Novation 136-36-52Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Novation (184)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Exactech Novation hip liner is the plastic cup component in total hip replacement. Most of the 38 reports describe the liner wearing out years after implantation, sometimes with bone loss around the implant, hip pain, or the hip becoming unstable and dislocating, leading to revision surgery. There are no deaths and 38 injuries. Two open recalls: one from August 2022 involving liners packaged in non-conforming bags that may speed up wear, and one from June 2021 on the risk of edge-loading and premature wear with certain implant configurations.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
38
Deaths reported
0
Injuries reported
38
Recalls naming it
2

Summary

The FDA's MAUDE database contains 38 adverse event reports that list this exact device (DI 10885862024329), received between Dec 20, 2019 and Feb 25, 2026 — classified as 38 injuries.

The most frequently reported specific device problem is “Naturally Worn” (33 reports).

2 FDA recall records name this device's identifier, 2 still open.

14 other recalls cover devices cleared under the same FDA submission (K070479); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 4 reportsJun 2024: 4 reportsJul 2024: 1 reportsAug 2024: 4 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 11, 2022Class IIOpen, Classified

    Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.

    Z-1729-2022 · Root cause: Process control · Recall details · FDA record ↗

  2. Jun 29, 2021Class IIOpen, Classified

    Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.

    Z-2128-2021 · Root cause: Unknown/Undetermined by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn33
  • Insufficient Device Problem Information5
  • Adverse Event Without Identified Device or Use Problem3
  • Fracture2
  • Loosening of Implant Not Related to Bone-Ingrowth2
  • Crack1
  • Mechanical Problem1

Patient problems

  • Failure of Implant19
  • Insufficient Information17
  • Osteolysis16
  • Implant Pain5
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4
  • Pain4
  • Synovitis4
  • Scar Tissue3

Recent adverse event reports

The 25 most recent of 38 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 25, 2026Naturally Worn

    AS REPORTED, APPROXIMATELY 10 YEARS AND 9 MONTHS POST THE INITIAL RIGHT TOTAL HIP ARTHROPLASTY, THE PATIENT REPORTED A GRINDING SENSATION IN THEIR HIP AND WAS REVISED DUE TO BEING UNSTABLE AND DISLOCATING POSTERIORLY. AN IMAGE PROVIDED SHOWS WEAR OF THE LINER. THERE WAS NO REPORTED BREAKAGE OF A DEVICE OR SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.

    Patient problems: Crepitus; Joint Dislocation; Joint Laxity

    MAUDE report 24442043 ↗
  • InjuryOct 15, 2025Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A LEFT TOTAL HIP ARTHROPLASTY. SUBSEQUENTLY, THE PATIENT EXPERIENCED WEAR OF THE ACETABULAR LINER AND HIP PAIN. AS A RESULT, THE PATIENT UNDERWENT LEFT HIP REVISION APPROXIMATELY SIX YEARS AND NINE MONTHS AFTER INITIAL IMPLANTATION. NO FURTHER PATIENT IMPACT REPORTED. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Pain

    MAUDE report 23286649 ↗
  • InjuryNov 27, 2024Naturally Worn; Naturally Worn

    AS REPORTED, APPROXIMATELY 3 MONTHS POST INITIAL LEFT TOTAL HIP ARTHROPLASTY (THA), THE PATIENT UNDERWENT REVISION AS THE POLY WAS WORN AND NEEDED REPLACING. DURING THE REVISION, IT WAS DETERMINED THAT THE CUP ALSO NEEDED TO BE REMOVED DUE TO BONE VOID. THE PATIENT WAS REVISED TO COMPETITOR ACETABULAR COMPONENTS AND THE SURGEON IMPLANTED THE EXACTECH BIOLOX DELTA OPTION 28MM HEAD AND +0 ADAPTER BECAUSE THE STEM STAYED INTACT. THE EVENT WAS NOT RELATED TO THE BREAKAGE OF A DEVICE AND DID NOT LEAD TO A SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.

