Skip to content

Optetrak 208-22-22Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak (336)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10885862054210), received between Nov 22, 2021 and Sep 20, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Naturally Worn” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K954208); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2021Class IIOpen, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0019-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn2
  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Device Problem Information1

Patient problems

  • Ambulation or Postural Difficulties1
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Discomfort1
  • Emotional Changes1
  • Failure of Implant1
  • Insufficient Information1
  • Loss of Range of Motion1
  • Osteolysis1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjurySep 20, 2024Naturally Worn; Naturally Worn

    LEGAL CASE ¿ USA (MDL NO. 3044). CASE-(B)(4) RK REV 1 IS ASSOCIATED WITH THIS CASE. IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 73 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, SIGNIFICANT PAIN AND DISCOMFORT, GAIT IMPAIRMENT, POOR BALANCE, DIFFICULTY WALKING, COMPONENT PART LOOSENING, SOFT TISSUE DAMAGE, BONE LOSS, MENTAL AND EMOTIONAL DISTRESS. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE SERIAL NUMBER (B)(6) IS CONFIRMED TO BE A PART OF RECALL NUMBER: Z-0019-2022.

    Patient problems: Ambulation or Postural Difficulties; Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Discomfort; Emotional Changes; Loss of Range of Motion; Osteolysis; Pain

    MAUDE report 20271864 ↗
  • InjuryDec 21, 2023Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A RIGHT KNEE REVISION ON (B)(6) 2014. THIS DEVICE HAS NOT BEEN EXPLANTED. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE PATIENT HAS FILED A SHORT-FORM COMPLAINT IN A COORDINATED ACTION IN (B)(6) WITH MASTER CASE NO.(B)(4). THE CONSOLIDATED LONG FORM COMPLAINT THAT APPLIES TO CASES FILED IN THIS COORDINATED ACTION ALLEGES THAT PATIENTS FILING SUITS IN THIS COORDINATED ACTION WERE REQUIRED ¿TO UNDERGO REVISION SURGERIES DUE TO SEVERE, PAIN, SWELLING, AND INSTABILITY¿ DUE TO ¿WEAR OF THE POLYETHYLENE COMPONENTS AND RESULTING COMPONENT LOOSENING AND/OR OTHER FAILURE FAILURES CAUSING SERIOUS COMPLICATIONS INCLUDING TISSUE DAMAGE, OSTEOLYSIS, PERMANENT BONE LOSS, AND OTHER INJURIES.¿ BECAUSE THE PATIENT HAS FILED A SHORT-FORM COMPLAINT IN THIS COORDINATED ACTION, THE PATIENT APPEARS TO ALLEGE THAT THE PATIENT WAS INJURED AS A RESULT OF WEAR OF AN EXACTECH POLYETHYLENE DEVICE.

    Patient problems: Insufficient Information

    MAUDE report 18373355 ↗
  • InjuryNov 22, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED, A POLY SWAP WAS SCHEDULED FOR (B)(6) 2021. ALL NEEDED MATERIALS (IMPLANTS AND INSTRUMENTS) WERE PREPARED THE DAY BEFORE. THE IMPLANT RECORD WAS OBTAINED. THE SURGEON WAS TOLD HIS ONLY OPTIONS WERE REPLACING THE 22MM INSERT THAT WAS IN THERE OR INCREASING TO A 26MM OR 30MM. EARLY IN THE DAY ON THE DATE OF SURGERY WE (REP AND SURGEON'S STAFF) LOOKED AT THE PICS OF THE CASE AND INFORMED THE SURGEONS OF OUR CONCERNS. THE FEMUR NEEDED TO BE REVISED AND DUE TO THE SIZES OF EXISTING IMPLANTS OUR SYSTEMS PARAMETERS MAY NO COVER THE GAP AND A TOTAL FEMUR MAY BE NEEDED. THE SURGEON CANCELED THE PROCEDURE ON THE LEFT KNEE AND PROCEEDED TO SCOPE THE RIGHT KNEE. PATIENT WAS SCHEDULED FOR A POLY SWAP. THAT WAS DEEMED INADEQUATE IN A GROUP DISCUSSION. THE PROCEDURE WAS CANCELED AND REFERRED TO A REVISION SPECIALIST. PATIENT CONTINUED ON WITH SCOPE OF RIGHT KNEE BUT LEFT KNEE REVISION WAS CANCELLED AND REFERRED TO ANOTHER HSS SURGEON. THE WERE NO DEVICES REMOVED TO BE RETURNED FOR ANALYSIS. ADDITIONAL INFORMATION RECEIVED INDICATES THAT LEFT KNEE REVISION WAS SCHEDULED DUE TO MASSIVE INSTABILITY AND PAIN.

    Patient problems: Failure of Implant

    MAUDE report 12856389 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K954208. One submission often covers many models, so check the product and lot details in each record.

  • Z-0582-2024 · Nov 21, 2023 · Open, Classified
    Exactech Optetrak Knee System, Catalog Numbers: a) 200-02-26, b) 200-02-29, c) 200-02-32, d) 200-02-35, e) 200-02-38, f) 200-02-41, g) 200-03-32, h) 200-07-29, i) 200-07-

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →