NA 209-30-00Recalls & reported problems
Exactech, Inc. · TORQUE LIMITING DRIVER · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10885862056191), received between Jul 20, 2020 and Apr 12, 2021 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Break” (8 reports).
5 other recalls cover devices cleared under the same FDA submission (K150890, K954208); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break8
Patient problems
- No Consequences Or Impact To Patient6
- Insufficient Information2
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 12, 2021Break; Break
DURING A REVISION FROM A PRIMARY TO A CC, WHILE LOCKING THE INSERT TO THE BASE PLATE, THE BALL TIP DRIVER BROKE, RESULTING IN MAJOR DIFFICULTIES TO RETRIEVE THE BROKEN TIP LEFT IN THE SCREW HEAD. ALL PIECES WERE RETRIEVED, WITH LESS THAN 5 MIN DELAY TO SURGERY. THE TIP WAS DISCARDED AND WILL NOT RETURN, THE DRIVER WILL BE RETURNING. NO PATIENT INFO PROVIDED.
Patient problems: Insufficient Information
MAUDE report 11654183 ↗MalfunctionApr 12, 2021Break; Break
DURING A REVISION FROM A PRIMARY TO A CC, WHILE LOCKING THE INSERT TO THE BASE PLATE, THE BALL TIP DRIVER BROKE, RESULTING IN MAJOR DIFFICULTIES TO RETRIEVE THE BROKEN TIP LEFT IN THE SCREW HEAD. ALL PIECES WERE RETRIEVED, WITH LESS THAN 5 MIN DELAY TO SURGERY. THE TIP WAS DISCARDED AND WILL NOT RETURN, THE DRIVER WILL BE RETURNING. NO PATIENT INFO PROVIDED.
Patient problems: Insufficient Information
MAUDE report 11654176 ↗MalfunctionDec 21, 2020Break; Break
DURING A REVISION FROM A PRIMARY TO A CC, WHILE LOCKING THE INSERT TO THE BASE PLATE, THE BALL TIP DRIVER BROKE, RESULTING IN MAJOR DIFFICULTIES TO RETRIEVE THE BROKEN TIP LEFT IN THE SCREW HEAD. ALL PIECES WERE RETRIEVED, WITH LESS THAN 5 MIN DELAY TO SURGERY. THE TIP WAS DISCARDED AND WILL NOT RETURN, THE DRIVER WILL BE RETURNING. NO PATIENT INFO PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11049979 ↗MalfunctionDec 21, 2020Break; Break
DURING A REVISION FROM A PRIMARY TO A CC, WHILE LOCKING THE INSERT TO THE BASE PLATE, THE BALL TIP DRIVER BROKE, RESULTING IN MAJOR DIFFICULTIES TO RETRIEVE THE BROKEN TIP LEFT IN THE SCREW HEAD. ALL PIECES WERE RETRIEVED, WITH LESS THAN 5 MIN DELAY TO SURGERY. THE TIP WAS DISCARDED, AND WILL NOT RETURN. THE DRIVER WILL BE RETURNING. NO PATIENT INFO. PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11049884 ↗MalfunctionDec 21, 2020Break; Break
DURING A REVISION FROM A PRIMARY TO A CC, WHILE LOCKING THE INSERT TO THE BASE PLATE, THE BALL TIP DRIVER BROKE, RESULTING IN MAJOR DIFFICULTIES TO RETRIEVE THE BROKEN TIP LEFT IN THE SCREW HEAD. ALL PIECES WERE RETRIEVED, WITH LESS THAN 5 MIN DELAY TO SURGERY. THE TIP WAS DISCARDED AND WILL NOT RETURN, THE DRIVER WILL BE RETURNING. NO PATIENT INFO PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11049668 ↗MalfunctionJul 20, 2020Break; Break
DURING A REVISION FROM A PRIMARY TO A CC, WHILE LOCKING THE INSERT TO THE BASE PLATE, THE BALL TIP DRIVER BROKE, RESULTING IN MAJOR DIFFICULTIES TO RETRIEVE THE BROKEN TIP LEFT IN THE SCREW HEAD. ALL PIECES WERE RETRIEVED, WITH LESS THAN 5 MIN DELAY TO SURGERY. THE TIP WAS DISCARDED AND WILL NOT RETURN, THE DRIVER WILL BE RETURNING. NO PATIENT INFO PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10297795 ↗MalfunctionJul 20, 2020Break; Break
DURING A REVISION FROM A PRIMARY TO A CC, WHILE LOCKING THE INSERT TO THE BASE PLATE, THE BALL TIP DRIVER BROKE, RESULTING IN MAJOR DIFFICULTIES TO RETRIEVE THE BROKEN TIP LEFT IN THE SCREW HEAD. ALL PIECES WERE RETRIEVED, WITH LESS THAN 5 MIN DELAY TO SURGERY. THE TIP WAS DISCARDED AND WILL NOT RETURN, THE DRIVER WILL BE RETURNING. NO PATIENT INFO PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10297640 ↗MalfunctionJul 20, 2020Break; Break
DURING A REVISION FROM A PRIMARY TO A CC. IT HAS HAPPENED WHILE LOCKING THE INSERT TO THE BASE PLATE, RESULTING IN MAJOR DIFFICULTIES TO RETRIEVE THE BROKEN TIP LEFT IN THE SCREW HEAD. ALL PIECES WERE RETRIEVED, WITH LESS THAN 5 MIN DELAY TO SURGERY. THE TIP WAS DISCARDED AND WILL NOT RETURN, THE DRIVER WILL BE RETURNING. NO PATIENT INFO PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10296710 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K150890, K954208. One submission often covers many models, so check the product and lot details in each record.
- Z-0583-2024 · Nov 21, 2023 · Open, ClassifiedExactech Optetrak Logic Knee System, Catalog Numbers: a) 02-012-47-3011, b) 02-012-47-5009, c) 02-012-47-5011, d) 02-012-65-3013
- Z-0582-2024 · Nov 21, 2023 · Open, ClassifiedExactech Optetrak Knee System, Catalog Numbers: a) 200-02-26, b) 200-02-29, c) 200-02-32, d) 200-02-35, e) 200-02-38, f) 200-02-41, g) 200-03-32, h) 200-07-29, i) 200-07-
- Z-1364-2020 · Jan 27, 2020 · Open, ClassifiedTruliant Splined Stem Extension, 18mm x 120mm - Product Usage: The devices are used in a cemented total knee prosthesis system, which provides femoral components, tibial inserts, s
- Z-3209-2018 · May 8, 2018 · Open, ClassifiedOptetrak Logic CC Stem Extension, 22mm x 160mm The OPTETRAK Logic CC Total Knee System is indicated for use in skeletally mature individuals undergoing primary surgery for total k
- Z-2704-2017 · Apr 7, 2017 · TerminatedExactech Optetrak Logic CC 15mm Tibial Augment Screws
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.