Novation 132-32-54Recalls & reported problems
Exactech, Inc. · Class 2 · Implant
All models of Novation (184)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10885862023339), received between Aug 10, 2022 and Dec 18, 2023 — classified as 3 injuries.
The most frequently reported specific device problem is “Naturally Worn” (3 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 11, 2022Class IIOpen, Classified
Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.
Z-1729-2022 · Root cause: Process control · Recall details · FDA record ↗
- Jun 29, 2021Class IIOpen, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
Z-2133-2021 · Root cause: Unknown/Undetermined by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Naturally Worn3
- Adverse Event Without Identified Device or Use Problem1
- Device Dislodged or Dislocated1
Patient problems
- Failure of Implant2
- Pain2
- Ambulation or Postural Difficulties1
- Implant Pain1
- Joint Dislocation1
- Osteolysis1
- Swelling/ Edema1
- Synovitis1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryDec 18, 2023Naturally Worn; Adverse Event Without Identified Device or Use Problem; Naturally Worn; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED, A MALE PATIENT, WHO HAD AN INITIAL RIGHT HIP IMPLANTED ON (B)(6), 2014, UNDERWENT A REVISION PROCEDURE ON (B)(6), 2023, APPROXIMATELY 9 YEARS 10 MONTHS POST THE INITIAL PROCEDURE. THE PATIENT WAS BROUGHT BACK FOR RIGHT HIP PAIN. DURING THE PROCEDURE, THE FEMUR WAS DISLOCATED, AND THE FEMORAL HEAD REMOVED. THE STEM WAS CHECKED AND FOUND TO BE WELL FIXED. ATTENTION TURNED TO THE CUP. THE LINER WAS REMOVED WITH THE HELP OF A BONE SCREW. ALL THREE SCREWS WERE REMOVED, AND THE CUP WAS FOUND TO BE LOOSE AND REMOVED. FEMORAL HEAD ALLOGRAFT WAS IMPACTED INTO THE ACETABULUM WITH BONE GRAFT. A NEW COMPETITOR¿S DUAL MOBILITY CUP WAS THEN IMPLANTED ALONG WITH A NEW FEMORAL HEAD. THERE WERE NO SURGICAL DELAYS OR DEVICE BREAKAGES DURING THE PROCEDURE. AN X-RAY WAS PROVIDED. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO DEVICE RETURNS ARE AVAILABLE. THE PATIENT REQUESTED TO KEEP THEM. DEVICE IMAGES WERE PROVIDED. NO FURTHER INFORMATION.
Patient problems: Implant Pain; Pain
MAUDE report 18346649 ↗InjuryJul 25, 2023Naturally Worn; Naturally Worn
AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A RIGHT HIP REPLACEMENT ON (B)(6) 2010. APPROXIMATELY 2 YEARS AND 7 MONTHS AFTER THE INITIAL PROCEDURE THE PATIENT HAD A RIGHT HIP REVISION ON (B)(6) 2023. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. REVISION OP REPORT ON 22 JUN 2023. DIAGNOSIS: SEVERE OSTEOLYSIS FEMORAL AND ACETABULAR COMPONENTS THE PROXIMAL FEMUR WAS EVALUATED THERE WAS EXTENSIVE OSTEOLYSIS THE TROCHANTER WAS MODESTLY MOBILE. THE ENTIRE ACETABULAR COMPONENT SUPERIORLY THERE WAS A SIGNIFICANT LOSS OF BONE DUE TO OSTEOLYSIS AS THERE WAS INFERIORLY. THE PATIENT WAS AWAKENED TRANSPORTED TO THE RECOVERY ROOM THERE WERE NO COMPLICATIONS TO THE PROCEDURE.
Patient problems: Ambulation or Postural Difficulties; Failure of Implant; Osteolysis; Pain; Swelling/ Edema; Synovitis
MAUDE report 17389609 ↗InjuryAug 10, 2022Device Dislodged or Dislocated; Naturally Worn; Device Dislodged or Dislocated; Naturally Worn
AS REPORTED, APPROXIMATELY 6.5 YEARS POST INITIAL LEFT THA, SURGEON MADE DECISION TO REVISE HIP DUE TO PATIENT DISLOCATING POSTERIORLY MULTIPLE TIMES. PATIENT HAD A 60MM NOVATION CROWN CUP CLUSTER-HOLE SHELL GROUP, A NOVATION LIPPED LINER GROUP 4 32MM, AND A 32MM +0 COCR HEAD ON A SIZE 10 NOVATION ELEMENT HA COATED PRESS FIT STEM. SURGEON REVISED PATIENT VIA POLY SWAP BY GOING POSTERIOR APPROACH AND FOUND THE LIPPED LINER FACED ANTERIOR. SURGEON OBSERVED THAT THERE WAS SOME ADVANCED WEAR WITHIN THE POLY AND THE HEAD APPEARED TO BE MOVING AROUND WITHIN THE POLY ABNORMALLY AS IT WASN'T HELD WITHIN THE LINER AS NORMAL. SURGEON REMOVED LINER AND REMOVED HEAD. SURGEON SWITCHED TO A CERAMIC 36 OPTION HEAD WITH A +3.5 SLEEVE AND WENT WITH A NOVATION XLE 10 DEGREE FACE CHANGING LINER GROUP 4 36MM.
Patient problems: Failure of Implant; Joint Dislocation
MAUDE report 15198652 ↗
Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record