Skip to content

Optetrak 208-24-13Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak (336)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
2

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10885862054340), received between Dec 21, 2021 and Dec 22, 2025 — classified as 6 injuries.

The most frequently reported specific device problem is “Naturally Worn” (5 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 21, 2023Class IIOpen, Classified

    Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag.

    Z-0582-2024 · Root cause: Packaging change control · Recall details · FDA record ↗

  2. Aug 30, 2021Class IIOpen, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0019-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn5
  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Device Problem Information1

Patient problems

  • Failure of Implant3
  • Insufficient Information2
  • Pain2
  • Synovitis2
  • Ambulation or Postural Difficulties1
  • Bone Fracture(s)1
  • Scar Tissue1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 22, 2025Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL TOTAL KNEE ARTHROPLASTY WITH A COMPETITOR'S DEVICE. SUBSEQUENTLY, THE PATIENT DEVELOPED AN INFECTION AND WAS REVISED TO EXACTECH DEVICE. AFTER BEING CONVERTED TO EXACTECH DEVICES, THE PATIENT THEN DEVELOPED PAIN AND POLY WEAR RESULTING IN ANOTHER REVISION. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Pain

    MAUDE report 23866650 ↗
  • InjuryJun 24, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 48 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Ambulation or Postural Difficulties; Scar Tissue; Synovitis

    MAUDE report 19595038 ↗
  • InjurySep 26, 2023Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A RIGHT KNEE REVISION OF (B)(6) 2016. APPROXIMATELY 6 YEARS AND 9 MONTHS AFTER THE FIRST REVISION THE PATIENT HAD A SECOND RIGHT KNEE REVISION ON (B)(6) 2023. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE PATIENT HAS FILED A SHORT-FORM COMPLAINT IN A MULTIDISTRICT LITIGATION TITLED IN RE: EXACTECH POLYETHYLENE ORTHOPEDIC PRODUCTS LIABILITY LITIGATION, MDL NO. 3044, AND PENDING IN THE EASTERN DISTRICT OF NEW YORK. PER THE TRANSFER ORDER CREATING THIS MULTIDISTRICT LITIGATION, ¿PLAINTIFFS¿ALLEGE THAT THEIR KNEE OR HIP REPLACEMENT DEVICES¿FAILED PREMATURELY BECAUSE OF DEGRADATION OF THE DEVICE¿S POLYETHYLENE COMPONENT, WHICH RESULTED IN THE PREMATURE REMOVAL (OR PLANNED REMOVAL) OF THE PROSTHESIS AT ISSUE.¿ BECAUSE THE PATIENT HAS FILED A SUIT IN THIS MULTIDISTRICT LITIGATION, THE PATIENT APPEARS TO ALLEGE THAT THE PATIENT WAS INJURED AS A RESULT OF PREMATURE WEAR OF AN EXACTECH POLYETHYLENE DEVICE.

    Patient problems: Bone Fracture(s); Insufficient Information; Pain; Synovitis

    MAUDE report 17815843 ↗
  • InjuryMay 21, 2023Naturally Worn; Naturally Worn

    DURING THE WEEK OF APRIL 17-25, REPRESENTATIVES OF EXACTECH, INC. AND EXACTECH SPAIN CONDUCTED AN IN PERSON SITE VISIT WITH (B)(6) HOSPITAL. AS A RESULT OF THAT SITE VISIT AND AT EXACTECH'S REQUEST, THE HOSPITAL PROVIDED A RETROSPECTIVE EXCEL LISTING OF REVISION RELATED TO POLYETHYLENE WEAR THAT HAVE BEEN PERFORMED SINCE 2010. THESE CASES HAVE NOT PREVIOUSLY BEEN REPORTED TO EXACTECH AS COMPLAINTS. THIS PATIENT (B)(6) WAS IMPLANTED WITH A 208-24-13-INSERTO TIBIAL CC 4,13MM, SERIAL NUMBER: (B)(6) ON (B)(6) 2021. THE INSERT WAS MANUFACTURED ON 21-AUG-2018 AND WAS PACKAGED IN A VACUUM BAG WITH NO EVOH. THE INSERT WAS MANUFACTURED ON 21-AUG-2018 AND WAS 2.5 YEARS SHELF AGE AT THE TIME OF IMPLANT. APPROXIMATELY TWO YEARS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION FOR WEAR OF THE POLYETHYLENE. NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 16971213 ↗
  • InjuryFeb 2, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED, APPROXIMATELY 2.52 YEARS POST LEFT SIDE TKA, THIS MALE PATIENT UNDERWENT REVISION DUE TO PERSISTENT STIFFNESS. NO ADDITIONAL INFORMATION IS AVAILABLE OR FORTHCOMING. THE DEVICE IS NOT AVAILABLE FOR EVALUATION. 510K: K954208.

    Patient problems: Failure of Implant

    MAUDE report 13428194 ↗
  • InjuryDec 21, 2021Naturally Worn; Naturally Worn

    DURING OFFICE VISIT, SURGEON DID A CONDYLAR CONSTRAINED LINER EXCHANGE ON A REVISION HE DONE IN 2003, ON (B)(6) 2021. THIS WAS A LINER EXCHANGE OF A SIZE 4/13 MM CONDYLAR CONSTRAINED INSERT. THE LINER HAD MEDIAL WEAR. ALSO THE SPINE STIFFENING SCREW HAD SHEARED OFF AT THE THREADED JUNCTION OF THE SCREW. SURGEON WAS UNABLE TO RETRIEVE THE REST OF THE SCREW THAT WAS LOCKED INTO THE STEM AND TIBIAL TRAY. THIS DID NOT AFFECT THE INSERTION OF A NEW CONDYLAR CONSTRAINED LINER, BUT WAS UNABLE TO PLACE A NEW SPINE STIFFENING SCREW IN DUE TO THE BROKEN PIECE STILL IN THERE. THERE WAS NO BREAKAGE TO DEVICE. THERE WAS NO SURGICAL DELAY OR PROLONGATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. THE DEVICE WILL BE RETURN.

    Patient problems: Failure of Implant

    MAUDE report 13057292 ↗

FDA clearance and classification

Cleared through 510(k) K954208 on Feb 26, 1996 (decision: Substantially Equivalent for Some Indications). Submitted by Exactech, Inc. as “Optetrak Constrained Condylar Knee For Cemented Use Only”. The FDA review took 172 days. FDA record ↗

The FDA regulates product code JWH (prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer) under 21 CFR 888.3560 as a Class II device. 21 CFR 888.3560 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Optetrak 208-24-13

Has the Optetrak 208-24-13 been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-0582-2024 (Class II), began on Nov 21, 2023: Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Optetrak 208-24-13?

The FDA's MAUDE database lists 6 adverse event reports for this device, received between Dec 21, 2021 and Dec 22, 2025. In those reports, the most common specific device problem is “Naturally Worn” (5); the most common patient problems are “Failure of Implant” (3), “Pain” (2) and “Synovitis” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Optetrak 208-24-13?

None of the reports are classified as deaths, 6 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Optetrak 208-24-13 FDA approved?

It was cleared through 510(k) K954208 on Feb 26, 1996. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the Optetrak 208-24-13 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the Optetrak 208-24-13, and is it still sold?

The FDA's device database lists Exactech, Inc. as the company responsible for it (brand Optetrak, model 208-24-13). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page