    Patient problems: Osteolysis

    MAUDE report 20792110 ↗
  • InjuryOct 25, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED A 73 YO PATIENT, INITIAL RIGHT HIP IMPLANTED IN (B)(6) 2010, UNDERWENT A REVISION PROCEDURE IN (B)(6) 2024, APPROXIMATELY 13 YEARS 11 MONTHS POST THE INITIAL PROCEDURE. POLY WEAR OVER TIME REPORTED. THE PATIENT WAS REVISED TO A 146-36-52 NV EXHL LAT LNR G2 36MM. THERE WERE NO SURGICAL DELAYS OR DEVICE BREAKAGES DURING THE PROCEDURE. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO X-RAYS WERE PROVIDED. THE PRODUCT WILL NOT BE RETURNED FOR ANALYSIS DUE TO HOSPITAL POLICIES. NO DEVICE IMAGES WERE PROVIDED. NO FURTHER INFORMATION.

    Patient problems: Insufficient Information

    MAUDE report 20537396 ↗
  • InjurySep 4, 2024Fracture; Naturally Worn; Fracture; Naturally Worn

    IT WAS REPORTED THAT THIS 67 Y/O FEMALE PATIENT'S RIGHT HIP WAS REVISED APPROXIMATELY 14 YEARS 3 MONTHS POST OP. LINER WORE OUT. FEMORAL HEAD WAS EXCHANGED FOR AN EXACTECH BIOLOX OPTION 36MM HEAD W/ +3.5MM ADAPTER BUT A COMPETITORS LINER WAS IMPLANTED AND CEMENTED TO OUR CUP. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. UNABLE TO OBTAIN IMAGES OR X-RAYS. PRODUCT NOT RETURNING: HOSPITAL POLICY.

    Patient problems: Insufficient Information; Osteolysis

    MAUDE report 20140503 ↗
  • InjuryAug 27, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THIS MALE PATIENT HAD A HIP PROSTHESIS CUP IMPLANTED BY EXACTECH. HE PRESENTED TO SURGEON WITH SWELLING IN HIS KNEE. AS A SECONDARY FINDING, A CHECK OF THE IMPLANTED HIP PROSTHESIS WAS CARRIED OUT. THE PATIENT WAS FREE OF SYMPTOMS IN THIS REGARD. THE X-RAY AND CT CHECK SHOWED AN INLAY THAT WAS ALMOST COMPLETELY WORN OUT CRANIALLY WITH A CLEAR DECENTRATION OF THE PROSTHESIS HEAD IN THE CUP AND CLEAR OSTEOLYSIS IN THE ACETABULUM, SO THAT THE INDICATION FOR PROSTHESIS REVISION WAS MADE. NEW CARDIAC ARRHYTHMIAS OCCURRED DURING THE INDUCTION OF SURGERY/ANESTHESIA, SO THE OPERATION WAS NOT CARRIED OUT. AFTER CARDIOLOGICAL CLARIFICATION, THE NEW OPERATION IS SCHEDULED FOR (B)(6) 2024.

    Patient problems: Arrhythmia; Osteolysis; Swelling/ Edema

    MAUDE report 20077065 ↗
  • InjuryAug 26, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT APPROXIMATELY 116 MONTHS AFTER A LEFT HIP REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS OSTEOLYSIS, METALLOSIS, POLY WEAR, AND FEMORAL LOOSENING. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Osteolysis

    MAUDE report 20068865 ↗
  • InjuryAug 26, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 116 MONTHS AFTER A LEFT HIP REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS OSTEOLYSIS, METALLOSIS, POLY WEAR, AND FEMORAL LOOSENING. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Osteolysis

    MAUDE report 20067919 ↗
  • InjuryAug 22, 2024Naturally Worn; Naturally Worn

    AS REPORTED, THE PATIENT HAD AN INITIAL THA ON AN UNKNOWN DATE. THE PATIENT WAS REVISED DUE TO WEAR. THE CUP WAS REMOVED AND A COMPETITOR'S DEVICE WAS IMPLANTED. AN EXACTECH HEAD WAS USED. THERE WAS NO REPORTED BREAKAGE OF A DEVICE OR SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.

    Patient problems: Insufficient Information

    MAUDE report 20044534 ↗
  • InjuryJul 10, 2024Crack; Naturally Worn; Crack; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 46 MONTHS AFTER A LEFT TOTAL HIP REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, LEFT HIP PAIN, SNAPPING, SCARRING, AND IRRITATION; EVIDENCE OF POLYETHYLENE IRRITATION; CRACKING AND ECCENTRIC WEAR OF POLYETHYLENE. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN SATISFACTORY CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant; Implant Pain; Scar Tissue; Synovitis

    MAUDE report 19709416 ↗
  • InjuryJun 25, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT THE PATIENT WAS IMPLANTED WITH A NOVATION GXL DEVICE ON THE RIGHT HIP AND THEN APPROXIMATELY 7 YEARS LATER THE PATIENT WAS REVISED. THERE WAS NO OTHER PATIENT/MEDICAL INFORMATION PROVIDED. NO X-RAYS OR IMAGES WERE PROVIDED. THERE IS NO DEVICE RETURN. THERE IS NO OTHER INFORMATION AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 19608872 ↗
  • InjuryJun 7, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 75 MONTHS AFTER A LEFT TOTAL HIP REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PAIN, OSTEOLYSIS, AND HETEROTOPIC BONE. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Implant Pain; Osteolysis

    MAUDE report 19484457 ↗
  • InjuryJun 6, 2024Mechanical Problem; Naturally Worn; Mechanical Problem; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 79 MONTHS AFTER A RIGHT TOTAL HIP REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PAIN, LIMITED ROM, LOSS OF MOBILITY, ANXIETY, EMOTIONAL DISTRESS. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Anxiety; Emotional Changes; Implant Pain; Loss of Range of Motion; Osteolysis

    MAUDE report 19479696 ↗
  • InjuryJun 3, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 195 MONTHS AFTER A TOTAL HIP REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 19453552 ↗
  • InjuryMay 16, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 82 MONTHS AFTER A LEFT TOTAL HIP REPLACEMENT PROCEDURE, THE PATIENT HAS EXPERIENCED PROSTHESIS WEAR. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 19329404 ↗
  • InjuryMay 15, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 79 MONTHS AFTER A RIGHT TOTAL HIP REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS POLYETHYLENE WEAR. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 19326162 ↗
  • InjuryMay 15, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 89 MONTHS AFTER A RIGHT TOTAL HIP REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 19326143 ↗
  • InjuryMay 8, 2024Naturally Worn; Adverse Event Without Identified Device or Use Problem; Naturally Worn; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT A PATIENT HAD A LEFT TOTAL HIP ARTHROPLASTY ON (B)(6) 2016 AND A SURGICAL REVISION ON (B)(6) 2024; APPROXIMATELY 7 YEARS 9 MONTHS AFTER IMPLANT. OPERATIVE REPORT OF (B)(6) 2024- PROCEDURE: REVISION LEFT TOTAL HIP REPLACEMENT/ REVISED TO EXACTECH DEVICES. THE FEMORAL COMPONENT WAS FOUND TO BE STABLE AND WELL POSITIONED. THE LINER IS REMOVED. THE CUP IS FOUND TO BE STABLE. THE OSTEOLYTIC LESIONS ARE DEBRIDED, THEN PACKED WITH CANCELLOUS CHIPS. THE PATIENT TOLERATED THE PROCEDURE WELL, WITHOUT COMPLICATION AND TRANSFERRED TO THE RECOVERY ROOM IN A STABLE POST- OPERATIVE CONDITION. (PREOPERATIVE DIAGNOSIS IS NOT PROVIDED) NO DEVICE RETURN. THERE IS NO ADDITIONAL INFORMATION PROVIDED/AVAILABLE.

    Patient problems: Osteolysis; Pain

    MAUDE report 19267140 ↗
  • InjuryApr 18, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT A PATIENT HAD A LEFT TOTAL HIP ARTHROPLASTY ON (B)(6) 2016 AND THEN EXPERIENCED A REVISION SURGICAL PROCEDURE ON (B)(6) 2022 APPROXIMATELY 5 YEARS AND 5 MONTHS AFTER INITIAL IMPLANT. NO OTHER PATIENT INFORMATION / MEDICAL HISTORY REPORTED. NO IMAGES OF THE DEVICES ARE PROVIDED. THE DEVICE WILL NOT BE RETURNED. THERE IS NO OTHER INFORMATION AVAILABLE.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 19132221 ↗
  • InjuryApr 18, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT A PATIENT HAD A RIGHT TOTAL HIP ARTHROPLASTY ON (B)(6) 2013, WITH NO REPORTED REVISION. NO OTHER PATIENT INFORMATION / MEDICAL HISTORY REPORTED. NO RADIOGRAPHIC IMAGES OF THE DEVICE ARE PROVIDED. THE DEVICE WILL NOT BE RETURNED. THERE IS NO OTHER INFORMATION AVAILABLE.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 19132201 ↗
  • InjuryFeb 6, 2024Naturally Worn; Naturally Worn

    AS PART OF THE MANUFACTURER'S RECALL CAMPAIGN, THE PATIENT PRESENTED HIMSELF FOR A CHECK-UP OF THE DEVICE IMPLANTED IN 2013 LEFT HIP PROSTHESIS. THE X-RAY CONTROL SHOWED A CLEAR DECENTERING OF THE PROSTHETIC HEAD A VERY LARGE OSTEOLYSIS IN THE ACETABULUM AS A SIGN OF INLAY WEAR. THERE IS AN ADDITIONAL FINDING PRONOUNCED LUMBAR SPINE SYNDROME IN POLYMYALGIA RHEUMATICA. DURING THE REVISION SURGERY ON (B)(6) 2023, HE WAS ABLE TO RADIOLOGICAL FINDINGS ARE CONFIRMED. BECAUSE THE SOCKET WAS LOOSE AND OSTEOLYSIS OCCURRED DUE TO THE SIZE THEN THE COMPLETE SOCKET CHANGE TO A REVISION HIP CUP WITH TAB AND ILIAC PIN. AS PART OF THE REPLACEMENT OPERATION, THE CHANGING THE PROSTHETIC HEAD.

    Patient problems: Failure of Implant; Implant Pain; Myalgia; Osteolysis

    MAUDE report 18650814 ↗
  • InjuryNov 7, 2023Naturally Worn; Naturally Worn

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A RIGHT HIP REPLACEMENT ON (B)(6) 2013. APPROXIMATELY 10 YEARS AFTER THE INITIAL PROCEDURE THE PATIENT HAD A RIGHT HIP REVISION ON (B)(6) 2023. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. REVISION OP REPORT ON (B)(6) 2023. DIAGNOSIS: FAILED RIGHT THA (ADVANCED WEAR OF POLYETHYLENE LINER WITH SIGNIFICANT PERIACETABULAR OSTEOLYSIS). ENCOUNTERED EXTENSIVE REACTIVE GRANULOMATOUS SYNOVIAL TISSUE. FEMORAL HEAD WAS REMOVED WITH TAMP AND A MALLET. BONE IMPLANT INTERFACE OF THE FEMUR STEM WAS CLEARED¿FOUND TO BE WELL FIXED AND REASONABLY POSITIONED. STEM NOT REVISED. POLYETHYLENE LINER WAS REMOVED. AN EXTENSIVE SYNOVECTOMY WAS PERFORMED TO REMOVE GRANULOMATOUS TISSUES. REMOVED THE ACETABULUM WITH MINIMAL BONE LOSS. AS EXPECTED FROM PREOP XR AND CT THERE WAS EXTENSIVE OSTEOLYSIS OF THE ILIUM, AND LESS SO OF THE PUBIS AND ISCHIUM. THE PATIENT TOLERATED THE PROCEDURE WELL WITHOUT ANY COMPLICATIONS AND THE PROCEDURE WENT AS PLANNED.

    Patient problems: Failure of Implant; Osteolysis; Synovitis

    MAUDE report 18083689 ↗
  • InjurySep 20, 2023Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information

    IT WAS REPORTED THAT THIS FEMALE PATIENT'S LEFT HIP WAS REVISED. SURGEON REPLACED LINER WITH NEUTRAL XLE CUP AND A +3.5 36MM BIOLOX DELTA OPTIONS HEAD. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO OTHER PATIENT INFORMATION /MEDICAL HISTORY REPORTED. PRODUCT NOT RETURNING - HOSPITAL KEPT LINER.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 17780273 ↗
  • InjurySep 1, 2023Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A LEFT HIP REPLACEMENT ON (B)(6) 2016. APPROXIMATELY 6 YEARS AND 2 MONTHS AFTER THE INITIAL PROCEDURE THE PATIENT HAD A LEFT HIP REVISION ON (B)(6) 2023. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Insufficient Information; Joint effusion; Osteolysis; Pain; Scar Tissue

    MAUDE report 17672670 ↗
  • InjuryJul 6, 2023Naturally Worn; Naturally Worn

    AS REPORTED VIA LEGAL DOCUMENTATION, A PATIENT HAD LEFT HIP REPLACEMENT SURGERY ON (B)(6) 2016. THEY SUBSEQUENTLY UNDERWENT LEFT HIP REVISION SURGERY ON (B)(6) 2023, APPROXIMATELY 6 YEARS 10 MONTHS AFTER THEIR PRIMARY PROCEDURE. (B)(6) 2023 OP REPORT STATES THAT THERE WAS SOME MILD OSTEOLYSIS AROUND THE RIM OF THE ACETABULUM, BUT THE STEM WAS WELL FIXED AND RETAINED. THE TRUNNION OF THE FEMORAL HEAD WAS NOTED TO BE IN GOOD CONDITION. THE ACETABULAR LINER HAD SIGNIFICANT WEAR, SOME CRACKING ALONG THE SUPERIOR RIM OF THE ACETABULUM. THE CUP WAS WELL FIXED AND IN GOOD POSITION. THE CENTRAL HOLE AND AROUND THE RIM DID NOT INDICATE SIGNIFICANT OSTEOLYTIC DEFECTS; BONE GRAFTING WAS NOT NEEDED. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Osteolysis

    MAUDE report 17275471 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K070479. One submission often covers many models, so check the product and lot details in each record.

  • Z-2116-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-28-XX, Acumatch GXL 0 Degree Liner, 28mm; Novation GXL Liner, Neutral, 28m
  • Z-2123-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-40-XX, Novation GXL Liner, Lipped Ant, 40mm; Novation GXL Liner, Lipped, 40
  • Z-2120-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-22-XX, Novation GXL Liner, G00, 22mm ID; and Novation GXL Liner, Lipped Ant
  • Z-2133-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-32-XX, Acumatch GXL 15 Degree Liner, 32mm; Novation GXL Liner, Lipped Ant,
  • Z-2126-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-28-XX, Novation GXL Liner, +5mm Lateralized, 28mm; and Novation GXL Liner,
  • Z-2121-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-28-XX, Acumatch GXL 15 Degree Liner, 28mm; Novation GXL Liner, G0, 28mm ID;
  • Z-2127-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-32-XX, Novation GXL Liner, +5mm Lateralized, 32mm. Orthopedic hip implant
  • Z-2117-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-32-XX, Acumatch GXL 0 Degree Liner, 32mm; and Novation GXL Liner, Neutral,
  • Z-2132-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-32-XX, Novation GXL Liner, 10 Deg Face, 32mm. Exactech Connexion GXL a
  • Z-2131-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-28-XX, Acumatch GXL 15 Degree, +5 Lat Liner, 28mm; Novation GXL 10 Deg Line
  • Z-2118-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-36-XX, Acumatch GXL 0 Degree Liner, 36mm; and Novation GXL Liner, Neutral,
  • Z-2122-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-36-XX, Acumatch GXL 15 Degree Liner, 36mm, and Novation GXL Liner, Lipped A
  • Z-0788-2013 · Nov 5, 2012 · Terminated
    Exactech NOVATION CROWN CUP +5mm LATERALIZED LINER, Use with Crown Cup Only, GXL UHMWPE, SN 0123456, REF 136-32-53. Exactech Hip Systems are indicated for use in surgery for hip r
  • Z-0787-2013 · Nov 5, 2012 · Terminated
    Exactech NOVATION CROWN CUP, NEUTRAL LINER, Use with Crown Cup Only, GXL UHMWPE, SN 0123456, REF 130-32-53. Exactech Hip Systems are indicated for use in surgery for hip replaceme

Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented devices →