Optetrak 200-01-05Recalls & reported problems
Exactech, Inc. · Class 2 · Implant
All models of Optetrak (336)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10885862039552), received between Feb 11, 2020 and Feb 14, 2020 — classified as 3 injuries.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).
6 other recalls cover devices cleared under the same FDA submission (K932690); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Loosening of Implant Not Related to Bone-Ingrowth1
Patient problems
- Bacterial Infection2
- Failure of Implant2
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryFeb 14, 2020Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth
AS REPORTED BY (B)(6), THIS (B)(6), 178 CM, (B)(6) (BMI=38), MALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT USING CEMENT OF THE LEFT KNEE VIA MEDIAL PARAPATELLAR APPROACH ON (B)(6) 2008. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. THE PATIENT WAS IMPLANTED WITH A OPTETRAK KNEE SYSTEM - CEMENTED CRUCIATE RETAINED (CR) FEMORAL COMPONENT - SIZE 5 (CATALOG 200-01-05, SERIAL (B)(4)), OPTETRAK KNEE SYSTEM - CEMENTED FINNED TIBIAL TRAY - SIZE 5F/5T (CATALOG 200-04-55, SERIAL (B)(4)), OPTETRAK KNEE SYSTEM - CRUCIATE RETAINED (CR) TIBIAL INSERT - SIZE 5 - 9MM (CATALOG 200-25-09, SERIAL (B)(4)) AND HERAEUS MEDICAL ANTIBIOTIC LOADED HIGH VISCOSITY BONE CEMENT (CATALOG 66017569, SERIAL (B)(4)). ON (B)(6) 2011, APPROXIMATELY 31 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO ASEPTIC LOOSENING FEMUR; ASEPTIC LOOSENING TIBIA; INFECTION. THE EXACTECH WERE REMOVED AND REPLACED.
Patient problems: Bacterial Infection; Failure of Implant
MAUDE report 9711175 ↗InjuryFeb 12, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED BY UNITED KINGDOM NATIONAL JOINT REGISTRY (UKNJR), A (B)(6) Y/O, MALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT OF THE RIGHT KNEE ON (B)(6) 2007. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. ON (B)(6) 2012, APPROXIMATELY 62 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO INSTABILITY, PAIN, AND STIFFNESS. FOLLOW-UPS FOR ADDITIONAL INFORMATION CANNOT BE PERFORMED FOR THIS COMPLAINT, INFORMATION WAS SUBMITTED AS COMPLETE BY THE UNITED KINGDOM NATIONAL JOINT REGISTRY (UKNJR); THERE IS NO IDENTIFICATION OF FACILITIES OR SURGEONS. ALL AVAILABLE INFORMATION ABOUT THE EVENT AND/OR PATIENT HAS BEEN PROVIDED. THESE DEVICES ARE USED FOR TREATMENT NOT DIAGNOSIS. BASED ON REVIEW OF ALL AVAILABLE INFORMATION, THERE IS NO EVIDENCE TO SUGGEST THAT THE REPORTED EVENT IS RELATED TO ANY DEVICE DESIGN OR MANUFACTURING ISSUES. THE EVENT REPORTED WAS LIKELY THE RESULT OF PATIENT CONDITIONS. THERE HAS BEEN NO PATIENT MEDICAL HISTORY/INFORMATION PROVIDED; THEREFORE, THE PATIENT RISK/CLINICAL FACTORS CANNOT BE ASSESSED.
Patient problems: Failure of Implant
MAUDE report 9703544 ↗InjuryFeb 11, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED BY UNITED KINGDOM NATIONAL JOINT REGISTRY (UKNJR), THIS (B)(6) YEARS OLD, MALE PATIENT UNDERWENT PRIMARY TOTAL PROSTHETIC REPLACEMENT OF THE LEFT KNEE ON (B)(6) 2007. THE INDICATION FOR THE INDEX PROCEDURE WAS OSTEOARTHRITIS. ON (B)(6) 2007, APPROXIMATELY 1-MONTH POST IMPLANTATION, THE PATIENT UNDERWENT REVISION DUE TO INFECTION. FOLLOW-UPS FOR ADDITIONAL INFORMATION CANNOT BE PERFORMED FOR THIS COMPLAINT, INFORMATION WAS SUBMITTED AS COMPLETE BY THE UNITED KINGDOM NATIONAL JOINT REGISTRY (UKNJR); THERE IS NO IDENTIFICATION OF FACILITIES OR SURGEONS. ALL AVAILABLE INFORMATION ABOUT THE EVENT AND/OR PATIENT HAS BEEN PROVIDED. THESE DEVICES ARE USED FOR TREATMENT NOT DIAGNOSIS. STERILE CERTIFICATION REVIEWED; DEVICES WERE STERILIZED AND RELEASED ACCORDING TO SPECIFICATIONS. BASED ON REVIEW OF ALL AVAILABLE INFORMATION, THERE IS NO EVIDENCE TO SUGGEST THAT THE REPORTED EVENT IS RELATED TO ANY DEVICE DESIGN OR MANUFACTURING ISSUES. THE EVENT REPORTED WAS LIKELY THE RESULT OF PATIENT CONDITIONS. THERE HAS BEEN NO PATIENT MEDICAL HISTORY INFORMATION PROVIDED; THEREFORE, THE PATIENT RISK/CLINICAL FACTORS CANNOT BE ASSESSED.
Patient problems: Bacterial Infection
MAUDE report 9695519 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K932690. One submission often covers many models, so check the product and lot details in each record.
- Z-2156-2024 · Apr 18, 2024 · Open, ClassifiedOPTETRAK One Peg Patella,, Item Numbers: a) 200-03-26, 26MM; b) 200-03-29, 29MM; c) 200-03-32, 32MM; d) 200-03-35, 35MM; e) 200-03-38, 38MM; f) 200-03-41, 41MM
- Z-2158-2024 · Apr 18, 2024 · Open, ClassifiedOPTETRAK, Advanced Patella, 3 Peg Implant, Item Numbers: a) 200-07-26, 26MM; b) 200-07-29, 29MM; c) 200-07-32, 32MM; d) 200-07-35, 35MM; e) 200-07-38, 38MM
- Z-2157-2024 · Apr 18, 2024 · Open, ClassifiedOPTETRAK Inset Patella, Item Numbers: a) 200-05-23, 23MM; b) 200-05-26, 26MM; c) 200-05-29, 29MM
- Z-2155-2024 · Apr 18, 2024 · Open, ClassifiedOPTETRAK Three Peg Patella, Item Numbers: a) 200-02-26, 26MM; b) 200-02-29, 29MM; c) 200-02-32, 32MM; d) 200-02-35, 35MM; e) 200-02-38, 38MM; f) 200-02-41, 41MM;
- Z-0582-2024 · Nov 21, 2023 · Open, ClassifiedExactech Optetrak Knee System, Catalog Numbers: a) 200-02-26, b) 200-02-29, c) 200-02-32, d) 200-02-35, e) 200-02-38, f) 200-02-41, g) 200-03-32, h) 200-07-29, i) 200-07-
- Z-0019-2022 · Aug 30, 2021 · Open, ClassifiedOPTETRAK Comprehensive Knee System, labeled as the following: a. OPTETRAK All-polyethylene CR Tibial Components, with Product Line Numbers: 200-11-XX, 200-12-XX, 200-13-XX, 200
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page