Skip to content

Exactech 170-40-00Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10885862155504), received between Sep 16, 2024 and Feb 19, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Naturally Worn” (2 reports).

7 other recalls cover devices cleared under the same FDA submission (K121392); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Naturally Worn2
  • Scratched Material1

Patient problems

  • Bone Shedding Debris1
  • Failure of Implant1
  • Metal Related Pathology1
  • Pain1
  • Scar Tissue1
  • Swelling/ Edema1
  • Synovitis1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 19, 2025Naturally Worn; Scratched Material; Naturally Worn; Scratched Material

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 36 MONTHS AFTER A LEFT TOTAL HIP REPLACEMENT, THE PATIENT UNDERWENT LEFT HIP REVISION TO ADDRESS PROSTHESIS WEAR. A POST OPERATIVE OP REPORT WAS PROVIDED. POST OP DIAGNOSIS NOTED CHRONICALLY FAILED TOTAL HIP ARTHROPLASTY DUE TO POLYETHYLENE WEAR. INTRAOPERATIVELY, THE SURGEON OBSERVED METALLOSIS AND SYNOVITIS AND FRAGMENTED POLYETHYLENE LINER PIECES IN THE JOINT CAUSING THE METALLOSIS. ADDITIONALLY, THE SURGEON OBSERVED A LINEAR SCRATCH ON THE SURFACE OF THE FEMORAL HEAD. BOTH LINER AND FEMORAL HEAD COMPONENTS WERE REPLACED. IT WAS NOTED THAT THE PATIENT'S WEIGHT CONTRIBUTED TO A MORE COMPLEX APPROACH AND POSITIONING DIFFICULTIES, INTRAOPERATIVELY. NO SURGICAL OR THIS CANNOT BE CONFIRMED FROM THE REPORTED INFORMATION AND THE DEVICES WERE NOT AVAILABLE FOR EVALUATION.

    Patient problems: Bone Shedding Debris; Failure of Implant; Metal Related Pathology; Scar Tissue; Synovitis

    MAUDE report 21411605 ↗
  • InjurySep 16, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THIS 59 Y/O MALE PATIENT'S LEFT HIP WAS REVISED APPROXIMATELY 5 YEARS POST OP. THE PATIENT RETURNED TO SURGEONS OFFICE COMPLAINING OF PAIN, SWELLING, AND DISSATISFACTION WITH THEIR LEFT PRIMARY TOTAL HIP. THE POLY AND A RECALLED IMPLANT WERE IN QUESTION AND SHOWN TO BE WORN ON X-RAY. THE PATIENT UNDERWENT AN ACETABULAR POLY IMPLANT AND A FEMORAL HEAD SWAP. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. PRODUCT NOT RETURNING: IMPLANTS ARE SENT TO THE LAB AND LEGAL.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 20233178 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K121392. One submission often covers many models, so check the product and lot details in each record.

  • Z-1732-2022 · Aug 11, 2022 · Open, Classified
    GXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. NOVATION GXL LINER NEUTRAL, 32MM ID GROUP 1 CUPS, REF 130-32-51 b. NOVATION GXL LINER NEUTRAL,
  • Z-2129-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-40-XX, Novation GXL Liner, +5mm Lateralized, 40mm. Orthopedic hip implant
  • Z-2123-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-40-XX, Novation GXL Liner, Lipped Ant, 40mm; Novation GXL Liner, Lipped, 40
  • Z-2118-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-36-XX, Acumatch GXL 0 Degree Liner, 36mm; and Novation GXL Liner, Neutral,
  • Z-2133-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-32-XX, Acumatch GXL 15 Degree Liner, 32mm; Novation GXL Liner, Lipped Ant,
  • Z-2119-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-40-XX, Novation GXL Liner, Neutral, 40mm. Orthopedic hip implant component
  • Z-2117-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-32-XX, Acumatch GXL 0 Degree Liner, 32mm; and Novation GXL Liner, Neutral,

Other Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate devices

All Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page