Skip to content

Optetrak Logic 02-012-49-1513Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak Logic (408)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10885862144423), received between Jun 26, 2022 and Apr 16, 2024 — classified as 4 injuries.

The most frequently reported specific device problem is “Naturally Worn” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K111400); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2021Class IIOpen, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0021-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn3
  • Adverse Event Without Identified Device or Use Problem2
  • Peeled/Delaminated1
  • Unstable1

Patient problems

  • Failure of Implant3
  • Pain3
  • Ambulation or Postural Difficulties1
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Balance Problems1
  • Discomfort1
  • Joint Laxity1
  • Osteomyelitis1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 16, 2024Unstable; Naturally Worn; Unstable; Naturally Worn

    IT WAS REPORTED THAT A 70 YO FEMALE PATIENT, INITIAL (B)(6) 2016, UNDERWENT A REVISION PROCEDURE ON (B)(6) 2024, APPROXIMATELY 7 YEARS 8 MONTHS POST THE INITIAL PROCEDURE. THE PATIENT RETURNED TO THE SURGEON¿S OFFICE WITH COMPLAINTS OF PAIN, SWELLING, INSTABILITY, AND DISSATISFACTION WITH THEIR TKA. THE PATIENT HAS A RECALLED POLY IMPLANT. UPON EXAMINATION, THE PATIENT WAS SCHEDULED TO HAVE A REVISION TKA POSSIBLE POLY SWAP VS FULL REVISION. THE PATIENT UNDERWENT POLY TIBIAL INSERT SWAP AND PATELLA SWAP AND WAS REVISED TO A TRULIANT CRC TIBIAL INSERT SZ 1.5 , 15MM SN: (B)(6) AND AN OPTETRAK 3 PEG PATELLA 29MM SN: (B)(6). THERE WERE NO DEVICE BREAKAGES OR SURGICAL DELAYS DURING THE PROCEDURE. NO X-RAYS WERE PROVIDED. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. THE EXPLANTED DEVICES ARE NOT AVAILABLE FOR RETURN. THEY WERE SENT TO LAB AND LEGAL AT THE HOSPITAL. A DEVICE IMAGE WAS PROVIDED. NO FURTHER INFORMATION.

    Patient problems: Failure of Implant; Joint Laxity; Pain; Swelling/ Edema

    MAUDE report 19122193 ↗
  • InjuryMar 30, 2023Naturally Worn; Adverse Event Without Identified Device or Use Problem; Naturally Worn; Adverse Event Without Identified Device or Use Problem

    AS REPORTED, THE PATIENT HAD AN INITIAL LEFT TKA ON (B)(6) 2017. THE PATIENT RETURNED WITH COMPLAINTS, INCLUDING PAIN AND DISSATISFACTION. THE PATIENT UNDERWENT A POLY SWAP ON (B)(6)2023 AND WAS REVISED TO A SIZE 15MM CRC POLY, SZ 1.5. PIECES OF POLY WERE REMOVED FROM THE PATIENT'S WOUND SITE. THERE WAS NO REPORTED SURGICAL DELAY/PROLONGATION.

    Patient problems: Failure of Implant; Pain

    MAUDE report 16643953 ↗
  • InjuryJul 14, 2022Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED VIA A LEGAL FILING THAT A FEMALE PATIENT UNDERWENT REVISION SURGERY APPROXIMATELY 3 YEARS 8 MONTHS POST HER INITIAL LEFT KNEE TKA. INITIALLY IMPLANTED WITH AN OPTETRAK DEVICE, INCLUDING A SIZE 1.5, 13 MM OPTETRAK LOGIC TIBIAL INSERTS MADE OF POLYETHYLENE. THE PLAINTIFF WAS DIAGNOSED WITH FAILED TOTAL ARTHROPLASTY WITH EARLY POLYETHYLENE FAILURE. BEFORE THE REVISION SURGERY, SHE WAS ALSO DIAGNOSED WITH OSTEOMYELITIS IN HER KNEE DUE TO THE BREAKDOWN OF THE POLYETHYLENE. DURING THE REVISION SURGERY, THE PLAINTIFF'S SURGEON STATED, "WE WERE ABLE TO LEVER THE POLYETHYLENE PLASTIC WITH A SMALL OSTEOTOME AND AFTER THIS WAS RETRIEVED, WE THEN INSPECTED THE PLASTIC, NOTED TO HAVE PITTING AND DELAMINATION OF THE MEDIAL ASPECT OF THE TIBIAL POLYETHYLENE. OTHERWISE, THE POLYETHYLENE LOOKED TO BE IN GOOD CONDITION LATERALLY."

    Patient problems: Osteomyelitis

    MAUDE report 15020156 ↗
  • InjuryJun 26, 2022Naturally Worn; Adverse Event Without Identified Device or Use Problem; Naturally Worn; Adverse Event Without Identified Device or Use Problem

    AS REPORTED BY LEGAL BRIEF, PATIENT'S INITIAL TOTAL KNEE ARTHROPLASTY WAS PERFORMED ON (B)(6) 2018. PATIENT'S UNDERWENT REVISION TOTAL KNEE ARTHROPLASTY ON (B)(6) 2021 DUE TO PREMATURE FAILURE OF HIS EXACTECH OPTETRAK TOTAL KNEE DEVICE.

    Patient problems: Ambulation or Postural Difficulties; Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Balance Problems; Discomfort; Failure of Implant; Pain; Scar Tissue; Synovitis

    MAUDE report 14824552 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K111400. One submission often covers many models, so check the product and lot details in each record.

  • Z-0583-2024 · Nov 21, 2023 · Open, Classified
    Exactech Optetrak Logic Knee System, Catalog Numbers: a) 02-012-47-3011, b) 02-012-47-5009, c) 02-012-47-5011, d) 02-012-65-3013

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page