip
  • Z-2133-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-32-XX, Acumatch GXL 15 Degree Liner, 32mm; Novation GXL Liner, Lipped Ant,
  • Z-2116-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-28-XX, Acumatch GXL 0 Degree Liner, 28mm; Novation GXL Liner, Neutral, 28m
  • Z-2113-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 104-32-XX, MCS +5GXL Liner 5/15 DEG and MCS GXL Liner 5/15 DEG. Orthopedic hip
  • Z-2124-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 134-28-XX, Acumatch GXL, Ext Cov Liner, 28mm. Orthopedic hip implant component
  • Z-2122-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-36-XX, Acumatch GXL 15 Degree Liner, 36mm, and Novation GXL Liner, Lipped A

Other Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate devices

All Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page