Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10885862023391), received between Oct 29, 2019 and Aug 21, 2024 — classified as 7 injuries.
The most frequently reported specific device problem is “Naturally Worn” (6 reports).
2 FDA recall records name this device's identifier, 2 still open.
10 other recalls cover devices cleared under the same FDA submission (K051556); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 11, 2022Class IIOpen, Classified
Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.
Z-1728-2022 · Root cause: Process control · Recall details · FDA record ↗
- Jun 29, 2021Class IIOpen, Classified
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
Z-2122-2021 · Root cause: Unknown/Undetermined by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Naturally Worn6
- Adverse Event Without Identified Device or Use Problem2
- Insufficient Device Problem Information1
Patient problems
- Failure of Implant6
- Insufficient Information2
- Osteolysis2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Implant Pain1
- No Code Available1
- Pain1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryAug 21, 2024Naturally Worn; Naturally Worn
IT WAS REPORTED THAT APPROXIMATELY UNKNOWN MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT HAS EXPERIENCED PROSTHESIS WEAR. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Insufficient Information; Osteolysis; Pain
MAUDE report 20012238 ↗InjuryJan 30, 2024Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information
AS REPORTED, THE PATIENT HAD AN INITIAL THA ON (B)(6) 2006. THE PATIENT IS PENDING REVISION; REASON UNKNOWN AND DATE UNKNOWN. NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant; Insufficient Information
MAUDE report 18611302 ↗InjuryJul 5, 2023Naturally Worn; Naturally Worn
AS REPORTED, THE PATIENT HAD AN INITIAL THA ON (B)(6)2007. THE PATIENT HAD A PREVIOUS REVISION ON (B)(6)2014. THE PATIENT WAS REVISED AGAIN ON (B)(6)2023 DUE TO PAIN, WEAR AND SEVERE OSTEOLYSIS ON CT SCAN, PER THE SURGEON. NO ISSUES WITH SURGERY. REVISED TO A COMPETITORS CAGE AND DUAL MOBILITY. A NEW 22MM METAL HEAD WAS ALSO USED. NO OTHER PATIENT INFORMATION/MEDICAL HISTORY REPORTED.
Patient problems: Failure of Implant; Implant Pain; Osteolysis
MAUDE report 17262963 ↗InjuryJun 19, 2022Naturally Worn; Naturally Worn
AS REPORTED, APPROXIMATELY TWELVE YEARS POST INITIAL LEFT THA, THE 66 Y/O MALE PATIENT HAD A REVISION DUE TO POLY WORN OVER TIME. PATIENT WAS REVISED WITH NEW ACUMATCH 36MM SIZE H 15 DEGREE LINER. HEAD ALSO REPLACED WITH BIOLOX DELTA 36MM +0. PT DEEMED STABLE. THERE WAS NO BREAKAGE OF DEVICE OR SURGICAL DELAY/PROLONGATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. PHOTOS RECEIVED. THE DEVICE IS NOT AVAILABLE DUE TO HOSPITAL POLICY.
Patient problems: Failure of Implant
MAUDE report 14740092 ↗InjuryNov 1, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED, APPROXIMATELY 11 YRS POSTOP THE INITIAL RTHA, THE (B)(6) Y/O MALE PATIENT PRESENTED COMPLAINING OF R HIP PAIN AND INSTABILITY. REVISED HEAD AND LINER. PATIENT WAS DEEMED STABLE. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION DUE TO HOSPITAL POLICY.
