Optetrak 200-01-03Recalls & reported problems
Exactech, Inc. · Class 2 · Implant
All models of Optetrak (336)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Exactech Optetrak knee implant is a cemented artificial knee joint. The 25 reports mostly describe implants coming loose years after surgery and needing repeat surgery, with pain, bone loss and infection mentioned. There are no deaths and 25 injuries among the reports. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 25
- Deaths reported
- 0
- Injuries reported
- 25
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 10885862039538), received between Feb 11, 2020 and Apr 20, 2023 — classified as 25 injuries.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (14 reports).
6 other recalls cover devices cleared under the same FDA submission (K932690); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem16
- Loosening of Implant Not Related to Bone-Ingrowth14
- Naturally Worn2
- Device Dislodged or Dislocated1
- Loss of Osseointegration1
Patient problems
- Failure of Implant14
- Bacterial Infection7
- Pain5
- No Information3
- Osteolysis2
- No Consequences Or Impact To Patient1
- Synovitis1
Recent adverse event reports
The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.
InjuryApr 20, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AS REPORTED BY LEGAL NOTIFICATION, THE PATIENT HAD AN INITIAL LEFT TKA ON (B)(6) 2014. THE PATIENT WAS REVISED ON (B)(6) 2022. JOINT FLUID WAS OBTAINED AND SENT FOR CULTURE AND SENSITIVITY X3. EXTENSIVE SYNOVITIS FROM THE POLYETHYLENE WAS NOTED. THE FEMORAL COMPONENT WAS GROSSLY LOOSE AND WAS REMOVED WITH OSTEOTOMES. THE TIBIAL COMPONENT AND PATELLA WERE ALSO GROSSLY LOOSE AND WERE REMOVED. THE PATIENT TOLERATED THE PROCEDURE WELL. NO OTHER PATIENT INFORMATION/MEDICAL HISTORY REPORTED.
Patient problems: Failure of Implant; Synovitis
MAUDE report 16782198 ↗InjuryFeb 14, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED BY (B)(6), THIS (B)(6) Y/O, 165 CM, (B)(6) (BMI=36), FEMALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE RIGHT KNEE VIA MEDIAL PARAPATELLAR APPROACH ON (B)(6) 2009. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. ON (B)(6) 2011, APPROXIMATELY 20 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO TIBIAL OSTEOLYSIS (¿LYSISTIBIA¿) AND PAIN. FOLLOW-UPS FOR ADDITIONAL INFORMATION CANNOT BE PERFORMED FOR THIS COMPLAINT, INFORMATION WAS SUBMITTED AS COMPLETE BY THE (B)(6); THERE IS NO IDENTIFICATION OF FACILITIES OR SURGEONS. ALL AVAILABLE INFORMATION ABOUT THE EVENT AND/OR PATIENT HAS BEEN PROVIDED. THESE DEVICES ARE USED FOR TREATMENT NOT DIAGNOSIS. BASED ON REVIEW OF ALL AVAILABLE INFORMATION, THERE IS NO EVIDENCE TO SUGGEST THAT THE REPORTED EVENT IS RELATED TO ANY DEVICE DESIGN OR MANUFACTURING ISSUES. THE EVENT REPORTED WAS LIKELY THE RESULT OF PATIENT CONDITIONS. THE PATIENT HAS A HIGH BMI WHICH INDICATES THAT THEY ARE OVERWEIGHT, THIS CAN INCREASE BIOMECHANICAL STRESSORS ON NATURAL AND ARTIFICIAL JOINTS. NO ADDITIONAL PATIENT MEDICAL HISTORY INFORMATION IS PROVIDED.
Patient problems: Osteolysis; Pain
MAUDE report 9712555 ↗InjuryFeb 14, 2020Loss of Osseointegration; Device Dislodged or Dislocated; Loosening of Implant Not Related to Bone-Ingrowth; Loss of Osseointegration; Device Dislodged or Dislocated; Loosening of Implant Not Related to Bone-Ingrowth
AS REPORTED BY (B)(6), THIS (B)(6) Y/O, FEMALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE LEFT KNEE VIA MEDIAL PARAPATELLAR APPROACH ON (B)(6) 2010. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. THE PATIENT WAS IMPLANTED WITH A OPTETRAK KNEE SYSTEM - CEMENTED CRUCIATE RETAINED (CR) FEMORAL COMPONENT - SIZE 3 (CATALOG 200-01-03, SERIAL (B)(4)), OPTETRAK KNEE SYSTEM - CEMENTED FINNED TIBIAL TRAY - SIZE 3F/2T (CATALOG 200-04-32, SERIAL (B)(4)), OPTETRAK KNEE SYSTEM - THREE PEGS PATELLA - DIAMETER 29MM (CATALOG 200-02-29 SERIAL (B)(4)) AND HERAEUS MEDICAL ANTIBIOTIC LOADED HIGH VISCOSITY BONE CEMENT (CATALOG 66017569, SERIAL (B)(4)). ON (B)(6) 2015, APPROXIMATELY 62 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO ASEPTIC LOOSENING TIBIA; INSTABILITY; LYSISTIBIA. THE EXACTECH KNEE BRAND REMOVED UNAVAILABLE AND REPLACED.
