Adverse event reports
11
Deaths reported
0
Injuries reported
11
Recalls naming it
2

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10885862054289), received between Sep 7, 2022 and Mar 27, 2025 — classified as 11 injuries.

The most frequently reported specific device problem is “Naturally Worn” (10 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 6 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 21, 2023Class IIOpen, Classified

    Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag.

    Z-0582-2024 · Root cause: Packaging change control · Recall details · FDA record ↗

  2. Aug 30, 2021Class IIOpen, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0019-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn10
  • Loosening of Implant Not Related to Bone-Ingrowth5
  • Unstable2
  • Adverse Event Without Identified Device or Use Problem1
  • Fracture1
  • Insufficient Device Problem Information1
  • Lack of Effect1

Patient problems

  • Osteolysis7
  • Pain7
  • Implant Pain6
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available5
  • Joint Laxity5
  • Synovitis5
  • Failure of Implant3
  • Ambulation or Postural Difficulties2

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 27, 2025Unstable; Naturally Worn; Unstable; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 48 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, AMBULATION DIFFICULTIES, BALANCE PROBLEMS, DISCOMFORT, OSTEOLYSIS, AND PAIN. INITIAL SURGERY AND REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Ambulation or Postural Difficulties; Balance Problems; Discomfort; Osteolysis; Pain

    MAUDE report 21710983 ↗
  • InjuryJun 26, 2024Naturally Worn; Lack of Effect; Naturally Worn; Lack of Effect

    IT WAS REPORTED THAT APPROXIMATELY 16 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PAIN. INITIAL SURGERY AND REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Implant Pain; Pain; Scar Tissue; Synovitis

    MAUDE report 19612907 ↗
  • InjuryJun 20, 2024Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT APPROXIMATELY 175 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PAIN, GROSSLY LOOSE FEMUR, TIBIAL INSERT, TIBIA, PATELLA, BROKEN POLY LOCKING SCREW, MASSIVE BONE LOSS IN THE FEMUR AND TIBIA, MARKED BONE LOSS IN THE PATELLA, ADVANCED WEAR AND OXIDATION ON TIBIAL INSERT AND PATELLA. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Implant Pain; Joint Laxity; Osteolysis; Pain; Synovitis

    MAUDE report 19574248 ↗
  • InjuryJun 20, 2024Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT APPROXIMATELY 175 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PAIN, GROSSLY LOOSE FEMUR, TIBIAL INSERT, TIBIA, PATELLA, BROKEN POLY LOCKING SCREW, MASSIVE BONE LOSS IN THE FEMUR AND TIBIA, MARKED BONE LOSS IN THE PATELLA, ADVANCED WEAR AND OXIDATION ON TIBIAL INSERT AND PATELLA. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Implant Pain; Joint Laxity; Osteolysis; Pain; Synovitis

    MAUDE report 19574221 ↗
  • InjuryJun 20, 2024Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT APPROXIMATELY 175 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PAIN, GROSSLY LOOSE FEMUR, TIBIAL INSERT, TIBIA, PATELLA, BROKEN POLY LOCKING SCREW, MASSIVE BONE LOSS IN THE FEMUR AND TIBIA, MARKED BONE LOSS IN THE PATELLA, ADVANCED WEAR AND OXIDATION ON TIBIAL INSERT AND PATELLA. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Implant Pain; Joint Laxity; Osteolysis; Pain; Synovitis

    MAUDE report 19574164 ↗
  • InjuryJun 20, 2024Fracture; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth; Fracture; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT APPROXIMATELY 175 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PAIN, GROSSLY LOOSE FEMUR, TIBIAL INSERT, TIBIA, PATELLA, BROKEN POLY LOCKING SCREW, MASSIVE BONE LOSS IN THE FEMUR AND TIBIA, MARKED BONE LOSS IN THE PATELLA, ADVANCED WEAR AND OXIDATION ON TIBIAL INSERT AND PATELLA. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Implant Pain; Joint Laxity; Osteolysis; Pain; Synovitis

    MAUDE report 19574001 ↗
  • InjuryJun 5, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 60 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PAIN, DISCOMFORT, AND DIFFICULTY WALKING. PLAINTIFF HAS SUFFERED, AND MAY CONTINUE TO SUFFER, SEVERE AND PERMANENT PERSONAL INJURIES, INCLUDING POLYETHYLENE WEAR AND DEVICE FAILURE, PAINFUL KNEE REVISION SURGERY TO REMOVE OR REVISE THE DEFECTIVE DEVICE, CONTINUED REHABILITATION, MEDICAL CARE, AND POSSIBLE ADDITIONAL SURGERIES. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Ambulation or Postural Difficulties; Discomfort; Failure of Implant; Implant Pain

    MAUDE report 19466379 ↗
  • InjuryMay 17, 2024Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT APPROXIMATELY 48 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS LOOSENING OF THE PATELLAR COMPONENT. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. DURING THE PROCEDURE, THE FEMORAL AND TIBIAL COMPONENTS WERE EXAMINED AND FOUND TO BE WELL FIXED WITH NO EVIDENCE OF LOOSENING. THERE WAS NOT ANY NOTE OF SIGNIFICANT SYNOVITIS OR SIGNS OF A LARGE EFFUSION ASSOCIATED WITH THE EXACTECH RECALL AND POLYETHYLENE PROBLEMS. THE DECISION WAS MADE TO LEAVE THE FEMORAL AND WELL FIXED TIBIAL COMPONENTS INTACT. THE PATIENT WAS REVISED WITH A COMPETITOR'S PATELLAR COMPONENT. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Osteolysis

    MAUDE report 19344107 ↗
  • InjuryMar 14, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT A PATIENT HAD A LEFT TOTAL KNEE ARTHROPLASTY FIRST REVISION ON (B)(6) 2015, AND THEN EXPERIENCED SECOND REVISION SURGICAL PROCEDURE ON (B)(6) 2017 APPROXIMATELY 1 YEAR AND 8 MONTHS AFTER FIRST REVISION IMPLANT. NO IMAGES WERE PROVIDED. THERE IS NO OTHER INFORMATION AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 18901178 ↗
  • InjuryDec 19, 2023Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A FOURTH RIGHT KNEE REVISION ON (B)(6) 2013. THIS DEVICE HAS NOT BEEN EXPLANTED. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE PATIENT HAS FILED A SHORT-FORM COMPLAINT IN A COORDINATED ACTION IN (B)(6) WITH MASTER CASE NO. (B)(4). THE CONSOLIDATED LONG FORM COMPLAINT THAT APPLIES TO CASES FILED IN THIS COORDINATED ACTION ALLEGES THAT PATIENTS FILING SUITS IN THIS COORDINATED ACTION WERE REQUIRED ¿TO UNDERGO REVISION SURGERIES DUE TO SEVERE, PAIN, SWELLING, AND INSTABILITY¿ DUE TO ¿WEAR OF THE POLYETHYLENE COMPONENTS AND RESULTING COMPONENT LOOSENING AND/OR OTHER FAILURE FAILURES CAUSING SERIOUS COMPLICATIONS INCLUDING TISSUE DAMAGE, OSTEOLYSIS, PERMANENT BONE LOSS, AND OTHER INJURIES.¿ BECAUSE THE PATIENT HAS FILED A SHORT-FORM COMPLAINT IN THIS COORDINATED ACTION, THE PATIENT APPEARS TO ALLEGE THAT THE PATIENT WAS INJURED AS A RESULT OF WEAR OF AN EXACTECH POLYETHYLENE DEVICE.

    Patient problems: Insufficient Information

    MAUDE report 18354762 ↗
  • InjurySep 7, 2022Unstable; Naturally Worn; Unstable; Naturally Worn

    IT WAS REPORTED VIA A LEGAL NOTIFICATION THAT MALE PATIENT, INITIALLY IMPLANTED WITH OPTETRAK KNEE REPLACEMENT SYSTEM IN 2011, EXACT DATE UNKNOWN, FOLLOWING SEEING NEWS THAT HIS KNEE REPLACEMENT WAS RECALLED, OBTAINED MRI¿S AND X-RAYS OF HIS KNEE WHICH CONFIRMED THAT THE RECALLED POLYETHYLENE INSERT HAD DEGRADED TO SUCH EXTENT THAT IT WAS CAUSING EXTENSIVE BONE DAMAGE. HE UNDERWENT REVISION SURGERY ON (B)(6) 2022 APPROXIMATELY 11 YEARS 6 MONTHS POST THE INITIAL PROCEDURE. THE KNEE REPLACEMENT SYSTEM WAS REMOVED, AND BONE RECONSTRUCTION WAS PERFORMED. NO DEVICES WILL BE RETURNING AS THE CUSTOMER RETAINS ALL DEVICES.

    Patient problems: Joint Laxity; No Clinical Signs, Symptoms or Conditions; Osteolysis; Pain; Swelling/ Edema

    MAUDE report 15374494 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates