Novation 167-02-00Recalls & reported problems
Exactech, Inc. · NOVATION A+, BROACH HANDLE, RIGHT · Class 2
All models of Novation (184)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10885862036025), received between Oct 30, 2020 and Nov 8, 2022 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K080980); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break2
- Component Misassembled1
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 8, 2022Component Misassembled; Component Misassembled
IT WAS REPORTED THAT THE SPRINGS ON THE LOCKING MECHANISM OF THE BROACH HANDLE HAVE BROKEN. THIS HAS HAPPENED OVER TIME BY HEAVY STRIKES WITH THE MALLET AND CARELESS BEHAVIOR IN CENTRAL STERILE. THE SPRINGS AND LOCKING MECHANISMS ARE DAMAGED ON THE HANDLES. NO OTHER INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15752898 ↗MalfunctionFeb 9, 2022Break; Break
AS REPORTED, DURING IMPACTION, THE SPRING ON THE BROACH HANDLE WAS BROKEN BY SURGEON. THIS SITUATION CAME TO NO SURPRISE BECAUSE THIS HAS HAPPENED TO SURGEON ON NUMEROUS OCCASIONS USING A PLUS SIZED MALLET WHILE BROACH FEMUR. THIS INCIDENT CAUSED NO DELAY, THE SURGICAL SITE WAS CONFIRMED CLEAN BY SURGEON AND THE PROCEDURE CARRIED ON AS PLANNED. PATIENT IS STABLE CONDITION. THE DEVICE WILL BE RETURN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13500541 ↗InjuryDec 18, 2020Break; Break
SURGEON BROKE THE SPRING IN THE HANDLE DURING IMPACTION. THE PATIENT WAS NOT AFFECTED. THERE WAS NO HARM TO THE PATIENT. THE CASE WENT ON AS PLANNED WITH NO ISSUE. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. PATIENT IN STABLE CONDITION AND NOT AFFECTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11036121 ↗MalfunctionOct 30, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
BROACH HANDLES SPRING WAS BROKEN DURING IMPACTION. PATIENT WAS NOT HARMED. HAD TWO HANDLES IN THE ROOM. NO SURGICAL DELAY. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10762947 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K080980. One submission often covers many models, so check the product and lot details in each record.
- Z-0201-2016 · Sep 23, 2015 · TerminatedNovation Element Press-Fit, Cementless, HA Coated, Collared, Standard Offset, Size 9, Femoral Hip Implant Stems, Intended for press-fit fixation
Other Prosthesis, hip, semi-constrained, uncemented, metal / polymer, non-porous, calcium phosphate devices
- Novation 130-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls156 reports
- Novation 130-36-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls79 reports
- CORAIL 3L92511Depuy (Ireland) · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate158 reports
- Acumatch A Series 138-36-29Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate3 recalls1 report
- Novation 136-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls43 reports
- Novation 130-32-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls33 reports
- Novation 132-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls26 reports
- Novation 136-36-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls22 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page