Optetrak Logic 02-012-49-1509Recalls & reported problems
Exactech, Inc. · Class 2 · Implant
All models of Optetrak Logic (408)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10885862144409), received between Jun 19, 2020 and May 1, 2024 — classified as 3 injuries.
The most frequently reported specific device problem is “Naturally Worn” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K111400); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 30, 2021Class IIOpen, Classified
Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.
Z-0021-2022 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Naturally Worn3
Patient problems
- Failure of Implant2
- Insufficient Information1
- Pain1
- Swelling/ Edema1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryMay 1, 2024Naturally Worn; Naturally Worn
AS REPORTED, SIX YEARS AFTER THE INITIAL LEFT THA SURGERY, THE 64 YEAR OLD FEMALE PATIENT COMPLAINED OF KNEE SWELLING AND PAIN. THE SURGEON DIAGNOSED WEAR OF THE TIBIAL INSERT AND A REVISION WAS PERFORMED ON AN UNKNOWN DATE. BREAKAGE AND WEAR WAS OBSERVED IN THE RETRIEVED TIBIAL INSERT. NO METALLOSIS WAS FOUND. THERE WAS NO PROBLEM WITH THE LOCKING MECHANISM OF THE TIBIAL TRAY. ALL PROLIFERATIVE TISSUE WAS REMOVED AND ONLY THE TIBIAL INSERT WAS REPLACED. ALL PARTS/PIECES WERE REMOVED FROM THE PATIENT. THERE WAS NO REPORTED SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.
Patient problems: Failure of Implant; Pain; Swelling/ Edema
MAUDE report 19219940 ↗InjuryJul 17, 2023Naturally Worn; Naturally Worn
IT WAS REPORTED THAT A FEMALE PATIENT, INITIAL RIGHT KNEE IMPLANTED ON (B)(6) 2019, UNDERWENT A REVISION PROCEDURE ON (B)(6) 2023, APPROXIMATELY 3 YEARS 11 MONTHS POST THE INITIAL PROCEDURE. THE REVISION WAS DUE TO POLY WEAR. THE LINER WAS EXCHANGED. THERE WERE NO SURGICAL DELAYS OR DEVICE BREAKAGES DURING THE PROCEDURE. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO PATIENT HISTORY, X-RAYS, OR DEVICE IMAGES WERE PROVIDED. THE EXPLANTED DEVICE IS NOT AVAILABLE FOR RETURN. THE HOSPITAL DISPOSED OF THE DEVICE. NO FURTHER INFORMATION.
Patient problems: Insufficient Information
MAUDE report 17333907 ↗InjuryJun 19, 2020Naturally Worn; Naturally Worn
AS REPORTED, APPROXIMATELY 2 YEARS AND 8 MONTHS POSTOP THE INITIAL IMPLANT, THE (B)(6) Y/O FEMALE PATIENT, PREVIOUSLY TREATED FOR RIGHT KNEE WINDSWEPT DEFORMITY, HAD SEVERE RIGHT KNEE PAIN, THEN A PATELLOFEMORAL ARTHROSIS WAS SUSPECTED BY THE X-RAY FINDINGS. THE SURGEON PREFERS THE SURGICAL PHILOSOPHY WITHOUT RESURFACING PATELLA FOR PRIMARY KNEE REPLACEMENT SURGERY AND HE HAS EXCELLENT LONG-TERM CLINICAL RESULTS FOR THIS PHILOSOPHY, HOWEVER, HE PLANNED TO RESURFACE THE PATELLA USING OUR PATELLA COMPONENT FOR THIS CASE. WHEN THE KNEE WAS EXPOSED, ABNORMAL PROLIFERATIVE SYNOVIUM WAS SEEN AROUND THE KNEE, AND REMARKABLE POLYETHYLENE WEAR OF THE TIBIAL INSERT WAS ALSO FOUND AT THE ANTERO-LATERAL PART OF THE INSERT. THEN HE REMOVED THE INSERT TO CHECK THE CONDITION OF POSTERIOR ASPECT OF THE KNEE, HE CONFIRMED THAT POSTER-MEDICAL PART OF THE INSERT WAS SEVERELY DAMAGED BY DELAMINATION WEAR. AS HE WAS CONCERNED ABOUT USING SAME TYPE TO THE SLOPED INSERT, HE PERFORMED TRIAL REDUCTION USING THE NEUTRAL INSERT OF 9 MM THICKNESS AND DECIDED TO USE IT FOR THIS REPLACEMENT. HE ALSO ADDED V-SHAPE ARTHROTOMY TO ADDRESS THE JOINT STIFFNESS IN FLEXION. THE SURGERY WAS FINISHED INSERTING THE NEUTRA
Patient problems: Failure of Implant
MAUDE report 10177026 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K111400. One submission often covers many models, so check the product and lot details in each record.
- Z-0583-2024 · Nov 21, 2023 · Open, ClassifiedExactech Optetrak Logic Knee System, Catalog Numbers: a) 02-012-47-3011, b) 02-012-47-5009, c) 02-012-47-5011, d) 02-012-65-3013
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page