Skip to content

Optetrak 200-72-09Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak (336)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
0
Injuries reported
8
Recalls naming it
1

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10885862041210), received between Dec 13, 2022 and Apr 8, 2026 — classified as 8 injuries.

The most frequently reported specific device problem is “Naturally Worn” (8 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2021Open, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0019-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn8
  • Adverse Event Without Identified Device or Use Problem1
  • Fracture1
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Material Integrity Problem1
  • Peeled/Delaminated1
  • Structural Problem1
  • Unstable1

Patient problems

  • Failure of Implant4
  • Pain4
  • Metal Related Pathology3
  • Swelling/ Edema3
  • Synovitis3
  • Implant Pain2
  • Joint Laxity2
  • Osteolysis2

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 8, 2026Naturally Worn

    AS REPORTED, APPROXIMATELY 7.5 YEARS POST THE INITIAL RIGHT TOTAL KNEE ARTHROPLASTY, THE PATIENT COMPLAINED OF KNEE SWELLING AND PAIN AND WAS DIAGNOSED WITH WEAR OF THE TIBIAL INSERT. THE PATIENT WAS REVISED AND DAMAGE AND WEAR WERE OBSERVED ON THE RETRIEVED TIBIAL INSERT, AS WELL AS THE TIBIAL TRAY. THE SURGEON DECIDED THAT ONLY THE INSERT AND TIBIAL TRAY NEEDED TO BE REPLACED WITH A TRULIANT CRC SIZE 2 15 MM INSERT AND TIBIAL TRAY 2F/1T. A STEM EXTENTION 14MM×40MM, 1/2 TIBIAL AUGMENT SIZE 1T-5MM (X2), 1/2 TIBIAL AUGMENT RM/LL SIZE 1T-10MM AND 1/2 TIBIAL AUGMENT RL/LM SIZE 1T-10MM WERE ALSO USED. THERE WAS NO REPORTED SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 24815251 ↗
  • InjuryJan 7, 2025Fracture; Structural Problem; Naturally Worn; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth; Fracture; Structural Problem; Naturally Worn; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth

    LEGAL CASE ¿ USA. PATIENT ID: (B)(6) + RIGHT KNEE. CASE (B)(4) IS ASSOCIATED WITH THIS CASE. IT WAS REPORTED THAT APPROXIMATELY 123 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT AN ARTHROSCOPY PROCEDURE DUE TO SYNOVITIS AND LOOSENING. PATIENT BEGAN EXPERIENCING PAIN, POPPING, CLICKING AND CATCHING OF THE RIGHT KNEE AFTER A FALL APPROXIMATELY 7 MONTHS PRIOR. ARTHROSCOPIC SURGERY OPERATIVE NOTES WERE PROVIDED. POSTOPERATIVE DIAGNOSIS INDICATED SYNOVITIS, TIBIAL COMPONENT LOOSENING, METALLOSIS, AND FRACTURED TIBIAL POLYETHYLENE WITH LOOSE FRAGMENTS. SURGEON PERFORMED A THREE COMPARTMENT SYNOVECTOMY AND LOOSE BODY REMOVAL FROM WITHIN THE NOTCH. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. NO INTRAOPERATIVE OR POSTOPERATIVE COMPLICATIONS. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Joint Laxity; Metal Related Pathology; Pain; Peeling; Synovitis

    MAUDE report 21092507 ↗
  • InjurySep 3, 2024Peeled/Delaminated; Naturally Worn; Peeled/Delaminated; Naturally Worn

    IT WAS REPORTED THAT A FEMALE PATIENT, INITIAL LEFT KNEE IMPLANTED IN (B)(6) 2016, UNDERWENT A REVISION PROCEDURE IN (B)(6) 2024, APPROXIMATELY 8 YEARS 5 MONTHS POST THE INITIAL PROCEDURE. THE PATIENT WHO HAD A TOTAL KNEE ARTHROPLASTY USING CR SLOPE ++ TIBIAL INSERT COMPLAINED OF KNEE SWELLING AND PAIN. THE SURGEON DIAGNOSED THE WEAR AND DAMAGE OF THE TIBIAL INSERT, THEN REVISION SURGERY FOR THE REPLACEMENT OF THE TIBIAL INSERT WAS PERFORMED. BREAKAGE AND WEAR WERE OBSERVED IN THE RETRIEVED TIBIAL INSERT. NO PARTS OR PIECES OF THE TIBIAL TRAY FELL INTO THE WOUND SITE. THE SURGEON DECIDED THAT NO PROBLEM OF LOCKING MECHANISM OF THE TIBIAL TRAY WAS FOUND, AND THE INSERT ONLY WAS REPLACED TO A NEW TRULIANT CR INSERT NEUTRAL SIZE 2 12MM. THERE WERE NO SURGICAL DELAYS DURING THE PROCEDURE. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. AN X-RAY WAS PROVIDED. THE EXPLANTED DEVICE IS NOT RETURNING: A RETRIEVED INSERT WOULD BE USED FOR CLARIFICATION TO A PATIENT. DEVICE IMAGES OF THE EXPLANTED DEVICE WAS PROVIDED. NO FURTHER INFORMATION.

    Patient problems: Implant Pain; Swelling/ Edema

    MAUDE report 20131558 ↗
  • InjurySep 3, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 136 MONTHS AFTER A RIGHT KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, SCAR TISSUE AND SYNOVITIS. OP REPORT NOTED POLY HAD SEVERE WEAR PATTERN AND PULSED. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Scar Tissue; Synovitis

    MAUDE report 20125373 ↗
  • InjuryMay 18, 2023Naturally Worn; Naturally Worn

    DURING THE WEEK OF APRIL (B)(6), REPRESENTATIVES OF EXACTECH, INC. AND EXACTECH SPAIN CONDUCTED AN INPERSON SITE VISIT WITH UBARMIN HOSPITAL. AS A RESULT OF THAT SITE VISIT AND AT EXACTECH'S REQUEST, THE HOSPITAL PROVIDED A RETROSPECTIVE EXCEL LISTING OF REVISION RELATED TO POLYETHYLENE WEAR THAT HAVE BEEN PERFORMED SINCE 2010. THESE CASES HAVE NOT PREVIOUSLY BEEN REPORTED TO EXACTECH AS COMPLAINTS. THIS PATIENT ((B)(6)) WAS IMPLANTED WITH A 200-72-09-INSERTO TIBIAL SLOPE++ CR 2-9 MM, SERIAL NUMBER (B)(6) ON (B)(6) 2015. THE INSERT WAS MANUFACTURED ON 4/17/2013 AND WAS PACKAGED IN A VACUUM BAG WITH NO EVOH. THE INSERT WAS MANUFACTURED ON 4/17/2013 AND WAS 2.01 YEARS SHELF AGE AT THE TIME OF IMPLANT. ON (B)(6) 2022 APPROXIMATELY 7.0 YEARS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION FOR WEAR OF THE POLYETHYLENE. NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 16958346 ↗
  • InjuryApr 13, 2023Unstable; Naturally Worn; Unstable; Naturally Worn

    AS REPORTED PER LEGAL TEAM, THE 77 Y/O FEMALE PATIENT THAT HAD AN ORIGINAL KNEE REPLACEMENT ROUGHLY 10 YEARS AGO. THE PATIENT IS NEEDING A FULL KNEE REVISION DUE TO LYSIS ON THE BONE AND POLY WEAR.

    Patient problems: Failure of Implant; Implant Pain; Joint Laxity; Metal Related Pathology; Osteolysis; Synovitis

    MAUDE report 16736632 ↗
  • InjuryMar 27, 2023Material Integrity Problem; Naturally Worn; Material Integrity Problem; Naturally Worn

    ADDITIONAL INFORMATION RECEIVED BY THE LEGAL DEPARTMENT ON (B)(6) 2023 : ON (B)(6) 2023 , THE PATIENT UNDERWENT A LEFT TOTAL KNEE REPLACEMENT. ON (B)(6) 2017, THE PATIENT UNDERWENT REVISION OF THE EXACTECH KNEE DEVICE SECONDARY TO POLYETHYLENE LINER WEAR WITH RESULTING OSTEOLYSIS, LINER BREAKAGE, AND ASEPTIC LOOSENING. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Ambulation or Postural Difficulties; Balance Problems; Discomfort; Failure of Implant; Joint effusion; Osteolysis; Pain; Scar Tissue

    MAUDE report 16617385 ↗
  • InjuryDec 13, 2022Naturally Worn; Naturally Worn

    AS REPORTED, THIS PATIENT HAD AN INITIAL RIGHT TKA ON (B)(6) 2015. THE PATIENT COMPLAINED OF KNEE SWELLING AND PAIN AND THE SURGEON DIAGNOSED THE WEAR OF THE TIBIAL INSERT. THE PATIENT WAS REVISED ON (B)(6) 2022. BREAKAGE AND WEAR WAS OBSERVED IN THE RETRIEVED TIBIAL INSERT, AS WELL AS METALLOSIS. THE SURGEON REMOVED PARTS AND PIECES OF POLYETHYLENE PARTICLES AND PROLIFERATIVE TISSUES WERE RESECTED BY THE SURGEON APPROPRIATELY.

    Patient problems: Failure of Implant; Metal Related Pathology; Pain; Swelling/ Edema

    MAUDE report 15969400 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page