Optetrak 208-24-13Recalls & reported problems
Exactech, Inc. · Class 2 · Implant
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10885862054340), received between Dec 21, 2021 and Dec 22, 2025 — classified as 6 injuries.
The most frequently reported specific device problem is “Naturally Worn” (5 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Nov 21, 2023Open, Classified
Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag.
Z-0582-2024 · Root cause: Packaging change control · FDA record ↗
- Aug 30, 2021Open, Classified
Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.
Z-0019-2022 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Naturally Worn5
- Adverse Event Without Identified Device or Use Problem1
- Insufficient Device Problem Information1
Patient problems
- Failure of Implant3
- Insufficient Information2
- Pain2
- Synovitis2
- Ambulation or Postural Difficulties1
- Bone Fracture(s)1
- Scar Tissue1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryDec 22, 2025Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL TOTAL KNEE ARTHROPLASTY WITH A COMPETITOR'S DEVICE. SUBSEQUENTLY, THE PATIENT DEVELOPED AN INFECTION AND WAS REVISED TO EXACTECH DEVICE. AFTER BEING CONVERTED TO EXACTECH DEVICES, THE PATIENT THEN DEVELOPED PAIN AND POLY WEAR RESULTING IN ANOTHER REVISION. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Pain
MAUDE report 23866650 ↗InjuryJun 24, 2024Naturally Worn; Naturally Worn
IT WAS REPORTED THAT APPROXIMATELY 48 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Ambulation or Postural Difficulties; Scar Tissue; Synovitis
MAUDE report 19595038 ↗InjurySep 26, 2023Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information
AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A RIGHT KNEE REVISION OF (B)(6) 2016. APPROXIMATELY 6 YEARS AND 9 MONTHS AFTER THE FIRST REVISION THE PATIENT HAD A SECOND RIGHT KNEE REVISION ON (B)(6) 2023. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE PATIENT HAS FILED A SHORT-FORM COMPLAINT IN A MULTIDISTRICT LITIGATION TITLED IN RE: EXACTECH POLYETHYLENE ORTHOPEDIC PRODUCTS LIABILITY LITIGATION, MDL NO. 3044, AND PENDING IN THE EASTERN DISTRICT OF NEW YORK. PER THE TRANSFER ORDER CREATING THIS MULTIDISTRICT LITIGATION, ¿PLAINTIFFS¿ALLEGE THAT THEIR KNEE OR HIP REPLACEMENT DEVICES¿FAILED PREMATURELY BECAUSE OF DEGRADATION OF THE DEVICE¿S POLYETHYLENE COMPONENT, WHICH RESULTED IN THE PREMATURE REMOVAL (OR PLANNED REMOVAL) OF THE PROSTHESIS AT ISSUE.¿ BECAUSE THE PATIENT HAS FILED A SUIT IN THIS MULTIDISTRICT LITIGATION, THE PATIENT APPEARS TO ALLEGE THAT THE PATIENT WAS INJURED AS A RESULT OF PREMATURE WEAR OF AN EXACTECH POLYETHYLENE DEVICE.
Patient problems: Bone Fracture(s); Insufficient Information; Pain; Synovitis
MAUDE report 17815843 ↗InjuryMay 21, 2023Naturally Worn; Naturally Worn
DURING THE WEEK OF APRIL 17-25, REPRESENTATIVES OF EXACTECH, INC. AND EXACTECH SPAIN CONDUCTED AN IN PERSON SITE VISIT WITH (B)(6) HOSPITAL. AS A RESULT OF THAT SITE VISIT AND AT EXACTECH'S REQUEST, THE HOSPITAL PROVIDED A RETROSPECTIVE EXCEL LISTING OF REVISION RELATED TO POLYETHYLENE WEAR THAT HAVE BEEN PERFORMED SINCE 2010. THESE CASES HAVE NOT PREVIOUSLY BEEN REPORTED TO EXACTECH AS COMPLAINTS. THIS PATIENT (B)(6) WAS IMPLANTED WITH A 208-24-13-INSERTO TIBIAL CC 4,13MM, SERIAL NUMBER: (B)(6) ON (B)(6) 2021. THE INSERT WAS MANUFACTURED ON 21-AUG-2018 AND WAS PACKAGED IN A VACUUM BAG WITH NO EVOH. THE INSERT WAS MANUFACTURED ON 21-AUG-2018 AND WAS 2.5 YEARS SHELF AGE AT THE TIME OF IMPLANT. APPROXIMATELY TWO YEARS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION FOR WEAR OF THE POLYETHYLENE. NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME.
Patient problems: Failure of Implant; Insufficient Information
MAUDE report 16971213 ↗InjuryFeb 2, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED, APPROXIMATELY 2.52 YEARS POST LEFT SIDE TKA, THIS MALE PATIENT UNDERWENT REVISION DUE TO PERSISTENT STIFFNESS. NO ADDITIONAL INFORMATION IS AVAILABLE OR FORTHCOMING. THE DEVICE IS NOT AVAILABLE FOR EVALUATION. 510K: K954208.
Patient problems: Failure of Implant
MAUDE report 13428194 ↗InjuryDec 21, 2021Naturally Worn; Naturally Worn
DURING OFFICE VISIT, SURGEON DID A CONDYLAR CONSTRAINED LINER EXCHANGE ON A REVISION HE DONE IN 2003, ON (B)(6) 2021. THIS WAS A LINER EXCHANGE OF A SIZE 4/13 MM CONDYLAR CONSTRAINED INSERT. THE LINER HAD MEDIAL WEAR. ALSO THE SPINE STIFFENING SCREW HAD SHEARED OFF AT THE THREADED JUNCTION OF THE SCREW. SURGEON WAS UNABLE TO RETRIEVE THE REST OF THE SCREW THAT WAS LOCKED INTO THE STEM AND TIBIAL TRAY. THIS DID NOT AFFECT THE INSERTION OF A NEW CONDYLAR CONSTRAINED LINER, BUT WAS UNABLE TO PLACE A NEW SPINE STIFFENING SCREW IN DUE TO THE BROKEN PIECE STILL IN THERE. THERE WAS NO BREAKAGE TO DEVICE. THERE WAS NO SURGICAL DELAY OR PROLONGATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. THE DEVICE WILL BE RETURN.
Patient problems: Failure of Implant
MAUDE report 13057292 ↗
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.