Exactech 142-28-00Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10885862026774), received between Oct 14, 2020 and Mar 7, 2025 — classified as 4 injuries.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 other recall cover devices cleared under the same FDA submission (K041906); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Detachment of Device or Device Component1
  • Naturally Worn1

Patient problems

  • Joint Dislocation3
  • Fall1
  • Implant Pain1
  • Metal Related Pathology1
  • Osteolysis1
  • Synovitis1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 7, 2025Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 41 MONTHS AFTER A RIGHT TOTAL HIP REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS ECCENTRIC POLYETHYLENE WEAR AND OSTEOLYSIS. INITIAL SURGERY AND REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. FINDINGS INDICATED EXTREME SYNOVITIS AND METALLOSIS, AND THE SURGEON WAS ABLE TO REMOVE THE CUP WITH HIS FINGERS. THE SURGEON PERFORMED A REVISION OF THE ACETABULAR CUP/LINER AND FEMORAL HEAD. PATIENT WAS TAKEN BACK TO THE POSTANESTHESIA RECOVERY ROOM IN A STABLE FASHION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Implant Pain; Metal Related Pathology; Osteolysis; Synovitis

    MAUDE report 21549384 ↗
  • InjuryFeb 12, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THIS PATIENT'S HIP WAS REVISED APPROXIMATELY 3.5 YEARS POST INITIAL OPERATION. THE PATIENT FELL DOWN AND HAD DISLOCATION OF BIPOLAR CUP. NEXT DAY, THIS PATIENT WENT TO SEE A DOCTOR AND SURGEON FOUND OUT DISASSOCIATION OF FEMORAL HEAD FROM BIPOLAR CUP VIA IMAGING SYSTEM WHEN HE PERFORMED A CLOSED/MANUAL REDUCTION BY USING IMAGING SYSTEM. IN THIS SITUATION, HE DECIDED TO PERFORM REVISION SURGERY AND FEMORAL HEAD AND BIPOLAR CUP WERE REPLACED TO NEW ONES. THIS FEMORAL HEAD CAME OFF THIS BIPOLAR CUP WITHOUT DISASSOCIATION OF LOCKING RING.

    Patient problems: Fall; Joint Dislocation

    MAUDE report 21364024 ↗
  • InjuryFeb 6, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED, APPROXIMATELY 44 DAYS POST INITIAL LEFT SIDE THA, THE FEMALE PATIENT WAS BROUGHT BACK FOR A DISLOCATED LEFT BIPOLAR HIP. THE BIPOLAR SHELL AND 28 HEAD WERE REMOVED. THE STEM CHECKED AND FOUND TO BE WELL FIXED. A COMPETITOR'S DUAL MOBILITY CUP AND AN EXACTECH 28 +3.5 HEAD WAS THEN IMPLANTED. THERE WAS NO BREAKAGE OF DEVICE OR SURGICAL DELAY/PROLONGATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. PHOTO RECEIVED. SALES REP WAS UNABLE TO OBTAIN X-RAYS. THE DEVICE IS NOT AVAILABLE FOR EVALUATION AS IT WAS DISCARDED BY HOSPITAL. NO OTHER PATIENT INFORMATION/MEDICAL HISTORY REPORTED.

    Patient problems: Joint Dislocation

    MAUDE report 16308514 ↗
  • InjuryOct 14, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON (B)(6) 2000,14 DAYS POST-OP, A LEFT HIP DISLOCATION OF THE BIPOLAR CUP FROM THE INNER HEAD FIXED TO FEMORAL STEM WAS OBSERVED WHEN THE POST-OP X-RAYS OF THE PATIENT WAS TAKEN. REVISION SURGERY OF THE BIPOLAR CUP AND THE INNER HEAD WAS IMMEDIATELY PERFORMED. THE DISLOCATED BIPOLAR CUP OF DIA. 54MM AND THE DIA. 28MM+0MM INNER METAL HEAD WERE WELL FIXED TO FEMORAL STEM AND WERE REMOVED. REPLACED TO BIPOLAR CUP OF DIA. 53MM AND DIA. 28MM + 3.5MM INNER HEAD. THE LOCKING RING OF THE RETRIEVED BIPOLAR CUP WAS FIRMLY ASSEMBLED TO THE METALLIC SHELL AND ANY DAMAGES OF THE POLYETHYLENE LOCKING RING WAS NOT OBSERVED VISUALLY AT ALL. THERE WAS A 30-45 MIN DELAY IN SURGERY, (NOTE: THIS SURGICAL DELAY MEANS THE SURGICAL TIME OF THIS REVISION SURGERY COMPARED AGAINST THE SURGICAL TIME OF THE PRIMARY SURGERY. THE ACTUAL PRIMARY AND REVISION SURGERY HAVE BEEN SMOOTHLY FINISHED.) THIS DID NOT CAUSE ANY ADVERSE EVENT TO THE PATIENT. PATIENT CONTINUES IN FOLLOW-UP. DEVICES WILL RETURN.

    Patient problems: Joint Dislocation

    MAUDE report 10677366 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K041906. One submission often covers many models, so check the product and lot details in each record.

  • Z-1727-2022 · Aug 11, 2022 · Open, Classified
    GXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ACUMATCH +5MM LAT LINER 15 DEG 36MM ID, SIZE H, REF 138-36-08 b. ACUMATCH +5MM LAT LINER 15 DE

Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page