Patient problems: Failure of Implant
MAUDE report 12733825 ↗InjuryJul 1, 2020Naturally Worn; Naturally Worn
AS REPORTED, APPROXIMATELY 14 YEARS POSTOP THE INITIAL LEFT THA THIS MALE PATIENT PRESENTED COMPLAINING OF PAIN. THE LEFT HIP ACETABULAR COMPONENTS WERE FOUND TO BE MALPOSITIONED AND WORN. A REVISION WAS COMPLETED. THE FEMORAL STEM WAS LEFT IMPLANTED, THE FEMORAL HEAD WAS REMOVED AND REPLACED. ALL COMPONENTS ON THE ACETABULAR SIDE WERE REPLACED WITH COMPETITOR DEVICES DUE TO MALPOSITIONING AND WEAR. THIS PATIENT HAD THE FEMORAL COMPONENTS IN THE LEFT HIP REVISED 1 YEAR AGO. THE FEMORAL SIDE WAS REMOVED AND REPLACED WITH COMPETITOR DEVICES, BUT ACETABULAR SIDE REMAINED IMPLANTED, JUST SWAPPED OUT THE POLY LINER. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. DEVICES WILL NOT BE RETURNED DUE TO HOSPITAL POLICY.
Patient problems: Failure of Implant
MAUDE report 10220601 ↗InjuryOct 29, 2019Naturally Worn; Adverse Event Without Identified Device or Use Problem; Naturally Worn
AS REPORTED, APPROXIMATELY 9 ½ YEARS POST THE INITIAL THA, A 70 Y/O MALE PATIENT WAS REVISED. THE SURGEON OPENED THE JOINT IN STANDARD FASHION AND DISLOCATED THE HIP. HE REMOVED OLD HEAD AND LINER. HE THEN IMPLANTED A NEW HEAD AND LINER IN THE SAME SIZES. REPORTEDLY, THIS WAS MORE OF A "TIRE CHANGE." THE SURGEON THEN RELOCATED THE HIP AND CLOSED THE WOUND. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. DEVICES WERE DISPOSED OF BY THE HOSPITAL. BASED ON REVIEW OF ALL AVAILABLE INFORMATION, THERE IS NO EVIDENCE TO SUGGEST THAT THE REPORTED EVENT IS RELATED TO ANY DESIGN, MANUFACTURING, OR PATIENT RELATED ISSUES. THE CAUSE OF THE REVISION CANNOT BE CONCLUSIVELY DETERMINED SINCE THE DEVICES WERE NOT RETURNED. THE MOST LIKELY CAUSE IS EXPECTED WEAR.
Patient problems: Failure of Implant; No Code Available
MAUDE report 9250160 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K051556. One submission often covers many models, so check the product and lot details in each record.
- Z-2133-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-32-XX, Acumatch GXL 15 Degree Liner, 32mm; Novation GXL Liner, Lipped Ant,
- Z-2124-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 134-28-XX, Acumatch GXL, Ext Cov Liner, 28mm. Orthopedic hip implant component
- Z-2114-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 104-36-XX, MCS +5GXL Liner 5/15 DEG and MCS GXL Liner 5/15 DEG. Orthopedic hip
- Z-2131-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-28-XX, Acumatch GXL 15 Degree, +5 Lat Liner, 28mm; Novation GXL 10 Deg Line
- Z-2117-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-32-XX, Acumatch GXL 0 Degree Liner, 32mm; and Novation GXL Liner, Neutral,
- Z-2113-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 104-32-XX, MCS +5GXL Liner 5/15 DEG and MCS GXL Liner 5/15 DEG. Orthopedic hip
- Z-2118-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-36-XX, Acumatch GXL 0 Degree Liner, 36mm; and Novation GXL Liner, Neutral,
- Z-2116-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-28-XX, Acumatch GXL 0 Degree Liner, 28mm; Novation GXL Liner, Neutral, 28m
- Z-2112-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 104-28-XX, MCS +5GXL Liner 5/15 DEG and MCS GXL Liner 5/15 DEG. Orthopedic hip
- Z-2121-2021 · Jun 29, 2021 · Open, ClassifiedExactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-28-XX, Acumatch GXL 15 Degree Liner, 28mm; Novation GXL Liner, G0, 28mm ID;
Other Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate devices
- Novation 130-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls156 reports
- Novation 130-36-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls79 reports
- CORAIL 3L92511Depuy (Ireland) · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate158 reports
- Acumatch A Series 138-36-29Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate3 recalls1 report
- Novation 136-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls43 reports
- Novation 130-32-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls33 reports
- Novation 132-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls26 reports
- Novation 136-36-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls22 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page