Skip to content

Optetrak 234-02-05Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak (336)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10885862063212), received between May 15, 2023 and Apr 17, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth2

Patient problems

  • Osteolysis2
  • Pain2
  • Synovitis2
  • Ambulation or Postural Difficulties1
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Aseptic loosening1
  • Cyst(s)1
  • Discomfort1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 17, 2026Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL TOTAL KNEE ARTHROPLASTY ON THE LEFT SIDE. SUBSEQUENTLY, THE PATIENT EXPERIENCED COMPLETE POLYETHYLENE LINER WEAR AND DELAMINATION LINER WITH RESULTING SYNOVITIS, CYST FORMATION, OSTEOLYSIS OF THE FEMUR AND TIBIA, AND ASEPTIC LOOSENING OF THE FEMORAL AND TIBIAL COMPONENTS. AS A RESULT, APPROXIMATELY SIX YEARS AND EIGHT MONTHS POST-SURGERY, THE PATIENT UNDERWENT A REVISION. NO FURTHER IMPACT TO THE PATIENT WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS IS REPORT 2 OF 3 FOR THIS EVENT/PATIENT.

    Patient problems: Ambulation or Postural Difficulties; Aseptic loosening; Cyst(s); Metal Related Pathology; Osteolysis; Pain; Synovitis; Unspecified Infection; Unspecified Musculoskeletal problem

    MAUDE report 24913483 ↗
  • InjuryMay 15, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED VIA LEGAL DOCUMENTATION, THIS PATIENT HAD LEFT KNEE REPLACEMENT SURGERY ON (B)(6)2009. HE HAD LEFT KNEE REVISION ON (B)(6)2023, APPROXIMATELY 13 YEARS 5 MONTHS AFTER THEIR INITIAL IMPLANTATION. HE HAS DEVELOPED PROGRESSIVE PAIN AND DEBILITATION RELATED TO HIS LEFT KNEE. HIS X-RAYS SHOWED WHAT APPEARED TO BE LOOSENING OF THE FEMORAL COMPONENT WITH DEBONDING OF THE ANTERIOR FLANGE WITH THE CEMENT MANTLE. THE PATELLAR COMPONENT WAS WELL FIXED HOWEVER WAS SEVERELY WORN AND THEREFORE, REPLACED. THE FEMORAL COMPONENT HAD SIGNIFICANT DEBONDING OF THE CEMENT PROSTHETIC INTERFACE UNDER THE ANTERIOR FLANGE AND WAS EASILY REMOVED. THERE ALSO APPEARED TO BE LOOSENING OF THE TIBIAL COMPONENT AS IT WAS REMOVED FROM ITS CEMENT MANTLE RELATIVELY EASILY. THE POLYETHYLENE LINER HAD MINIMAL WEAR EXCEPT FOR ON THE ANTERIOR PORTION OF THE TIBIAL SPINE. THERE WAS SYNOVIAL HYPERTROPHY CONSISTENT WITH COMPONENT LOOSENING.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Discomfort; Failure of Implant; Osteolysis; Pain; Synovitis

    MAUDE report 16933641 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates