Skip to content

Optetrak 208-04-44Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak (336)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885862053428), received between Dec 30, 2024 and Dec 30, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).

34 other recalls cover devices cleared under the same FDA submission (K023186); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Bone Fracture(s)1
  • Pain1
  • Synovitis1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 30, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT APPROXIMATELY 81 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, LOOSENING, PAIN, SYNOVITIS, BONE FRACTURE(S). REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Bone Fracture(s); Pain; Synovitis

    MAUDE report 21035801 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K023186. One submission often covers many models, so check the product and lot details in each record.

  • Z-0131-2018 · Sep 19, 2017 · Terminated
    Optetrak Offset Tibial Tray and Screws, Size 1F/1T & 2F/1T, Alpha, Ref 208-04-21, Sterile, RX.
  • Z-0136-2018 · Sep 19, 2017 · Terminated
    Optetrak Offset Tibial Tray and Screws, Size 3F/3T, Alpha, Ref 208-04-33, Sterile, RX.
  • Z-0145-2018 · Sep 19, 2017 · Terminated
    Optetrak Offset Tibial Tray and Screws, Size 5F/5T, Beta, Ref 208-04-54, Sterile, RX.
  • Z-0157-2018 · Sep 19, 2017 · Terminated
    Optetrak Constrained Condylar Offset Screw, Size PP, Ref 208-04-81, Sterile, RX.
  • Z-0156-2018 · Sep 19, 2017 · Terminated
    Optetrak Constrained Condylar Offset Screw, Size MM, Ref 208-04-80, Sterile, RX.
  • Z-0128-2018 · Sep 19, 2017 · Terminated
    Optetrak Offset Tibial Tray and Screws, Size 1DeltaF/0T, Beta, Ref 208-04-12, Sterile, RX.
  • Z-0150-2018 · Sep 19, 2017 · Terminated
    Optetrak Constrained Condylar Offset Screw, Size FF, Ref 208-04-74, Sterile, RX.
  • Z-0154-2018 · Sep 19, 2017 · Terminated
    Optetrak Constrained Condylar Offset Screw, Size KK, Ref 208-04-78, Sterile, RX.
  • Z-0152-2018 · Sep 19, 2017 · Terminated
    Optetrak Constrained Condylar Offset Screw, Size HH, Ref 208-04-76, Sterile, RX.
  • Z-0132-2018 · Sep 19, 2017 · Terminated
    Optetrak Offset Tibial Tray and Screws, Size 1F/1T & 2F/1T, Beta, Ref 208-04-22, Sterile, RX.
  • Z-0129-2018 · Sep 19, 2017 · Terminated
    Optetrak Offset Tibial Tray and Screws, Size 1DeltaF/1T, Alpha, Ref 208-04-13, Sterile, RX.
  • Z-0135-2018 · Sep 19, 2017 · Terminated
    Optetrak Offset Tibial Tray and Screws, Size 3F/2T, Alpha, Ref 208-04-31, Sterile, RX.
  • Z-0142-2018 · Sep 19, 2017 · Terminated
    Optetrak Offset Tibial Tray and Screws, Size 5F/4T, Alpha, Ref 208-04-51, Sterile, RX.
  • Z-0126-2018 · Sep 19, 2017 · Terminated
    Optetrak Offset Tibial Tray and Screws, Size 0F/0T, Alpha, Ref 208-04-01, Sterile, RX.
  • Z-0159-2018 · Sep 19, 2017 · Terminated
    Optetrak Offset Tibial Tray and Screws, Size 3F/2T, Beta, Ref 208-04-32, Sterile, RX.

…and 19 more.

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page