Skip to content

Optetrak Logic 02-012-44-2015Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak Logic (408)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
10
Deaths reported
0
Injuries reported
10
Recalls naming it
1

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10885862173508), received between Nov 9, 2020 and Apr 21, 2025 — classified as 10 injuries.

The most frequently reported specific device problem is “Naturally Worn” (9 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2021Class IIOpen, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0021-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn9
  • Unstable4
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Physical Resistance/Sticking1

Patient problems

  • Osteolysis5
  • Synovitis5
  • Failure of Implant4
  • Implant Pain4
  • Joint Laxity4
  • Pain4
  • Ambulation or Postural Difficulties2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 21, 2025Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT A 73 YO FEMALE PATIENT, INITIAL RIGHT KNEE IMPLANTED IN (B)(6) 2018, UNDERWENT A REVISION PROCEDURE IN (B)(6) 2025, APPROXIMATELY 6 YEARS 9 MONTHS POST THE INITIAL PROCEDURE. THE SURGEON PERFORMED A SYNOVECTOMY SURROUNDING THE AFFECTED KNEE AND FOUND SEVERE OSTEOLYSIS PRESENT. THE PATELLA EXHIBITED MEDIAL TO LATERAL WEAR ALONG THE TRACKING PATH. THE TIBIAL INSERT HAD MILD MEDIAL WEAR. ADDITIONAL WEAR ON THE POSTERIOR POST FOUND. THE FEMORAL AND TIBIAL COMPONENTS WERE WELL FIXED. THE PATELLA AND PSC INSERT WERE REPLACED WITH SAME SIZE/TYPE COMPONENT. THERE WERE NO DEVICE BREAKAGES OR SURGICAL DELAYS DURING THE PROCEDURE. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO X-RAYS WERE ABLE TO BE OBTAINED. THE EXPLANTED DEVICES ARE NOT AVAILABLE FOR RETURN. THEY ARE BEING RETAINED BY THE HOSPITAL FOR ANALYSIS. DEVICE IMAGES WERE PROVIDED. NO FURTHER INFORMATION. THIS IS 1 OF 2 REPORTS FOR THIS EVENT.

    Patient problems: Osteolysis

    MAUDE report 21877160 ↗
  • InjuryJan 21, 2025Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 68 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, DISCOMFORT, DISTRESS, PAIN, SCAR TISSUE, SYNOVITIS, AND OSTEOLYSIS. INITIAL SURGERY AND REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Discomfort; Distress; Osteolysis; Pain; Scar Tissue; Synovitis

    MAUDE report 21197569 ↗
  • InjurySep 30, 2024Unstable; Naturally Worn; Unstable; Naturally Worn

    LEGAL CASE ¿ USA. (B)(4) LK REV 1 IS ASSOCIATED WITH THIS CASE. IT WAS REPORTED THAT APPROXIMATELY 66 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR PAINFUL AND UNSTABLE LEFT KNEE; EXTENSIVE SYNOVITIS, OSTEOLYSIS ON FEMUR AND PATELLA, ASEPTIC LOOSENING OF FEMORAL, WEAR OF THE LINER WITH DELAMINATION, WEAR OF THE PATELLA. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE SERIAL NUMBER: (B)(6) IS CONFIRMED TO BE A PART OF RECALL NUMBER: Z-0021-2022.

    Patient problems: Joint Laxity; Osteolysis; Pain; Synovitis

    MAUDE report 20337233 ↗
  • InjurySep 4, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 75 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT HAS EXPERIENCED PROSTHESIS WEAR. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Aseptic loosening; Cyst(s); Joint effusion; Synovitis

    MAUDE report 20136469 ↗
  • InjuryJun 11, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 122 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PAIN, LARGE EFFUSION, SYMPTOMATIC POLYETHYLENE WEAR, OXIDATION. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Insufficient Information; Joint effusion; Pain

    MAUDE report 19508790 ↗
  • InjuryJun 4, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 10 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PAIN, STIFFNESS, AND WEAKNESS. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Implant Pain

    MAUDE report 19463231 ↗
  • InjuryAug 30, 2023Unstable; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth; Unstable; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED VIA LEGAL DOCUMENTATION, A PATIENT HAD LEFT KNEE REPLACEMENT SURGERY ON (B)(6) 2016. THEY UNDERWENT LEFT KNEE REVISION SURGERY ON (B)(6) 2023, APPROXIMATELY 6 YEARS 5 MONTHS POST PRIMARY SURGERY. REVISION OP REPORT STATED THAT THE IMPLANTS WERE GROSSLY UNSTABLE IN THE AP AND ML DIRECTION. THE FEMORAL CONE COMPONENT WAS DISSECTED AND EASILY REMOVED WITH A GENTLE TAP, THE IMPLANT SEPARATING AT THE CEMENT INTERFACE. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Implant Pain; Joint Laxity

    MAUDE report 17654653 ↗
  • InjuryApr 24, 2023Unstable; Naturally Worn; Unstable; Naturally Worn

    PER A REPORT FROM THE LEGAL DEPARTMENT THE PATIENT HAD AN INITIAL RIGHT KNEE REPLACEMENT IN (B)(6) 2018. APPROXIMATELY 4 YEARS AND 8 MONTHS AFTER THE INITIAL SURGERY, THE PATIENT HAD A RIGHT KNEE REVISION ON (B)(6) 2022 DUE TO ACCELERATED AND PREVENTABLE WEAR. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Ambulation or Postural Difficulties; Balance Problems; Implant Pain; Insufficient Information; Joint Laxity; Osteolysis; Pain; Swelling/ Edema; Synovitis

    MAUDE report 16794382 ↗
  • InjuryFeb 27, 2023Unstable; Naturally Worn; Unstable; Naturally Worn

    AS REPORTED VIA LEGAL DOCUMENTATION, A PATIENT HAD BILATERAL KNEE REPLACEMENT ON (B)(6)2018. THEY SUBSEQUENTLY UNDERWENT LEFT KNEE REVISION SURGERY ON (B)(6) 2023, APPROXIMATELY 4 YEARS 7 MONTHS POST PRIMARY PROCEDURE. REVISION OP REPORT POSTOPERATIVE DIAGNOSIS: S/P RECALLED TKA WITH POLY WEAR, SYNOVITIS AND ASEPTIC LOOSENING OF FEMUR. THE PATELLAR BUTTON WAS FOUND TO BE INTACT BUT WITH SOME SURFACE WEAR. THE POLYETHYLENE WAS NOTED TO HAVE SOME DAMAGE THAT WAS NOT CLEARLY VISIBLE BUT COULD PALPATE DEFT. THE FEMORAL COMPONENT WAS NOTED TO HAVE COMPLETELY DEBONDED FROM ITS CEMENT MANTLE. THE CEMENT MANTLE WAS INTACT, AND WHEN IT WAS REMOVED, THERE WAS NOTED TO BE AORI TYPE IIB BONE LOSS, WORSE ON THE LATERAL COMPARED TO MEIAL SIDE, AND OSTEOLYSIS AFFECTING BOTH POSTERIOR CONDYLES. THE TIBIAL COMPONENT WAS FOUND TO BE GROSSLY INTACT, NOT LOOSE. THE PATIENT WAS TRANSFERRED TO THE RECOVERY ROOM IN STABLE CONDITION. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Ambulation or Postural Difficulties; Discomfort; Failure of Implant; Implant Pain; Joint Laxity; Osteolysis; Swelling/ Edema; Synovitis

    MAUDE report 16450198 ↗
  • InjuryNov 9, 2020Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT A PATIENT EXPERIENCED A RIGHT KNEE REVISION SURGERY ON (B)(6) 2013, DUE TO RANGE OF MOTION/STIFFNESS. INITIAL SURGERY DATE WAS (B)(6) 2012. NO ADDITIONAL INFORMATION ABOUT THE EVENT OR PATIENT HAS BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 10808430 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page