Skip to content

Optetrak 208-51-15Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak (336)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10885862054517), received between May 21, 2023 and Feb 22, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2021Class IIOpen, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0019-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Device Problem Information1
  • Naturally Worn1

Patient problems

  • Failure of Implant2
  • Insufficient Information2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 22, 2024Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    AS REPORTED, APPROXIMATELY TEN YEARS POST INITIAL TKA, THE PATIENT HAD A POLYETHYLENE EXCHANGED ON AN OPTETRAK CC KNEE. IMAGES AND X-RAYS RECEIVED. NO PATIENT INFORMATION PROVIDED.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 18759907 ↗
  • MalfunctionMay 21, 2023Naturally Worn; Naturally Worn

    DURING THE WEEK OF APRIL 17-25, REPRESENTATIVES OF EXACTECH, INC. AND EXACTECH SPAIN CONDUCTED AN IN-PERSON SITE VISIT WITH (B)(6) HOSPITAL. AS A RESULT OF THAT SITE VISIT AND AT EXACTECH¿S REQUEST, THE HOSPITAL PROVIDED A RETROSPECTIVE EXCEL LISTINGS OF REVISION RELATED TO POLYETHYLENE WEAR THAT HAVE BEEN PERFORMED SINCE 2010. THESE CASES HAVE NOT PREVIOUSLY BEEN REPORTED TO EXACTECH AS COMPLAINTS. THIS PATIENT (B)(6) WAS IMPLANTED WITH A 208-51-15- CC TIBIAL INSERT SZ 1D, 15MM, SERIAL NUMBER: (B)(6) 0N (B)(6) 2020. THE INSERT WAS MANUFACTURED ON 13-FEB-2018 AND WAS PACKAGED IN A VACUUM BAG WITH NO EVOH. THE INSERT WAS MANUFACTURED ON 13-FEB-2018 AND WAS 2.4 YEARS SHELF AGE AT THE TIME OF IMPLANT. APPROXIMATELY THREE YEARS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION FOR WEAR OF THE POLYETHYLENE. NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 16971350 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page