Skip to content

Optetrak 208-01-05Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak (336)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10885862053244), received between Sep 19, 2023 and Jan 27, 2025 — classified as 4 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (4 reports).

1 other recall cover devices cleared under the same FDA submission (K954208); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth4
  • Unstable3
  • Insufficient Device Problem Information1

Patient problems

  • Joint Laxity3
  • Pain3
  • Osteolysis2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Bone Fracture(s)1
  • Cyst(s)1
  • Failure of Implant1
  • Implant Pain1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 27, 2025Unstable; Loosening of Implant Not Related to Bone-Ingrowth; Unstable; Loosening of Implant Not Related to Bone-Ingrowth

    ADDITIONAL INFORMATION RECEIVED FROM LEGAL ON 28-DEC-2023. LEGAL CASE ¿ USA . PATIENT ID: (B)(6) LK REV 2. RELATED (B)(4) LK REV 1. IMPLANT DATE: (B)(6) 2011, DR. (B)(6), EXPLANT DATE: (B)(6) 2017, DR. (B)(6). NO DEVICE RETURN ANTICIPATED DUE TO THIS BEING A LEGAL CASE. UNABLE TO LOCATE SURGEON/PATIENT/SERIAL NUMBER IN EBI. SERIAL NUMBERS PROVIDED. (B)(6), 208-25-22 - INSERT, TIBIA CONSTRAINED CONDYLAR. SERIAL NUMBER (B)(6) IS CONFIRMED TO HAVE BEEN PACKAGED IN A VACUUM BAG THAT DOES NOT CONTAIN EVOH. 510K: K954208. CONCOMITANTS: OPTETRAK FEMORAL CONSTRAINED CONDYLAR CEMENTED (CAT. NO. UNKNOWN, SN (B)(6)). EXTENSION, FEMORAL FLUTED STEM WITH SLOT (CAT. NO. UNKNOWN, SN (B)(6)). AUGMENT, FEMORAL POSTERIOR (CAT. NO. UNKNOWN, SN (B)(6)). AUGMENT, FEMORAL DISTAL (CAT. NO. UNKNOWN, SN (B)(6)). AUGMENT, FEMORAL DISTAL (CAT. NO. UNKNOWN, SN (B)(6)). ADAPTER, FEMORAL TAPER WITH SCREW (CAT. NO. UNKNOWN, SN (B)(6)). STRYKER SIMPLEX WITH TOBRA BONE CEMENT. ORIGINAL DESCRIPTION SUMMARY: LEGAL CASE - (B)(6). PATIENT ID: (B)(6) LK REV2. REFER TO CASE- (B)(4) LK REV1. AS REPORTED VIA LEGAL DOCUMENTATION, A PATIENT HAD THEIR FIRST LEFT KNEE REVISION SURGERY ON (B)(6) 2011. THEY SUBSEQUENTLY UNDERWE

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant; Joint Laxity; Pain; Scar Tissue; Synovitis

    MAUDE report 21246512 ↗
  • InjuryJan 7, 2025Unstable; Loosening of Implant Not Related to Bone-Ingrowth; Unstable; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED BY LEGAL DOCUMENTATION, THAT APPROXIMATELY 9 YEARS AND 4 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT EXPERIENCED PROSTHESIS WEAR, INSTABILITY, FEMORAL LOOSENING, CYST(S), JOINT EFFUSION, OSTEOLYSIS, SWELLING/ EDEMA, JOINT LAXITY AND IMPLANT PAIN. REVISION OPERATIVE REPORT- THE PATIENT PRESENTED WITH A PAINFUL CONSTRAINED RIGHT KNEE REPLACEMENT. THE PATIENT WAS NOTED TO HAVE MARKED WEAR OF THE POLYETHYLENE ON X-RAY, BUT NOT GROSS EVIDENCE OF LOOSENING. HE HAD A LARGE EFFUSION AND POPLITEAL CYST THAT WAS PAINFUL. THE PATIENT HAS FAILED CONSERVATIVE MANAGEMENT INCLUDING ACTIVITY MODIFICATION AND ANTI-INFLAMMATORY MEDICATIONS. THERE WAS VERY EXTENSIVE SYNOVITIS AND A VERY THOROUGH SYNOVECTOMY OF THE JOINT WAS PERFORMED. THE POLYETHYLENE LINER WAS REMOVED. THERE WAS NOTED TO BE SEVERE WEAR OF THE LINER WITH DELAMINATION. NEXT THE FEMORAL COMPONENT WAS NOTED TO BE LOOSE AND THERE WAS A FRACTURE OF THE STEM AT THE FEMORAL COMPONENT JUNCTION. THE FEMORAL COMPONENT AND STEM WERE REMOVED. THERE WAS SIGNIFICANT FIBROUS TISSUE BENEATH IT WITH MARKED OSTEOLYSIS. THERE WAS NO EVIDENCE OF INFECTION GROSSLY ALTHOUGH MULTIPLE SPECIMENS WERE SENT TO PATHOLOGY. ATTE

    Patient problems: Cyst(s); Implant Pain; Joint Laxity; Joint effusion; Osteolysis; Swelling/ Edema

    MAUDE report 21095535 ↗
  • InjuryOct 25, 2024Unstable; Loosening of Implant Not Related to Bone-Ingrowth; Unstable; Loosening of Implant Not Related to Bone-Ingrowth

    LEGAL CASE - (B)(6). IT WAS REPORTED THAT A 65 YO PATIENT, INITIAL LEFT KNEE IMPLANTED ON (B)(6) 2014, UNDERWENT A REVISION PROCEDURE ON (B)(6) 2022, APPROXIMATELY 8 YEARS POST THE INITIAL PROCEDURE. THE PATIENT HAD BEEN SEEN BY THE SURGEON IN AN OUTPATIENT CLINIC. THE PATIENT HAD CONSTANT PAIN AND INSTABILITY THAT FAILED CONSERVATIVE TREATMENT. THE PRE-PROCEDURE DIAGNOSIS INDICATED ASEPTIC LOOSENING OF COMPONENTS, ARTHROPLASTY WEAR OF THE POLYETHYLENE LINER. THE POLYETHYLENE WAS REMOVED WITH AN OSTEOTOME WITHOUT DIFFICULTY. THE FEMORAL COMPONENT WAS NOTED TO BE GROSSLY LOOSE AND WAS DIS-IMPACTED WITHOUT DIFFICULTY. THE TIBIAL COMPONENT WAS NOTED TO BE GROSSLY LOOSE AND WAS DIS-IMPACTED OFF THE TIBIAL PLATEAU WITHOUT DIFFICULTY AND THERE WAS MINIMAL BONE LOSS. THE TIBIAL STEM AND BONE CEMENT WERE REMOVED. COMPETITOR¿S DEVICES WERE RE-IMPLANTED. THE PATIENT WAS TRANSFERRED TO THE POST ANESTHESIA CARE UNIT IN STABLE CONDITION WITH STABLE VITAL SIGNS HAVING TOLERATED THE PROCEDURE WELL WITHOUT APPARENT COMPLICATIONS. REVISED TO STRYKER. SERIAL #: (B)(6), CATEGORY #: 208-25-13 - CC TIBIAL INSERT SZ 5, 13MM, 510K: K954208 JWH Z-0019-2022 CONCOMITANTS: SERIAL #: (B)(6), CATEGORY #: 204-0

    Patient problems: Joint Laxity; Pain

    MAUDE report 20532451 ↗
  • InjurySep 19, 2023Insufficient Device Problem Information; Loosening of Implant Not Related to Bone-Ingrowth; Insufficient Device Problem Information; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A RIGHT KNEE REVISION ON (B)(6) 2015. APPROXIMATELY 6 YEARS AND 11 MONTHS AFTER THE FIRST REVISION THE PATIENT HAD A SECOND RIGHT KNEE REVISION ON (B)(6) 2022. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE PATIENT HAS FILED A SHORT-FORM COMPLAINT IN A MULTIDISTRICT LITIGATION TITLED IN RE: EXACTECH POLYETHYLENE ORTHOPEDIC PRODUCTS LIABILITY LITIGATION, MDL NO. 3044, AND PENDING IN THE EASTERN DISTRICT OF (B)(6). PER THE TRANSFER ORDER CREATING THIS MULTIDISTRICT LITIGATION, ¿PLAINTIFFS¿ALLEGE THAT THEIR KNEE OR HIP REPLACEMENT DEVICES FAILED PREMATURELY BECAUSE OF DEGRADATION OF THE DEVICE¿S POLYETHYLENE COMPONENT, WHICH RESULTED IN THE PREMATURE REMOVAL (OR PLANNED REMOVAL) OF THE PROSTHESIS AT ISSUE.¿ BECAUSE THE PATIENT HAS FILED A SUIT IN THIS MULTIDISTRICT LITIGATION, THE PATIENT APPEARS TO ALLEGE THAT THE PATIENT WAS INJURED AS A RESULT OF PREMATURE WEAR OF AN EXACTECH POLYETHYLENE DEVICE.

    Patient problems: Bone Fracture(s); Insufficient Information; Osteolysis; Pain

    MAUDE report 17772195 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K954208. One submission often covers many models, so check the product and lot details in each record.

  • Z-0582-2024 · Nov 21, 2023 · Open, Classified
    Exactech Optetrak Knee System, Catalog Numbers: a) 200-02-26, b) 200-02-29, c) 200-02-32, d) 200-02-35, e) 200-02-38, f) 200-02-41, g) 200-03-32, h) 200-07-29, i) 200-07-

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates