Skip to content

Optetrak 200-04-34Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak (336)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885862039743), received between May 27, 2025 and May 27, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Fracture” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K932776); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1
  • Naturally Worn1
  • Unstable1

Patient problems

  • Ambulation or Postural Difficulties1
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Implant Pain1
  • Joint Laxity1
  • Metal Related Pathology1
  • Osteolysis1
  • Swelling/ Edema1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 27, 2025Fracture; Unstable; Naturally Worn; Fracture; Unstable; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 97 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, FRACTURE OF THE POLY LINER AND POSTERIOR MEDIAL TIBIAL COMPONENT, METALLOSIS, INSTABILITY, LIMITED MOBILITY, SWELLING, KNEE PAIN, AND A GRINDING SENSATION. INITIAL SURGERY AND REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. X-RAYS REVEALED METAL ON METAL BILATERAL ARTICULATION WITH FRACTURE POSTERIOR MEDIAL RIGHT TIBIAL COMPONENT AS WELL AS POLY AND FRACTURE LEFT POSTERIOR LATERAL TIBIAL COMPONENT AND POLY. INTRAOPERATIVE FINDINGS INDICATED SEVERE BLACK TISSUE STAINING/METALLOSIS THROUGHOUT THE KNEE AND OSTEOLYSIS OF THE FEMORAL. THE POLY LINER WAS TO HAVE SEVERE WEAR. THERE WAS MILD LOOSENING OF FEMORAL COMPONENT. THE TIBIAL COMPONENT HAD FRACTURED POSTERIOR MEDIALLY WHERE THE METAL FEMORAL COMPONENT HAD BEEN ARTICULATING WITH THE METAL POSTERIOR EDGE OF THE TIBIAL COMPONENT. MULTIPLE SMALL FRAGMENTS OF THE FRACTURED POLY WERE PRESENT AND REMOVED. THE PATELLA HAD SEVERE WEAR PRESENT. THE PATIENT WAS REVISED TO A COMPETITOR¿S DEVICES. THE PATIENT APPEARED TO HAVE TOLERATED THE PROCEDURE WELL AND W

    Patient problems: Ambulation or Postural Difficulties; Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Implant Pain; Joint Laxity; Metal Related Pathology; Osteolysis; Swelling/ Edema

    MAUDE report 22091378 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K932776. One submission often covers many models, so check the product and lot details in each record.

  • Z-0927-06 · Dec 19, 2005 · Terminated
    Product is Optetrak Cemented Finned Tibial Tray Sz. 4F/5T. Knee prosthesis. Product is labeled in part: ''OPTETRAK FINNED TIBIAL TRAY A/P53 mm M/L 83mm***STERILE, SINGLE USE ONL
  • Z-1379-05 · Jun 2, 2005 · Terminated
    The product is a Finned Tibial Tray. Product is labeled in part: ''OPTETRAK FINNED TIBIAL TRAY A/P 53 mm ML 83mm***4F/5T Ti Alloy FITS ONLY SZ. 4 FEMORAL CEMENTED***STERILE, SING

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates