Optetrak 208-04-43Recalls & reported problems
Exactech, Inc. · Class 2 · Implant
All models of Optetrak (336)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885862053411), received between Jul 31, 2023 and Jul 31, 2023 — classified as 1 injury.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).
34 other recalls cover devices cleared under the same FDA submission (K023186); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth1
- Mechanical Problem1
- Migration1
Patient problems
- Ambulation or Postural Difficulties1
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Aseptic loosening1
- Implant Pain1
- Loss of Range of Motion1
- Osteolysis1
- Pain1
- Swelling/ Edema1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJul 31, 2023Mechanical Problem; Loosening of Implant Not Related to Bone-Ingrowth; Migration; Mechanical Problem; Loosening of Implant Not Related to Bone-Ingrowth; Migration
AS REPORTED VIA LEGAL DOCUMENTATION THE MALE PATIENT HAD A RIGHT KNEE REPLACEMENT ON (B)(6) 2009. APPROXIMATELY 13 YEARS AND 7 MONTHS AFTER THE INITIAL PROCEDURE THE PATIENT HAD A RIGHT KNEE REVISION ON (B)(6) 2023. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. REVISION OP REPORT ON (B)(6) 2023; DIAGNOSIS: FAILED RIGHT KNEE ARTHROPLASTY SECONDARY TO TIBIAL LOOSENING AND MASSIVE OSTEOLYSIS. FINDINGS CONSISTENT WITH POLYETHYLENE-INDUCED SYNOVITIS. A SYNOVECTOMY WAS PERFORMED. IT WAS CLEAR THAT THE TIBIA HAD FAILED INTO VARUS WITH SIGNIFICANT MEDIAL BONE LOSS. THERE WAS DELAMINATION OF THE TIBIAL POLYETHYLENE. THERE WAS AN AREA OF MASSIVE OSTEOLYSIS UNDER THE ANTERIOR TROCHLEA THAT HAD ALREADY BEEN APPRECIATED BY THE CT SCAN. THERE WAS ALSO SIGNIFICANT OSTEOLYSIS OF THE MEDIAL FEMORAL CONDYLE. THE PATIENT WAS TRANSFERRED TO THE PACU IN STABLE CONDITION WITHOUT COMPLICATION.
Patient problems: Ambulation or Postural Difficulties; Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Aseptic loosening; Implant Pain; Loss of Range of Motion; Osteolysis; Pain; Swelling/ Edema; Synovitis
MAUDE report 17423128 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K023186. One submission often covers many models, so check the product and lot details in each record.
- Z-0130-2018 · Sep 19, 2017 · TerminatedOptetrak Offset Tibial Tray and Screws, Size 1DeltaF/1T, Beta, Ref 208-04-14, Sterile, RX.
- Z-0129-2018 · Sep 19, 2017 · TerminatedOptetrak Offset Tibial Tray and Screws, Size 1DeltaF/1T, Alpha, Ref 208-04-13, Sterile, RX.
- Z-0131-2018 · Sep 19, 2017 · TerminatedOptetrak Offset Tibial Tray and Screws, Size 1F/1T & 2F/1T, Alpha, Ref 208-04-21, Sterile, RX.
- Z-0127-2018 · Sep 19, 2017 · TerminatedOptetrak Offset Tibial Tray and Screws, Size 1DeltaF/0T, Alpha, Ref 208-04-11, Sterile, RX.
- Z-0128-2018 · Sep 19, 2017 · TerminatedOptetrak Offset Tibial Tray and Screws, Size 1DeltaF/0T, Beta, Ref 208-04-12, Sterile, RX.
- Z-0137-2018 · Sep 19, 2017 · TerminatedOptetrak Offset Tibial Tray and Screws, Size 3F/3T, Beta, Ref 208-04-34, Sterile, RX.
- Z-0154-2018 · Sep 19, 2017 · TerminatedOptetrak Constrained Condylar Offset Screw, Size KK, Ref 208-04-78, Sterile, RX.
- Z-0126-2018 · Sep 19, 2017 · TerminatedOptetrak Offset Tibial Tray and Screws, Size 0F/0T, Alpha, Ref 208-04-01, Sterile, RX.
- Z-0159-2018 · Sep 19, 2017 · TerminatedOptetrak Offset Tibial Tray and Screws, Size 3F/2T, Beta, Ref 208-04-32, Sterile, RX.
- Z-0148-2018 · Sep 19, 2017 · TerminatedOptetrak Constrained Condylar Offset Screw, Size DD, Ref 208-04-72, Sterile, RX.
- Z-0156-2018 · Sep 19, 2017 · TerminatedOptetrak Constrained Condylar Offset Screw, Size MM, Ref 208-04-80, Sterile, RX.
- Z-0134-2018 · Sep 19, 2017 · TerminatedOptetrak Offset Tibial Tray and Screws, Size 1F/2T & 2F/2T, Beta, Ref 208-04-24, Sterile, RX.
- Z-0132-2018 · Sep 19, 2017 · TerminatedOptetrak Offset Tibial Tray and Screws, Size 1F/1T & 2F/1T, Beta, Ref 208-04-22, Sterile, RX.
- Z-0135-2018 · Sep 19, 2017 · TerminatedOptetrak Offset Tibial Tray and Screws, Size 3F/2T, Alpha, Ref 208-04-31, Sterile, RX.
- Z-0142-2018 · Sep 19, 2017 · TerminatedOptetrak Offset Tibial Tray and Screws, Size 5F/4T, Alpha, Ref 208-04-51, Sterile, RX.
…and 19 more.
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page