Skip to content

Optetrak 210-01-04Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak (336)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10885862057419), received between Feb 18, 2025 and Nov 17, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K954208); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth2
  • Mechanical Problem1
  • Unstable1

Patient problems

  • Pain2
  • Swelling/ Edema2
  • Synovitis2
  • Ambulation or Postural Difficulties1
  • Aseptic loosening1
  • Joint Laxity1
  • Joint effusion1
  • Loss of Range of Motion1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 17, 2025Mechanical Problem; Loosening of Implant Not Related to Bone-Ingrowth; Mechanical Problem; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A RIGHT TOTAL KNEE ARTHROPLASTY. SUBSEQUENTLY, IT WAS REPORTED THAT THE PATIENT EXPERIENCED PAIN, SWELLING, STIFFNESS, LOSS OF MOTION, WEAKNESS AND DIFFICULTY AMBULATING. AS A RESULT, THE PATIENT UNDERWENT A RIGHT KNEE REVISION APPROXIMATELY 13 YEARS AND 7 MONTHS AFTER INITIAL IMPLANTATION. PROVIDED OPERATIVE NOTES REPORT TIBIAL LOOSENING AND WEAR OF THE TIBIAL INSERT. IT WAS LATER REPORTED THAT THE FEMORAL COMPONENT WAS FOUND TO BE DEBONDED. THE PATIENT WAS LAST REPORTED TO BE STABLE. NO FURTHER PATIENT IMPACT REPORTED. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Ambulation or Postural Difficulties; Aseptic loosening; Loss of Range of Motion; Osteolysis; Pain; Swelling/ Edema; Synovitis

    MAUDE report 23575362 ↗
  • InjuryFeb 18, 2025Unstable; Loosening of Implant Not Related to Bone-Ingrowth; Unstable; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT APPROXIMATELY 87 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION SURGERY TO ADDRESS PROSTHESIS WEAR, EFFUSION AND INSTABILITY. A REVISION OPERATIVE REPORT WAS PROVIDED. POST OPERATIVE DIAGNOSIS NOTED INSTABILITY AND FEMORAL COMPONENT LOOSENING. INTRAOPERATIVELY THE SURGEON OBSERVED SIGNIFICANT EFFUSION AND SYNOVITIS AND DELAMINATION OF THE POLYETHYLENE COMPONENT. IT WAS NOTED THERE WAS NO EVIDENCE OF THE POLYETHYLENE TIBIAL COMPONENT LOOSENING. THE SURGEON OBSERVED THE TIBIA WAS WELL FIXED BUT THERE WAS DEBONDING OF THE FEMORAL COMPONENT AT THE IMPLANT CEMENT INTERFACE. NO MEDICAL OR SURGICAL INTERVENTIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Joint Laxity; Joint effusion; Pain; Swelling/ Edema; Synovitis

    MAUDE report 21401871 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K954208. One submission often covers many models, so check the product and lot details in each record.

  • Z-0582-2024 · Nov 21, 2023 · Open, Classified
    Exactech Optetrak Knee System, Catalog Numbers: a) 200-02-26, b) 200-02-29, c) 200-02-32, d) 200-02-35, e) 200-02-38, f) 200-02-41, g) 200-03-32, h) 200-07-29, i) 200-07-

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page