Skip to content

Novation 180-01-48Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Novation (184)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10885862036995), received between Nov 10, 2021 and Sep 2, 2026 — classified as 6 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (3 reports).

16 other recalls cover devices cleared under the same FDA submission (K070479); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth3
  • Naturally Worn2
  • Loss of Osseointegration1
  • Malposition of Device1

Patient problems

  • Aseptic loosening2
  • Failure of Implant2
  • Osteolysis2
  • Pain2
  • Ambulation or Postural Difficulties1
  • Bone Fracture(s)1
  • Metal Related Pathology1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjurySep 2, 2026Loss of Osseointegration

    Patient problems: Ambulation or Postural Difficulties; Aseptic loosening; Pain

    MAUDE report 50572095 ↗
  • InjuryAug 21, 2025Malposition of Device; Loosening of Implant Not Related to Bone-Ingrowth; Malposition of Device; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED A PATIENT UNDERWENT A LEFT HIP TOTAL ARTHROPLASTY. SUBSEQUENTLY, THIRTEEN AND HALF (13.5) YEARS LATER, THE PATIENT'S X-RAYS INDICATED A MALPOSITIONED ACETABULAR CUP AND POLYETHYLENE WEAR. THE PATIENT UNDERWENT A LEFT HIP REVISION FOUR (4) DAYS LATER. THE ACETABULAR CUP, LINER AND FEMORAL HEAD WERE REVISED. THE CUP AND LINER WERE NOTED TO BE A COMPETITOR'S IMPLANTS. INTRAOPERATIVELY, IT WAS NOTED THE PATIENT NEEDED BONE GRAFT DUE TO POLY WEAR, OSTEOLYSIS AND CUP REMOVAL. THERE WAS AN UNEXPECTED SURGICAL DELAY OF 30-45 MINUTES REPORTED DUE TO THE ACETABULAR CUP BEING LOOSE AND FALLING OUT. AS A RESULT, A FULL ACETABULAR COMPONENT REVISION WAS NEEDED INSTEAD OF THE PLANNED REVISION OF JUST THE POLYETHYLENE COMPONENTS AND THE FEMORAL HEAD EXCHANGE. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Aseptic loosening; Osteolysis

    MAUDE report 22859786 ↗
  • InjuryFeb 26, 2025Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 65 MONTHS AFTER A LEFT TOTAL HIP REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, METAL RELATED PATHOLOGY AND PAIN. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO IMAGES WERE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Metal Related Pathology; Pain

    MAUDE report 21471702 ↗
  • InjurySep 23, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE PATIENT HAD HIP PROSTHESES FROM EXACTECH IMPLANTED ON THE RIGHT ON (B)(6) 2018. DURING OUTPATIENT CHECK-UPS, WITH MILD LOAD-RELATED COMPLAINTS, DECENTRATION OF THE PROSTHESIS HEAD WAS NOTICED FOR THE FIRST TIME IN (B)(6) 2021 AND AN ACETABULAR FOREIGN BODY GRANULOMA WAS SUSPECTED. DUE TO A TRAUMATIC ACETABULAR FRACTURE WITH ACETABULAR LOOSENING, A REVISION WITH ACETABULAR REPLACEMENT TO A REVISION ACETABULAR CUP WAS CARRIED OUT IN (B)(6) 2023. IN RETROSPECT, IT IS LIKELY THAT THE FRACTURE WAS AT LEAST EXACERBATED BY THE OSTEOLYSIS CAUSED BY PREMATURE PE ABRASION. RELATED TO 1038671-2021-00269 (LEFT HIP REVISION).

    Patient problems: Bone Fracture(s); Osteolysis

    MAUDE report 20277809 ↗
  • InjuryFeb 9, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED BY THE LEGAL BRIEF, APPROXIMATELY FIVE YEARS POST INITIAL THA, THIS FEMALE PATIENT EXPERIENCE SEVERE PAIN IN HER RIGHT HIP, AND WAS SUBSEQUENTLY DIAGNOSED WITH LOOSENING OF THE HIP SYSTEM AND POLYETHYLENE WEAR OF THE ACETABULAR COMPONENT. PATIENT HAD SUSTAINED SEVERE AND PERMANENT INJURIES; SHE WAS REQUIRED TO UNDERGO MULTIPLE PERMANENT AND PAINFUL REVISION SURGERIES, INCLUDING THE REMOVAL AND RE-IMPLANTATION OF A REPLACEMENT PROSTHETIC HIP. SHE HAS REQUIRED MANY MONTHS OF HOSPITALIZATION, PHYSICAL THERAPY, AND OTHER CARE. SHE HAS AND WILL CONTINUE TO BE BEDRIDDEN. SHE HAS AND WILL CONTINUE TO EXPERIENCE SEVER AND PERMANENT PAIN; AND SHE HAS AND WILL CONTINUE TO SUFFER SEVERE AND PERMANENT COMPLICATIONS AND OTHER INJURIES TO THE RIGHT HIP.

    Patient problems: Failure of Implant

    MAUDE report 13500418 ↗
  • InjuryNov 10, 2021Naturally Worn; Naturally Worn

    AS REPORTED BY THE LEGAL BRIEF, PLAINTIFF UNDERWENT REVISION SURGERY TO REMOVE THE DEFECTIVE DEVICE, EXCISE A LARGE OSTEOLYTIC CYST, AND REBUILD BONE SUPPORT A NEW ACETABULAR CUP. THE DEVICE IMPLANTED INTO PLAINTIFF WAS DEFECTIVE AND UNREASONABLY DANGEROUS IN THAT THE ACETABULAR LINER CATASTROPHICALLY FRACTURED, WHICH ALLOWED THE FEMORAL HEAD TO ARTICULATE DIRECTLY AGAINST THE INITIAL CUP AND CAUSE SEVERE PAIN AND INJURY TO THE PLAINTIFF AND NECESSITATED A REVISION SURGERY ONLY EIGHT YEARS AFTER IMPLANT. PLAINTIFFS LEARNED FOR THE FIRST TIME THAT THE DEVICE WAS DEFECTIVE AFTER ITS REMOVAL ON OR ABOUT (B)(6) 2020. AS A DIRECT RESULT OF THE DEFECTIVE NATURE OF THE DEVICE, THE PLAINTIFF HAS SUFFERED AND CONTINUES TO SUFFER PERMANENT AND DEBILITATING INJURES AND DRAINAGES, INCLUDING BUT NOT LIMITED TO, SIGNIFICANT PAIN AND DISCOMFORT; GAIT IMPAIRMENT; POOR BALANCE; DIFFICULTY WALKING; COMPONENT PART LOOSENING, IMPINGEMENT AND/OR DETACHMENT; METALLOSIS; SOFT TISSUE DAMAGE; ADVERSE LOCAL TISSUE REACTION; AND OTHER INJURIES PRESENTLY UNDIAGNOSED, WHICH ALL REQUIRE ONGOING MEDICAL CARE. AS A FURTHER DIRECT RESULT, THE PLAINTIFF HAS SUSTAINED AND WILL SUSTAIN FUTURE DAMAGES, INCLUDING BUT N

    Patient problems: Failure of Implant

    MAUDE report 12787773 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K070479. One submission often covers many models, so check the product and lot details in each record.

  • Z-1729-2022 · Aug 11, 2022 · Open, Classified
    GXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. NOVATION GXL LINER, NEUTRAL, 28MM ID, GROUP 1 CUPS, REF 130-28-51 b. NOVATION GXL LINER, NEUTRA
  • Z-2117-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-32-XX, Acumatch GXL 0 Degree Liner, 32mm; and Novation GXL Liner, Neutral,
  • Z-2118-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-36-XX, Acumatch GXL 0 Degree Liner, 36mm; and Novation GXL Liner, Neutral,
  • Z-2122-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-36-XX, Acumatch GXL 15 Degree Liner, 36mm, and Novation GXL Liner, Lipped A
  • Z-2128-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-36-XX, Novation GXL Liner, +5mm Lateralized, 36mm. Orthopedic hip implant
  • Z-2116-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-28-XX, Acumatch GXL 0 Degree Liner, 28mm; Novation GXL Liner, Neutral, 28m
  • Z-2126-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-28-XX, Novation GXL Liner, +5mm Lateralized, 28mm; and Novation GXL Liner,
  • Z-2127-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-32-XX, Novation GXL Liner, +5mm Lateralized, 32mm. Orthopedic hip implant
  • Z-2121-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-28-XX, Acumatch GXL 15 Degree Liner, 28mm; Novation GXL Liner, G0, 28mm ID;
  • Z-2133-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-32-XX, Acumatch GXL 15 Degree Liner, 32mm; Novation GXL Liner, Lipped Ant,
  • Z-2132-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-32-XX, Novation GXL Liner, 10 Deg Face, 32mm. Exactech Connexion GXL a
  • Z-2131-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-28-XX, Acumatch GXL 15 Degree, +5 Lat Liner, 28mm; Novation GXL 10 Deg Line
  • Z-2123-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-40-XX, Novation GXL Liner, Lipped Ant, 40mm; Novation GXL Liner, Lipped, 40
  • Z-2120-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-22-XX, Novation GXL Liner, G00, 22mm ID; and Novation GXL Liner, Lipped Ant
  • Z-0787-2013 · Nov 5, 2012 · Terminated
    Exactech NOVATION CROWN CUP, NEUTRAL LINER, Use with Crown Cup Only, GXL UHMWPE, SN 0123456, REF 130-32-53. Exactech Hip Systems are indicated for use in surgery for hip replaceme

…and 1 more.

Other Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate devices

All Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate devices →

Sources and updates