Equinoxe 320-46-10Recalls & reported problems
Exactech, Inc. · Class 2 · Implant
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10885862086754), received between May 31, 2019 and Aug 22, 2024 — classified as 3 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
2 FDA recall records name this device's identifier, 2 still open.
11 other recalls cover devices cleared under the same FDA submission (K063569); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Mar 6, 2024Open, Classified
The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).
Z-1425-2024 · Root cause: Nonconforming Material/Component · FDA record ↗
- Nov 21, 2023Open, Classified
Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag.
Z-0579-2024 · Root cause: Packaging change control · FDA record ↗
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Detachment of Device or Device Component1
Patient problems
- Failure of Implant1
- Insufficient Information1
- Joint Dislocation1
- Joint Laxity1
- No Code Available1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryAug 22, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
APPROXIMATELY 4 YEAR(S), 3 MONTH(S) AND 19 DAY(S) POST-OPERATIVE OF A REVISED LEFT TSA, THE PATIENT PRESENTED WITH DISASSOCIATION OF POLYETHYLENE. PATIENT DESCRIBES INSTABILITY EVENT 2 WEEKS AGO. THE PATIENT UNDERWENT STANDARD REVERSE REVISION SURGERY AND THE OUTCOME OF THIS EVENT IS CONSIDERED RESOLVED. THE CLINICAL REPORT INDICATES THAT THIS EVENT IS POSSIBLY RELATED TO THE DEVICE(S) AND DEFINITELY NOT RELATED TO THE PROCEDURE. (D10) CONCOMITANT DEVICE(S): 320-01-46 - EQUINOXE REVERSE 46MM GLENOSPHERE: (B)(6), 320-15-05 - EQ REV LOCKING SCREW: (B)(6), 320-20-00 - EQ REVERSE TORQUE DEFINING SCREW KIT: (B)(6).
Patient problems: Insufficient Information; Joint Laxity
MAUDE report 20034198 ↗InjuryNov 8, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
AS REPORTED APPROXIMATELY 22 MONTHS POSTOP THE INITIAL LTSA THIS (B)(6) Y/O MALE PATIENT REACHED DOWN TO PULL UP A KICKSTAND ON HIS MOTORCYCLE AND FELT A POP. HE WAS REVISED TO THE FOLLOWING, DOWNSIZING FROM 46MM TO 42MM. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. DEVICE WILL NOT RETURN DUE TO FACILITY POLICY.
Patient problems: Joint Dislocation
MAUDE report 12770117 ↗InjuryMay 31, 2019Device Dislodged or Dislocated
PATIENT WAS WORKING OUT AND STRETCHING WHEN HE FELT SOMETHING HAPPEN IN HIS SHOULDER. WHEN DOCTOR WENT AND COMPLETED THE SURGERY, HE NOTICED THAT THE HUMERAL LINER HAD BECOME DISASSOCIATED FROM THE HUMERAL ADAPTER TRAY.
Patient problems: Failure of Implant; No Code Available
MAUDE report 8659489 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K063569. One submission often covers many models, so check the product and lot details in each record.
- Z-2561-2025 · Aug 15, 2025 · Open, ClassifiedExactech Equinoxe, Reverse Shoulder Humeral Liners, REF: 320-38-00,145-DEG PE 38MM HUM LINER +0/ 320-38-03,145-DEG PE 38MM HUM LINER +2.5/ 320-42-00,145-DEG PE 42MM HUM LINER +0/
- Z-1418-2024 · Mar 6, 2024 · Open, ClassifiedExactech Equinoxe REVERSE SHOULDER, 38mm Humeral Liner: a) +0mm, Item Number 320-38-00, b) +2.5mm, Item Number 320-38-03; Shoulder Arthroplasty
- Z-1424-2024 · Mar 6, 2024 · Open, ClassifiedExactech Equinoxe REVERSE SHOULDER,46mm Humeral Liner, a) +0mm, Item number 320-46-00, b) +2.5mm, Item Number 320-46-03; Shoulder Arthroplasty
- Z-1422-2024 · Mar 6, 2024 · Open, ClassifiedExactech Equinoxe REVERSE SHOULDER,42mm Humeral Liner, a) +0mm, Item number 320-42-00, b) +2.5mm, Item Number 320-42-03; Shoulder Arthroplasty
- Z-1419-2024 · Mar 6, 2024 · Open, ClassifiedExactech Equinoxe REVERSE SHOULDER,38mm Constrained Humeral Liner: a) +0mm, Item Number 320-38-10, b) +2.5mm, Item Number 320-38-13, Shoulder Arthroplasty
- Z-1423-2024 · Mar 6, 2024 · Open, ClassifiedExactech Equinoxe REVERSE SHOULDER,42mm Constrained Humeral Liner, a) +0mm, Item number 320-42-10, b) +2.5mm, Item Number 320-42-13; Shoulder Arthroplasty
- Z-3020-2020 · Aug 24, 2020 · TerminatedEquinoxe Reverse Shoulder Compression Screw/Locking Cap Kit 4.5 x 22 mm.
- Z-3019-2020 · Aug 24, 2020 · TerminatedEquinoxe Reverse Shoulder Compression Screw/Locking Cap Kit 4.5 x 18 mm.
- Z-0012-2019 · Aug 21, 2018 · Open, ClassifiedEquinoxe Reverse Drill Bit Kit, Catalog Number 321-20-00 Product Usage: Shoulder surgery
- Z-0900-2017 · Oct 25, 2016 · TerminatedEquinoxe Reverse Shoulder Fixed Angle Torque Defining Screw Kit
- Z-2117-2012 · Jun 15, 2012 · TerminatedEquinoxe Reverse Compression Screw Intended for Primary and Reverse Total Shoulder Arthroplasty.
Other Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented devices
- Equinoxe 320-38-00Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented3 recalls107 reports
- Equinoxe 320-42-00Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented3 recalls107 reports
- Equinoxe 320-42-03Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls87 reports
- Equinoxe 320-38-03Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls43 reports
- Equinoxe 320-42-13Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls41 reports
- Equinoxe 320-10-00Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented132 reports
- Equinoxe 320-36-00Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls27 reports
- Equinoxe 320-42-10Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls14 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.