Patient problems: Failure of Implant; No Consequences Or Impact To Patient
MAUDE report 9712045 ↗InjuryFeb 14, 2020Naturally Worn; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth; Naturally Worn; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth
AS REPORTED BY (B)(6), THIS (B)(6) (BMI=32), MALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE RIGHT KNEE VIA MEDIAL PARAPATELLAR APPROACH ON (B)(6)2009. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. THE PATIENT WAS IMPLANTED WITH A OPTETRAK KNEE SYSTEM - CEMENTED CRUCIATE RETAINED (CR) FEMORAL COMPONENT - SIZE 3 (CATALOG 200-01-03, SERIAL (B)(4)), OPTETRAK KNEE SYSTEM - CEMENTED FINNED TIBIAL TRAY - SIZE 3F/3T (CATALOG 200-04-33, SERIAL (B)(4)), OPTETRAK KNEE SYSTEM - CRUCIATE RETAINED (CR) TIBIAL INSERT - SIZE 3 - 9MM (CATALOG 200-23-09, SERIAL (B)(4)), OPTETRAK KNEE SYSTEM - THREE PEGS PATELLA - DIAMETER 32MM (CATALOG 200-02-32, SERIAL, (B)(4)) AND SCHERING-PLOUGH HIGH VISCOSITY BONE CEMENT (CATALOG 491803, SERIAL (B)(4)). ON (B)(6) 2015, APPROXIMATELY 71 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO ASEPTIC LOOSENING FEMUR; ASEPTIC LOOSENING PATELLA; INSTABILITY; LYSISFEMUR; STIFFNESS; WEAR OF POLYETHYLENE COMPONENT. THE EXACTECH WERE REMOVED AND REPLACED WITH ENDO-MODEL MODULAR ROTATING HINGE FEMORAL MEDIUM 65MM RIGHT AND TRIATHLON SYMMETRIC PATELLA - 36MM X 10MM.
Patient problems: Failure of Implant; Osteolysis
MAUDE report 9711720 ↗InjuryFeb 14, 2020Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AS REPORTED BY (B)(6), THIS (B)(6) Y/O, 157 CM, (B)(6) (BMI=31), FEMALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE LEFT KNEE VIA MEDIAL PARAPATELLAR APPROACH ON (B)(6) 2009. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. THE PATIENT WAS IMPLANTED WITH A OPTETRAK KNEE SYSTEM - CEMENTED CRUCIATE RETAINED (CR) FEMORAL COMPONENT - SIZE 3 (CATALOG 200-01-03, SERIAL (B)(4)), OPTETRAK KNEE SYSTEM - CEMENTED FINNED TIBIAL TRAY - SIZE 3F/3T (CATALOG 200-04-33, SERIAL (B)(4)), OPTETRAK KNEE SYSTEM - CRUCIATE RETAINED (CR) TIBIAL INSERT - SIZE 3 - 9MM (CATALOG 200-23-09, SERIAL (B)(4)) AND HERAEUS MEDICAL ANTIBIOTIC LOADED HIGH VISCOSITY BONE CEMENT (CATALOG 66017569, SERIAL (B)(4)). ON (B)(6) 2012, APPROXIMATELY 39 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO ASEPTIC LOOSENING PATELLA. THE EXACTECH KNEE BRAND REMOVED UNAVAILABLE AND REPLACED WITH PFC« SIGMA OVAL PATELLA 35MM.
Patient problems: Failure of Implant
MAUDE report 9711668 ↗InjuryFeb 14, 2020Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AS REPORTED BY (B)(6), THIS (B)(6) Y/O, 164 CM, (B)(6) (BMI=35), FEMALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE RIGHT KNEE VIA MEDIAL PARAPATELLAR APPROACH ON (B)(6) 2009. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. THE PATIENT WAS IMPLANTED WITH A OPTETRAK KNEE SYSTEM - CEMENTED CRUCIATE RETAINED (CR) FEMORAL COMPONENT - SIZE 3 (CATALOG 200-01-03, SERIAL (B)(4)), OPTETRAK KNEE SYSTEM - CEMENTED FINNED TIBIAL TRAY - SIZE 3F/3T (CATALOG 200-04-33, SERIAL (B)(4)), OPTETRAK KNEE SYSTEM - CRUCIATE RETAINED (CR) TIBIAL INSERT - SIZE 3 - 11MM (CATALOG 200-23-11, SERIAL (B)(4)), OPTETRAK KNEE SYSTEM - THREE PEGS PATELLA - DIAMETER 29MM (CATALOG 200-02-29, SERIAL (B)(4)) AND HERAEUS MEDICAL ANTIBIOTIC LOADED HIGH VISCOSITY BONE CEMENT (CATALOG 66017569, SERIAL (B)(4)). ON (B)(6) 2018, APPROXIMATELY 113 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO ASEPTIC LOOSENING TIBIA. THE EXACTECH FEMORAL COMPONENT REMOVED; TIBIAL COMPONENT REMOVED; TIBIAL LINER COMPONENT REMOVED AND REPLACED WITH PFC« SIGMA TC3 FEMORAL RIGHT SIZE 4, MBT REVISION TIBIAL TRAY CEMENTED SIZE 3 AND TC3 RP TIBIAL INSERT SIZE 4 10MM.
Patient problems: Failure of Implant
MAUDE report 9711560 ↗InjuryFeb 14, 2020Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AS REPORTED BY (B)(6), THIS (B)(6), FEMALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE LEFT KNEE VIA MEDIAL PARAPATELLAR APPROACH ON (B)(6) 2009. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. THE PATIENT WAS IMPLANTED WITH A OPTETRAK KNEE SYSTEM - CEMENTED CRUCIATE RETAINED (CR) FEMORAL COMPONENT - SIZE 3 (CATALOG 200-01-03, SERIAL (B)(4)), OPTETRAK KNEE SYSTEM - CEMENTED FINNED TIBIAL TRAY - SIZE 3F/2T (CATALOG 200-04-32, SERIAL (B)(4)), OPTETRAK KNEE SYSTEM - CRUCIATE RETAINED (CR) TIBIAL INSERT - SIZE 3 - 11MM (CATALOG 200-23-11, SERIAL (B)(4)) AND HERAEUS MEDICAL ANTIBIOTIC LOADED HIGH VISCOSITY BONE CEMENT (CATALOG 66017569, SERIAL (B)(4)). ON (B)(6) 2013, APPROXIMATELY 48 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO ASEPTIC LOOSENING TIBIA. THE EXACTECH WERE REMOVED AND REPLACED WITH SIGMA FEMORAL PS NON-LUGGED CEMENTED SIZE 2.5 LEFT, SIGMA MODULAR TIBIAL TRAY CEMENTED COBALT CHROME SIZE 2.5, SIGMA STABILISED GVF INSERT SIZE 2.5 15MM AND PFC« SIGMA OVAL PATELLA 35MM.
Patient problems: Failure of Implant
MAUDE report 9711528 ↗InjuryFeb 14, 2020Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AS REPORTED BY (B)(6) NATIONAL JOINT REGISTRY (UKNJR), THIS (B)(6) Y/O, 163 CM, (B)(6) KG (BMI=26), FEMALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE LEFT KNEE VIA MEDIAL PARAPATELLAR APPROACH ON (B)(6) 2009. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. THE PATIENT WAS IMPLANTED WITH A OPTETRAK KNEE SYSTEM - CEMENTED CRUCIATE RETAINED (CR) FEMORAL COMPONENT - SIZE 3 (CATALOG 200-01-03, SERIAL (B)(4), OPTETRAK KNEE SYSTEM - CEMENTED FINNED TIBIAL TRAY - SIZE 3F/2T (CATALOG 200-04-32, SERIAL (B)(4), OPTETRAK KNEE SYSTEM - CRUCIATE RETAINED (CR) TIBIAL INSERT - SIZE 3 - 13MM (CATALOG 200-23-13, SERIAL (B)(4), OPTETRAK KNEE SYSTEM - THREE PEGS PATELLA - DIAMETER 29MM (CATALOG 200-02-29, SERIAL (B)(4) AND ZIMMER BIOMET ANTIBIOTIC LOADED HIGH VISCOSITY BONE CEMENT (CATALOG 3-00394-0002, SERIAL (B)(4). ON (B)(6) 2013, APPROXIMATELY 49 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO ASEPTIC LOOSENING TIBIA. THE EXACTECH KNEE BRAND REMOVED UNAVAILABLE AND REPLACED WITH LEGION CONSTRAINED OXINIUM FEMORAL COMPONENT LEFT SZ 4, LEGION REVISION TIBIAL BASE LEFT SZ 3 AND GENESIS II CONSTRAINED ARTICULAR INSERT SIZE 3-4 15MM.
Patient problems: Failure of Implant
MAUDE report 9711499 ↗InjuryFeb 14, 2020Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth
AS REPORTED BY UNITED KINGDOM NATIONAL JOINT REGISTRY (UKNJR), THIS (B)(6) Y/O, 168 CM, (B)(6) KG (BMI=36), FEMALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE RIGHT KNEE VIA MEDIAL PARAPATELLAR APPROACH ON (B)(6) 2008. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. THE PATIENT WAS IMPLANTED WITH A OPTETRAK KNEE SYSTEM - CEMENTED CRUCIATE RETAINED (CR) FEMORAL COMPONENT - SIZE 3 (CATALOG 200-01-03, SERIAL (B)(4)), OPTETRAK KNEE SYSTEM - CEMENTED FINNED TIBIAL TRAY - SIZE 3F/3T (CATALOG 200-04-33, SERIAL (B)(4)), OPTETRAK KNEE SYSTEM - CRUCIATE RETAINED (CR) TIBIAL INSERT - SIZE 3 - 9MM (CATALOG 200-23-09, SERIAL (B)(4)) AND HERAEUS MEDICAL ANTIBIOTIC LOADED HIGH VISCOSITY BONE CEMENT (CATALOG 66017569, SERIAL (B)(4)). ON (B)(6) 2012, APPROXIMATELY 47 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO ASEPTIC LOOSENING TIBIA; PAIN; STIFFNESS. THE EXACTECH KNEE BRAND REMOVED UNAVAILABLE AND REPLACED WITH PFC« SIGMA CR NON-POROUS FEMORAL RIGHT SIZE 3, SIGMA MODULAR TIBIAL TRAY CEMENTED COBALT CHROME SIZE 3, SIGMA CRVD GVF INSERT SIZE 3 12.5MM, PFC« SIGMA OVAL PATELLA 41MM.
Patient problems: Failure of Implant; Pain
MAUDE report 9711259 ↗InjuryFeb 14, 2020Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth
AS REPORTED BY (B)(6) THIS (B)(6) Y/O, 155 CM, (B)(6) KG (BMI=29), FEMALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE LEFT KNEE VIA MEDIAL PARAPATELLAR APPROACH ON (B)(6) 2008. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. ON (B)(6) 2009, APPROXIMATELY 14 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO ASEPTIC LOOSENING OF THE FEMUR, PATIENT CONDITIONS (MALALIGNMENT), AND PAIN. FOLLOW-UPS FOR ADDITIONAL INFORMATION CANNOT BE PERFORMED FOR THIS COMPLAINT, INFORMATION WAS SUBMITTED AS COMPLETE BY THE (B)(6); THERE IS NO IDENTIFICATION OF FACILITIES OR SURGEONS. ALL AVAILABLE INFORMATION ABOUT THE EVENT AND/OR PATIENT HAS BEEN PROVIDED. THESE DEVICES ARE USED FOR TREATMENT NOT DIAGNOSIS. IT IS FOUND IN A REVIEW OF THE LABELING AND IFU 700-096-004, IT IS KNOWN THAT DEVICE SPECIFIC RISKS CONSIST OF: FRACTURE, MIGRATION, LOOSENING, SUBLUXATION, OR DISLOCATION OF THE PROTHESIS OR ANY OF ITS COMPONENTS, ANY OF WHICH MAY REQUIRE A SECOND SURGICAL INTERVENTION OR REVISION. AS PART OF THE PREOPERATIVE ASSESSMENT, THE SURGEON MUST ENSURE THAT NO BIOLOGICAL, BIOMECHANICAL, OR OTHER FACTORS EXIST THAT MIGHT ADVERSELY AFFECT THE SURGERY
Patient problems: Failure of Implant; Pain
MAUDE report 9711151 ↗InjuryFeb 14, 2020Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AS REPORTED BY (B)(6), THIS (B)(6) Y/O (BMI=34), FEMALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE RIGHT KNEE VIA MEDIAL PARAPATELLAR APPROACH ON (B)(6) 2008. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. ON (B)(6) 2010, APPROXIMATELY 31 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO ASEPTIC LOOSENING OF THE FEMUR, PATIENT CONDITIONS (MALALIGNMENT), AND RANGE OF MOTION (STIFFNESS). FOLLOW-UPS FOR ADDITIONAL INFORMATION CANNOT BE PERFORMED FOR THIS COMPLAINT, INFORMATION WAS SUBMITTED AS COMPLETE BY THE (B)(6); THERE IS NO IDENTIFICATION OF FACILITIES OR SURGEONS. ALL AVAILABLE INFORMATION ABOUT THE EVENT AND/OR PATIENT HAS BEEN PROVIDED. THESE DEVICES ARE USED FOR TREATMENT NOT DIAGNOSIS. IT IS FOUND IN A REVIEW OF THE LABELING AND IFU 700-096-004, IT IS KNOWN THAT DEVICE SPECIFIC RISKS CONSIST OF: FRACTURE, MIGRATION, LOOSENING, SUBLUXATION, OR DISLOCATION OF THE PROTHESIS OR ANY OF ITS COMPONENTS, ANY OF WHICH MAY REQUIRE A SECOND SURGICAL INTERVENTION OR REVISION. AS PART OF THE PREOPERATIVE ASSESSMENT, THE SURGEON MUST ENSURE THAT NO BIOLOGICAL, BIOMECHANICAL, OR OTHER FACTORS EXIST THAT MIGHT ADVERSELY AFFECT THE SU
Patient problems: Failure of Implant
MAUDE report 9711025 ↗InjuryFeb 13, 2020Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth
AS REPORTED BY (B)(6), THIS (B)(6) Y/O, FEMALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE LEFT KNEE VIA MEDIAL PARAPATELLAR APPROACH ON (B)(6) 2006. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. ON (B)(6) 2007, APPROXIMATELY 8 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO ASEPTIC LOOSENING OF THE TIBIA AND INSTABILITY. FOLLOW-UPS FOR ADDITIONAL INFORMATION CANNOT BE PERFORMED FOR THIS COMPLAINT, INFORMATION WAS SUBMITTED AS COMPLETE BY THE (B)(6); THERE IS NO IDENTIFICATION OF FACILITIES OR SURGEONS. ALL AVAILABLE INFORMATION ABOUT THE EVENT AND/OR PATIENT HAS BEEN PROVIDED. THESE DEVICES ARE USED FOR TREATMENT NOT DIAGNOSIS. IT IS FOUND IN A REVIEW OF THE LABELING AND IFU 700-096-004, IT IS KNOWN THAT DEVICE SPECIFIC RISKS CONSIST OF: FRACTURE, MIGRATION, LOOSENING, SUBLUXATION, OR DISLOCATION OF THE PROTHESIS OR ANY OF ITS COMPONENTS, ANY OF WHICH MAY REQUIRE A SECOND SURGICAL INTERVENTION OR REVISION. AS PART OF THE PREOPERATIVE ASSESSMENT, THE SURGEON MUST ENSURE THAT NO BIOLOGICAL, BIOMECHANICAL, OR OTHER FACTORS EXIST THAT MIGHT ADVERSELY AFFECT THE SURGERY AND/OR POSTOPERATIVE PERIOD. REVISIONS OR SURGICAL INTERV
Patient problems: Failure of Implant
MAUDE report 9709918 ↗InjuryFeb 13, 2020Naturally Worn; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth; Naturally Worn; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth
AS REPORTED BY (B)(6), THIS (B)(6) Y/O, FEMALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE LEFT KNEE VIA MEDIAL PARAPATELLAR APPROACH ON (B)(6) 2006. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. ON (B)(6) 2015, APPROXIMATELY 117 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO ASEPTIC LOOSENING OF THE TIBIA, INFECTION, PAIN, AND WEAR OF POLYETHYLENE COMPONENT. FOLLOW-UPS FOR ADDITIONAL INFORMATION CANNOT BE PERFORMED FOR THIS COMPLAINT, INFORMATION WAS SUBMITTED AS COMPLETE BY THE (B)(6); THERE IS NO IDENTIFICATION OF FACILITIES OR SURGEONS. ALL AVAILABLE INFORMATION ABOUT THE EVENT AND/OR PATIENT HAS BEEN PROVIDED. THESE DEVICES ARE USED FOR TREATMENT NOT DIAGNOSIS. IT IS FOUND IN A REVIEW OF THE LABELING AND IFU 700-096-004, SURGICAL RISK INCLUDES INFECTION AND PAIN. AS PART OF THE PREOPERATIVE ASSESSMENT, THE SURGEON MUST ENSURE THAT NO BIOLOGICAL, BIOMECHANICAL, OR OTHER FACTORS EXIST THAT MIGHT ADVERSELY AFFECT THE SURGERY AND/OR POSTOPERATIVE PERIOD. REVISIONS OR SURGICAL INTERVENTIONS ARE A KNOWN COMPLICATION FOUND IN JOINT REPLACEMENTS. ALL PATIENTS SHOULD BE INSTRUCTED ON THE LIMITATIONS OF THE PROSTHESIS AND THE
Patient problems: Bacterial Infection; Failure of Implant; Pain
MAUDE report 9709896 ↗InjuryFeb 13, 2020Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AS REPORTED BY (B)(6), THIS (B)(6) Y/O, FEMALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE RIGHT KNEE VIA MEDIAL PARAPATELLAR APPROACH ON (B)(6) 2006. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. ON (B)(6) 2009, APPROXIMATELY 46 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO ASEPTIC LOOSENING OF THE TIBIA. THE FOLLOW-UPS FOR ADDITIONAL INFORMATION CANNOT BE PERFORMED FOR THIS COMPLAINT, INFORMATION WAS SUBMITTED AS COMPLETE BY THE (B)(6); THERE IS NO IDENTIFICATION OF FACILITIES OR SURGEONS. ALL AVAILABLE INFORMATION ABOUT THE EVENT AND/OR PATIENT HAS BEEN PROVIDED. THESE DEVICES ARE USED FOR TREATMENT NOT DIAGNOSIS.IT IS FOUND IN A REVIEW OF THE LABELING AND IFU 700-096-004, IT IS KNOWN THAT DEVICE SPECIFIC RISKS CONSIST OF: FRACTURE, MIGRATION, LOOSENING, SUBLUXATION, OR DISLOCATION OF THE PROTHESIS OR ANY OF ITS COMPONENTS, ANY OF WHICH MAY REQUIRE A SECOND SURGICAL INTERVENTION OR REVISION. AS PART OF THE PREOPERATIVE ASSESSMENT, THE SURGEON MUST ENSURE THAT NO BIOLOGICAL, BIOMECHANICAL, OR OTHER FACTORS EXIST THAT MIGHT ADVERSELY AFFECT THE SURGERY AND/OR POSTOPERATIVE PERIOD. REVISIONS OR SURGICAL INTERVENT
Patient problems: Failure of Implant
MAUDE report 9709861 ↗InjuryFeb 13, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED BY UNITED KINGDOM NATIONAL JOINT REGISTRY (UKNJR), THIS (B)(6) Y/O, FEMALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE RIGHT KNEE VIA MEDIAL PARAPATELLAR APPROACH ON (B)(6) 2008. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. ON (B)(6) 2016, APPROXIMATELY 106 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO OTHER INDICATION FOR REVISION. THE FOLLOW-UPS FOR ADDITIONAL INFORMATION CANNOT BE PERFORMED FOR THIS COMPLAINT, INFORMATION WAS SUBMITTED AS COMPLETE BY THE UNITED KINGDOM NATIONAL JOINT REGISTRY (UKNJR); THERE IS NO IDENTIFICATION OF FACILITIES OR SURGEONS. ALL AVAILABLE INFORMATION ABOUT THE EVENT AND/OR PATIENT HAS BEEN PROVIDED. THESE DEVICES ARE USED FOR TREATMENT NOT DIAGNOSIS. BASED ON REVIEW OF ALL AVAILABLE INFORMATION, THERE IS NO EVIDENCE TO SUGGEST THAT THE REPORTED EVENT IS RELATED TO ANY DEVICE DESIGN OR MANUFACTURING ISSUES. THE EVENT REPORTED WAS LIKELY THE RESULT OF PATIENT CONDITIONS. THERE HAS BEEN NO PATIENT MEDICAL HISTORY/INFORMATION PROVIDED; THEREFORE, THE PATIENT RISK/CLINICAL FACTORS CANNOT BE ASSESSED.
Patient problems: No Information
MAUDE report 9708099 ↗InjuryFeb 13, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED BY (B)(6), THIS (B)(6) Y/O, 180 CM, (B)(6) (BMI=21), MALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT OF THE LEFT KNEE ON (B)(6) 2005. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. ON (B)(6) 2017, APPROXIMATELY 139 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO INFECTION. FOLLOW-UPS FOR ADDITIONAL INFORMATION CANNOT BE PERFORMED FOR THIS COMPLAINT, INFORMATION WAS SUBMITTED AS COMPLETE BY THE (B)(6); THERE IS NO IDENTIFICATION OF FACILITIES OR SURGEONS. ALL AVAILABLE INFORMATION ABOUT THE EVENT AND/OR PATIENT HAS BEEN PROVIDED. THESE DEVICES ARE USED FOR TREATMENT NOT DIAGNOSIS. BASED ON REVIEW OF ALL AVAILABLE INFORMATION, THERE IS NO EVIDENCE TO SUGGEST THAT THE REPORTED EVENT IS RELATED TO ANY DEVICE DESIGN OR MANUFACTURING ISSUES. THE EVENT REPORTED WAS LIKELY THE RESULT OF PATIENT CONDITIONS. THERE HAS BEEN NO PATIENT MEDICAL HISTORY/INFORMATION PROVIDED; THEREFORE, THE PATIENT RISK/CLINICAL FACTORS CANNOT BE ASSESSED.
Patient problems: Bacterial Infection
MAUDE report 9707851 ↗InjuryFeb 13, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED BY (B)(6), THIS (B)(6) Y/O, 165 CM, (B)(6) KG (BMI=37), FEMALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE LEFT KNEE VIA MEDIAL PARAPATELLAR APPROACH ON (B)(6) 2007. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. THE PATIENT WAS IMPLANTED WITH A OPTETRAK KNEE SYSTEM - CEMENTED CRUCIATE RETAINED (CR) FEMORAL COMPONENT - SIZE 3 (CATALOG 200-01-03, SERIAL (B)(4), OPTETRAK KNEE SYSTEM - CEMENTED FINNED TIBIAL TRAY - SIZE 3F/2T (CATALOG 200-04-32, SERIAL (B)(4), OPTETRAK KNEE SYSTEM - CRUCIATE RETAINED (CR) TIBIAL INSERT - SIZE 3 - 13MM (CATALOG 200-23-13, SERIAL (B)(4) AND STRYKER ANTIBIOTIC LOADED MEDIUM VISCOSITY BONE CEMENT (CATALOG 61969001, SERIAL (B)(4). ON (B)(6) 2008, APPROXIMATELY 13 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO OTHER INDICATION FOR REVISION. THE EXACTECH KNEE BRAND REMOVED UNAVAILABLE AND REPLACED WITH OPTETRAK KNEE SYSTEM - THREE PEGS PATELLA - DIAMETER 32MM AND ANTIBIOTIC SIMPLEX ERYTHROMYCIN/COLISTIN (SINGLE PACK).
Patient problems: No Information
MAUDE report 9707795 ↗InjuryFeb 13, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED BY UNITED KINGDOM NATIONAL JOINT REGISTRY (UKNJR), THIS (B)(6) Y/O, MALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE LEFT KNEE VIA MEDIAL PARAPATELLAR APPROACH ON (B)(6) 2007. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. ON (B)(6) 2016, APPROXIMATELY 118 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO OTHER INDICATION FOR REVISION. THE FOLLOW-UPS FOR ADDITIONAL INFORMATION CANNOT BE PERFORMED FOR THIS COMPLAINT, INFORMATION WAS SUBMITTED AS COMPLETE BY THE UNITED KINGDOM NATIONAL JOINT REGISTRY (UKNJR); THERE IS NO IDENTIFICATION OF FACILITIES OR SURGEONS. ALL AVAILABLE INFORMATION ABOUT THE EVENT AND/OR PATIENT HAS BEEN PROVIDED. THESE DEVICES ARE USED FOR TREATMENT NOT DIAGNOSIS. BASED ON REVIEW OF ALL AVAILABLE INFORMATION, THERE IS NO EVIDENCE TO SUGGEST THAT THE REPORTED EVENT IS RELATED TO ANY DEVICE DESIGN OR MANUFACTURING ISSUES. THE EVENT REPORTED WAS LIKELY THE RESULT OF PATIENT CONDITIONS. THERE HAS BEEN NO PATIENT MEDICAL HISTORY/INFORMATION PROVIDED; THEREFORE, THE PATIENT RISK/CLINICAL FACTORS CANNOT BE ASSESSED.
Patient problems: No Information
MAUDE report 9707626 ↗InjuryFeb 13, 2020Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AS REPORTED BY UNITED KINGDOM NATIONAL JOINT REGISTRY (UKNJR), THIS (B)(6) Y/O, FEMALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE LEFT KNEE VIA LATERAL PARAPATELLAR APPROACH ON (B)(6) 2005. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. ON (B)(6) 2015, APPROXIMATELY 121 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO ASEPTIC LOOSENING OF THE TIBIA. FOLLOW-UPS FOR ADDITIONAL INFORMATION CANNOT BE PERFORMED FOR THIS COMPLAINT, INFORMATION WAS SUBMITTED AS COMPLETE BY THE UNITED KINGDOM NATIONAL JOINT REGISTRY (UKNJR); THERE IS NO IDENTIFICATION OF FACILITIES OR SURGEONS. ALL AVAILABLE INFORMATION ABOUT THE EVENT AND/OR PATIENT HAS BEEN PROVIDED. THESE DEVICES ARE USED FOR TREATMENT NOT DIAGNOSIS. IT IS FOUND IN A REVIEW OF THE LABELING AND IFU 700-096-004, IT IS KNOWN THAT DEVICE SPECIFIC RISKS CONSIST OF: FRACTURE, MIGRATION, LOOSENING, SUBLUXATION, OR DISLOCATION OF THE PROTHESIS OR ANY OF ITS COMPONENTS, ANY OF WHICH MAY REQUIRE A SECOND SURGICAL INTERVENTION OR REVISION. SURGICAL RISK INCLUDES INFECTION. AS PART OF THE PREOPERATIVE ASSESSMENT, THE SURGEON MUST ENSURE THAT NO BIOLOGICAL, BIOMECHANICAL, OR OTHER FACTORS EX
Patient problems: Failure of Implant
MAUDE report 9707439 ↗InjuryFeb 13, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED BY (B)(6), THIS (B)(6) Y/O, 165 CM, (B)(6) (BMI=22), FEMALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE RIGHT KNEE VIA MEDIAL PARAPATELLAR APPROACH ON (B)(6) 2008. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. ON (B)(6) 2009, APPROXIMATELY 10 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO PAIN. THE UNDERLYING CAUSE FOR PAIN WAS NOT REPORTED. FOLLOW-UPS FOR ADDITIONAL INFORMATION CANNOT BE PERFORMED FOR THIS COMPLAINT, INFORMATION WAS SUBMITTED AS COMPLETE BY THE (B)(6); THERE IS NO IDENTIFICATION OF FACILITIES OR SURGEONS. ALL AVAILABLE INFORMATION ABOUT THE EVENT AND/OR PATIENT HAS BEEN PROVIDED. THESE DEVICES ARE USED FOR TREATMENT NOT DIAGNOSIS. BASED ON REVIEW OF ALL AVAILABLE INFORMATION, THERE IS NO EVIDENCE TO SUGGEST THAT THE REPORTED EVENT IS RELATED TO ANY DEVICE DESIGN OR MANUFACTURING ISSUES. THE EVENT REPORTED WAS LIKELY THE RESULT OF PATIENT CONDITIONS. THERE HAS BEEN NO PATIENT MEDICAL HISTORY INFORMATION PROVIDED; THEREFORE, THE PATIENT RISK/CLINICAL FACTORS CANNOT BE ASSESSED.
Patient problems: Pain
MAUDE report 9706436 ↗InjuryFeb 11, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED BY UNITED KINGDOM NATIONAL JOINT REGISTRY (UKNJR), A (B)(6) Y/O, 166 CM, (B)(6) KG (BMI=(B)(6)), FEMALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT OF THE RIGHT KNEE ON (B)(6) 2009. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. ON (B)(6) 2010, APPROXIMATELY 20 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO INFECTION. FOLLOW-UPS FOR ADDITIONAL INFORMATION CANNOT BE PERFORMED FOR THIS COMPLAINT, INFORMATION WAS SUBMITTED AS COMPLETE BY THE UNITED KINGDOM NATIONAL JOINT REGISTRY (UKNJR); THERE IS NO IDENTIFICATION OF FACILITIES OR SURGEONS. ALL AVAILABLE INFORMATION ABOUT THE EVENT AND/OR PATIENT HAS BEEN PROVIDED. THESE DEVICES ARE USED FOR TREATMENT NOT DIAGNOSIS. BASED ON REVIEW OF ALL AVAILABLE INFORMATION, THERE IS NO EVIDENCE TO SUGGEST THAT THE REPORTED EVENT IS RELATED TO ANY DEVICE DESIGN OR MANUFACTURING ISSUES. THE EVENT REPORTED WAS LIKELY THE RESULT OF PATIENT CONDITIONS. THERE HAS BEEN NO PATIENT MEDICAL HISTORY/INFORMATION PROVIDED; THEREFORE, THE PATIENT RISK/CLINICAL FACTORS CANNOT BE ASSESSED.
Patient problems: Bacterial Infection
MAUDE report 9697170 ↗InjuryFeb 11, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED BY (B)(6), THIS (B)(6) Y/O, 156 CM, (B)(6) KG (BMI=15), FEMALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE LEFT KNEE VIA MEDIAL PARAPATELLAR APPROACH ON (B)(6) 2008. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. ON (B)(6) 2009, APPROXIMATELY 12 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO INFECTION. FOLLOW-UPS FOR ADDITIONAL INFORMATION CANNOT BE PERFORMED FOR THIS COMPLAINT, INFORMATION WAS SUBMITTED AS COMPLETE BY THE (B)(6); THERE IS NO IDENTIFICATION OF FACILITIES OR SURGEONS. ALL AVAILABLE INFORMATION ABOUT THE EVENT AND/OR PATIENT HAS BEEN PROVIDED. THESE DEVICES ARE USED FOR TREATMENT NOT DIAGNOSIS. BASED ON REVIEW OF ALL AVAILABLE INFORMATION, THERE IS NO EVIDENCE TO SUGGEST THAT THE REPORTED EVENT IS RELATED TO ANY DEVICE DESIGN OR MANUFACTURING ISSUES. THE EVENT REPORTED WAS LIKELY THE RESULT OF PATIENT CONDITIONS. THERE HAS BEEN NO PATIENT MEDICAL HISTORY INFORMATION PROVIDED; THEREFORE, THE PATIENT RISK/CLINICAL FACTORS CANNOT BE ASSESSED.
Patient problems: Bacterial Infection
MAUDE report 9697144 ↗InjuryFeb 11, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED BY UNITED KINGDOM NATIONAL JOINT REGISTRY (UKNJR), THIS (B)(6) Y/O, 165 CM, (B)(6) KG (BMI=(B)(6)), FEMALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT OF THE LEFT KNEE ON (B)(6) 2008. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. ON (B)(6) 2009, APPROXIMATELY 4 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO INFECTION. FOLLOW-UPS FOR ADDITIONAL INFORMATION CANNOT BE PERFORMED FOR THIS COMPLAINT, INFORMATION WAS SUBMITTED AS COMPLETE BY THE UNITED KINGDOM NATIONAL JOINT REGISTRY (UKNJR); THERE IS NO IDENTIFICATION OF FACILITIES OR SURGEONS. ALL AVAILABLE INFORMATION ABOUT THE EVENT AND/OR PATIENT HAS BEEN PROVIDED. THESE DEVICES ARE USED FOR TREATMENT NOT DIAGNOSIS. STERILE CERTIFICATION REVIEWED; DEVICES WERE STERILIZED AND RELEASED ACCORDING TO SPECIFICATIONS. BASED ON REVIEW OF ALL AVAILABLE INFORMATION, THERE IS NO EVIDENCE TO SUGGEST THAT THE REPORTED EVENT IS RELATED TO ANY DEVICE DESIGN OR MANUFACTURING ISSUES. THE EVENT REPORTED WAS LIKELY THE RESULT OF PATIENT CONDITIONS. THERE HAS BEEN NO PATIENT MEDICAL HISTORY INFORMATION PROVIDED; THEREFORE, THE PATIENT RISK/CLINICAL FACTORS CANNOT BE ASSESSED.
Patient problems: Bacterial Infection
MAUDE report 9697123 ↗InjuryFeb 11, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED BY UNITED KINGDOM NATIONAL JOINT REGISTRY (UKNJR), A (B)(6) Y/O, 156 CM, (B)(6) KG (BMI=(B)(6)), FEMALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT OF THE RIGHT KNEE ON (B)(6) 2008. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. ON (B)(6) 2012, APPROXIMATELY 55 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO INFECTION. FOLLOW-UPS FOR ADDITIONAL INFORMATION CANNOT BE PERFORMED FOR THIS COMPLAINT, INFORMATION WAS SUBMITTED AS COMPLETE BY THE UNITED KINGDOM NATIONAL JOINT REGISTRY (UKNJR); THERE IS NO IDENTIFICATION OF FACILITIES OR SURGEONS. ALL AVAILABLE INFORMATION ABOUT THE EVENT AND/OR PATIENT HAS BEEN PROVIDED. THESE DEVICES ARE USED FOR TREATMENT NOT DIAGNOSIS. BASED ON REVIEW OF ALL AVAILABLE INFORMATION, THERE IS NO EVIDENCE TO SUGGEST THAT THE REPORTED EVENT IS RELATED TO ANY DEVICE DESIGN OR MANUFACTURING ISSUES. THE EVENT REPORTED WAS LIKELY THE RESULT OF PATIENT CONDITIONS. THERE HAS BEEN NO PATIENT MEDICAL HISTORY/INFORMATION PROVIDED; THEREFORE, THE PATIENT RISK/CLINICAL FACTORS CANNOT BE ASSESSED.
Patient problems: Bacterial Infection
MAUDE report 9695759 ↗InjuryFeb 11, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED BY UNITED KINGDOM NATIONAL JOINT REGISTRY (UKNJR), THIS (B)(6) YEARS OLD, MALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE RIGHT KNEE ON (B)(6) 2007. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. ON (B)(6) 2008, APPROXIMATELY 10 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO INFECTION AND STIFFNESS. FOLLOW-UPS FOR ADDITIONAL INFORMATION CANNOT BE PERFORMED FOR THIS COMPLAINT, INFORMATION WAS SUBMITTED AS COMPLETE BY THE UNITED KINGDOM NATIONAL JOINT REGISTRY (UKNJR); THERE IS NO IDENTIFICATION OF FACILITIES OR SURGEONS. ALL AVAILABLE INFORMATION ABOUT THE EVENT AND/OR PATIENT HAS BEEN PROVIDED. THESE DEVICES ARE USED FOR TREATMENT NOT DIAGNOSIS. BASED ON REVIEW OF ALL AVAILABLE INFORMATION, THERE IS NO EVIDENCE TO SUGGEST THAT THE REPORTED EVENT IS RELATED TO ANY DEVICE DESIGN OR MANUFACTURING ISSUES. THE EVENT REPORTED WAS LIKELY THE RESULT OF PATIENT CONDITIONS. THERE HAS BEEN NO PATIENT MEDICAL HISTORY INFORMATION PROVIDED; THEREFORE, THE PATIENT RISK/CLINICAL FACTORS CANNOT BE ASSESSED.
Patient problems: Bacterial Infection
MAUDE report 9695596 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K932690. One submission often covers many models, so check the product and lot details in each record.
- Z-2157-2024 · Apr 18, 2024 · Open, ClassifiedOPTETRAK Inset Patella, Item Numbers: a) 200-05-23, 23MM; b) 200-05-26, 26MM; c) 200-05-29, 29MM
- Z-2156-2024 · Apr 18, 2024 · Open, ClassifiedOPTETRAK One Peg Patella,, Item Numbers: a) 200-03-26, 26MM; b) 200-03-29, 29MM; c) 200-03-32, 32MM; d) 200-03-35, 35MM; e) 200-03-38, 38MM; f) 200-03-41, 41MM
- Z-2155-2024 · Apr 18, 2024 · Open, ClassifiedOPTETRAK Three Peg Patella, Item Numbers: a) 200-02-26, 26MM; b) 200-02-29, 29MM; c) 200-02-32, 32MM; d) 200-02-35, 35MM; e) 200-02-38, 38MM; f) 200-02-41, 41MM;
- Z-2158-2024 · Apr 18, 2024 · Open, ClassifiedOPTETRAK, Advanced Patella, 3 Peg Implant, Item Numbers: a) 200-07-26, 26MM; b) 200-07-29, 29MM; c) 200-07-32, 32MM; d) 200-07-35, 35MM; e) 200-07-38, 38MM
- Z-0582-2024 · Nov 21, 2023 · Open, ClassifiedExactech Optetrak Knee System, Catalog Numbers: a) 200-02-26, b) 200-02-29, c) 200-02-32, d) 200-02-35, e) 200-02-38, f) 200-02-41, g) 200-03-32, h) 200-07-29, i) 200-07-
- Z-0019-2022 · Aug 30, 2021 · Open, ClassifiedOPTETRAK Comprehensive Knee System, labeled as the following: a. OPTETRAK All-polyethylene CR Tibial Components, with Product Line Numbers: 200-11-XX, 200-12-XX, 200-13-XX, 200-1
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page