Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10885862158147), received between Sep 19, 2023 and Sep 29, 2026 — classified as 6 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (6 reports).

1 other recall cover devices cleared under the same FDA submission (K111400); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth6
  • Unstable2
  • Insufficient Device Problem Information1
  • Mechanical Problem1

Patient problems

  • Pain4
  • Synovitis4
  • Failure of Implant3
  • Insufficient Information3
  • Osteolysis3
  • Joint Laxity2
  • Ankylosis1
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjurySep 29, 2026Loosening of Implant Not Related to Bone-Ingrowth

    Patient problems: Insufficient Information

    MAUDE report 50933386 ↗
  • InjuryJun 23, 2025Unstable; Loosening of Implant Not Related to Bone-Ingrowth; Unstable; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT A MALE PATIENT, WHO HAD A RIGHT KNEE IMPLANTED, UNDERWENT A REVISION PROCEDURE APPROXIMATELY 5 YEARS POST THE INITIAL PROCEDURE. THE PATIENT PRESENTED WITH INCREASING KNEE PAIN, A SENSE OF INSTABILITY, AND SWELLING. PLAIN RADIOGRAPHS REVEALED POLYETHYLENE WEAR, LOOSENING OF BOTH THE FEMORAL AND TIBIAL COMPONENTS, AND OSTEOLYSIS AFFECTING BOTH BONES. DURING THE REVISION PROCEDURE, A FEMORAL AND TIBIAL BONE DEFECT WAS IDENTIFIED CONSISTENT WITH AORI TYPE 2B. THERE WAS ALSO A SIGNIFICANT PATELLAR DEFECT AND A NOTABLE SOFT-TISSUE REACTION CHARACTERIZED BY SYNOVITIS. THE PATIENT WAS REVISED TO A COMPETITOR¿S DEVICES. THERE WERE NO DEVICE BREAKAGES OR SURGICAL DELAYS DURING THE PROCEDURE. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. X-RAYS WERE PROVIDED. THE EXPLANTED DEVICES ARE NOT AVAILABLE FOR RETURN. NO FURTHER INFORMATION. THIS IS 1 OF 4 REPORTS

    Patient problems: Inadequate Osseointegration; Joint Laxity; Osteolysis; Pain; Swelling/ Edema; Synovitis

    MAUDE report 22312189 ↗
  • InjuryDec 12, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 89 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, OSTEOLYSIS WITH CYSTIC FORMATION, BONE DEGRADATION, ASEPTIC LOOSENING, PAIN, FEMORAL LOOSENING, LOOSE TIBIAL BASEPLATE. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Bone Fracture(s); Cyst(s); Failure of Implant; Joint Laxity; Osteolysis; Pain; Synovitis

    MAUDE report 20919801 ↗
  • InjuryAug 26, 2024Unstable; Loosening of Implant Not Related to Bone-Ingrowth; Unstable; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED VIA LEGAL DOCUMENTATION THE 59 Y/O MALE PATIENT HAD A RIGHT KNEE REPLACEMENT ON (B)(6) 2015. APPROXIMATELY 7 YEARS AND 2 MONTHS AFTER THE INITIAL PROCEDURE THE PATIENT HAD A RIGHT KNEE REVISION ON (B)(6) 2022. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 20065243 ↗
  • InjuryJun 7, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT APPROXIMATELY 90 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, LOOSENING, AND SYNOVITIS. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Implant Pain; Osteolysis; Pain; Synovitis

    MAUDE report 19483742 ↗
  • InjurySep 19, 2023Mechanical Problem; Insufficient Device Problem Information; Loosening of Implant Not Related to Bone-Ingrowth; Mechanical Problem; Insufficient Device Problem Information; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A RIGHT KNEE REPLACEMENT ON (B)(6) 2013. APPROXIMATELY 2 YEARS AND 8 MONTHS AFTER THE INITIAL PROCEDURE THE PATIENT HAD A RIGHT KNEE REVISION ON (B)(6) 2015. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE PATIENT HAS FILED A SHORT-FORM COMPLAINT IN A MULTIDISTRICT LITIGATION TITLED IN RE: EXACTECH POLYETHYLENE ORTHOPEDIC PRODUCTS LIABILITY LITIGATION, MDL NO. 3044, AND PENDING IN THE EASTERN DISTRICT OF NEW YORK. PER THE TRANSFER ORDER CREATING THIS MULTIDISTRICT LITIGATION, ¿PLAINTIFFS ALLEGE THAT THEIR KNEE OR HIP REPLACEMENT DEVICES¿FAILED PREMATURELY BECAUSE OF DEGRADATION OF THE DEVICE¿S POLYETHYLENE COMPONENT, WHICH RESULTED IN THE PREMATURE REMOVAL (OR PLANNED REMOVAL) OF THE PROSTHESIS AT ISSUE.¿ BECAUSE THE PATIENT HAS FILED A SUIT IN THIS MULTIDISTRICT LITIGATION, THE PATIENT APPEARS TO ALLEGE THAT THE PATIENT WAS INJURED AS A RESULT OF PREMATURE WEAR OF AN EXACTECH POLYETHYLENE DEVICE.

    Patient problems: Ankylosis; Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant; Insufficient Information; Loss of Range of Motion; Pain; Scar Tissue; Synovitis

    MAUDE report 17772085 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K111400. One submission often covers many models, so check the product and lot details in each record.

  • Z-0583-2024 · Nov 21, 2023 · Open, Classified
    Exactech Optetrak Logic Knee System, Catalog Numbers: a) 02-012-47-3011, b) 02-012-47-5009, c) 02-012-47-5011, d) 02-012-65-3013

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →