Skip to content

Optetrak Logic 02-012-44-2511Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak Logic (408)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
10
Deaths reported
0
Injuries reported
10
Recalls naming it
1

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10885862173522), received between Feb 21, 2022 and May 4, 2026 — classified as 10 injuries.

The most frequently reported specific device problem is “Naturally Worn” (9 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2021Class IIOpen, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0021-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn9
  • Unstable7
  • Insufficient Device Problem Information1
  • Lack of Effect1
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Mechanical Problem1

Patient problems

  • Pain8
  • Failure of Implant7
  • Swelling/ Edema7
  • Joint Laxity6
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4
  • Synovitis4
  • Osteolysis3
  • Ambulation or Postural Difficulties2

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 4, 2026Unstable; Naturally Worn

    AS REPORTED VIA LEGAL DOCUMENT, REVISION OPERATIVE REPORT, A PATIENT, HAD A FAILED RIGHT TOTAL KNEE REPLACEMENT REFRACTORY TO CONSERVATIVE MANAGEMENT. THE PATIENT WAS INDICATED FOR SURGERY. A THOROUGH SYNOVECTOMY WAS PERFORMED. THERE WAS A FRACTURE AT THE PROXIMAL TIP OF THE FEMORAL COMPONENT STEM WHICH WAS VISUALIZED AND CLEANED UP. THE FEMORAL FRACTURE WAS REDUCED WITH POINTED REDUCTION CLAMPS. FRESHENING CUTS WERE MADE TO THE FEMUR. THEY WERE ABLE TO OBTAIN AN ANATOMIC REDUCTION OF THE FRACTURE WHICH WAS STABLE WITH THE STRUT IN PLACE. THE FEMORAL IMPLANT WAS REMOVED WITH MINIMAL BONE LOSS. THE TIBIA WAS INTACT AND WELL FIXED. NEW FEMORAL COMPONENT AND INSERT WERE IMPLANTED. THE PATIENT WAS TRANSFERRED TO THE RECOVERY ROOM IN STABLE CONDITION. THERE IS NO MENTION OF ANY DEFECTS TO THE POLYETHYLENE LINER IN THIS OPERATIVE REPORT. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. ADDITIONAL INFORMATION INDICATED THE PATIENT UNDERWENT SURGERY ON HER RIGHT KNEE AND WAS IMPLANTED WITH A RECALLED OPTETRAK POLYETHYLENE TIBIAL INSERT (SERIAL #: (B)(6). THE PATIENT UNDERWENT RIGHT KNE

    Patient problems: Ambulation or Postural Difficulties; Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Bone Fracture(s); Discomfort; Joint Laxity; Osteolysis; Pain

    MAUDE report 25075746 ↗
  • InjuryAug 28, 2024Unstable; Naturally Worn; Unstable; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 31 MONTHS AFTER A TOTAL REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PAIN, SWELLING AND INSTABILITY. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Joint Laxity; Pain; Swelling/ Edema

    MAUDE report 20087786 ↗
  • InjuryJun 4, 2024Unstable; Naturally Worn; Unstable; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 31 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PAIN, INSTABILITY, AND POLYETHYLENE DEBRIS. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant; Implant Pain; Joint Laxity; Osteolysis; Pain; Swelling/ Edema; Synovitis

    MAUDE report 19463440 ↗
  • InjuryMay 22, 2024Loosening of Implant Not Related to Bone-Ingrowth; Lack of Effect; Loosening of Implant Not Related to Bone-Ingrowth; Lack of Effect

    IT WAS REPORTED THAT APPROXIMATELY 55 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, SEVERE POLY WEAR ON THE ARTICULAR SURFACE, FEMORAL COMPONENT WAS LOOSE AND DEBONDED FROM THE CEMENT, PATELLAR COMPONENT HAD MILD WEAR. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Ambulation or Postural Difficulties; Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Discomfort; Distress; Failure of Implant; Loss of Range of Motion; Pain; Swelling/ Edema

    MAUDE report 19374380 ↗
  • InjuryOct 3, 2023Mechanical Problem; Unstable; Naturally Worn; Mechanical Problem; Unstable; Naturally Worn

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A LEFT KNEE REPLACEMENT ON (B)(6)2019. APPROXIMATELY 4 YEARS AND 2 MONTHS AFTER THE INITIAL PROCEDURE THE PATIENT HAD A LEFT KNEE REVISION ON (B)(6) 2023. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. REVISION OP REPORT ON (B)(6) 2023. DIAGNOSIS: PERIPROSTHETIC OSTEOLYSIS. CLEAR SYNOVIAL FLUID, WITHOUT SIGNS OF INFECTION. COMPLETE SYNOVECTOMY WAS PERFORMED ON THE MEDIAL AND LATERAL GUTTERS. THERE WAS SIGNIFICANT SYNOVITIS AND THE APPEARANCE OF WEAR PARTICLES IN THE SYNOVIUM BEHIND THE PATELLA. THE POLYETHYLENE INSERT WAS REMOVED AND EXTENSIVE WEAR OF THE POLY, PARTICULARLY IN THE POSTERIOR PORTION OF THE INSERT. THERE WERE SOME AREAS OF CONTAINED OSTEOLYSIS IN THE FEMORAL BONE. THE PATIENT WAS TRANSFERRED TO THE RECOVERY ROOM IN STABLE CONDITION.

    Patient problems: Failure of Implant; Joint Laxity; Loss of Range of Motion; Osteolysis; Pain; Swelling/ Edema; Synovitis

    MAUDE report 17859510 ↗
  • InjurySep 7, 2023Unstable; Naturally Worn; Unstable; Naturally Worn

    AS REPORTED VIA LEGAL DOCUMENTATION, A PATIENT HAD LEFT KNEE REPLACEMENT SURGERY ON (B)(6) 2014. THEY SUBSEQUENTLY UNDERWENT LEFT KNEE REVISION SURGERY ON (B)(6) 2023, APPROXIMATELY 8 YEARS 9 MONTHS POST PRIMARY PROCEDURE. THE PATIENT HAS CLAIMED THE FOLLOWING INJURIES AND COMPLICATIONS: SURGICAL REVISION AND OTHER MEDICAL TREATMENT, PAIN, SWELLING, STIFFNESS, LOSS OF MOTION, WEAKNESS, INSTABILITY, AND OTHER INJURIES AND/OR COMPLICATIONS. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE PATIENT HAS FILED A SHORT-FORM COMPLAINT IN A MULTIDISTRICT LITIGATION TITLED IN RE: EXACTECH POLYETHYLENE ORTHOPEDIC PRODUCTS LIABILITY LITIGATION, MDL NO. 3044, AND PENDING IN THE EASTERN DISTRICT OF NEW YORK. PER THE TRANSFER ORDER CREATING THIS MULTIDISTRICT LITIGATION, ¿PLAINTIFFS¿ALLEGE THAT THEIR KNEE OR HIP REPLACEMENT DEVICES¿FAILED PREMATURELY BECAUSE OF DEGRADATION OF THE DEVICE¿S POLYETHYLENE COMPONENT, WHICH RESULTED IN THE PREMATURE REMOVAL (OR PLANNED REMOVAL) OF THE P

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant; Pain; Swelling/ Edema

    MAUDE report 17698775 ↗
  • InjuryAug 15, 2023Unstable; Naturally Worn; Unstable; Naturally Worn

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A RIGHT KNEE REPLACEMENT ON (B)(6)2015. APPROXIMATELY 6 YEARS AND 11 MONTHS AFTER THE INITIAL PROCEDURE THE PATIENT HAD A RIGHT KNEE REVISION ON (B)(6) 2022. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. REVISION OP REPORT ON (B)(6) 2022. DIAGNOSIS: FAILED RIGHT KNEE SECONDARY TO ACCELERATED WEAR OF THE TIBIAL POLYETHYLENE BEARING. THE RIGHT KNEE WAS ASPIRATED OF SYNOVIAL FLUID. A SUBTOTAL SYNOVECTOMY WAS PERFORMED REMOVING THE HYPERTROPHIC SYNOVIUM FROM THE SUPRAPATELLAR POUCH AND WORKING DOWN THE MEDIAL AND LATERAL GUTTERS. THERE WAS MODERATE TO SEVERE PITTING OF THE ARTICULAR WEIGHTBEARING SURFACES ON THE INSERT. THE PATIENT WAS TAKEN TO THE RECOVERY ROOM IN STABLE CONDITION.

    Patient problems: Failure of Implant; Joint Laxity; Pain; Swelling/ Edema; Synovitis

    MAUDE report 17548988 ↗
  • InjuryMay 8, 2023Unstable; Naturally Worn; Insufficient Device Problem Information; Unstable; Naturally Worn; Insufficient Device Problem Information

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A RIGHT KNEE REPLACEMENT ON (B)(6) 2015 . APPROXIMATELY 6 YEARS AND 11 MONTHS AFTER THE INITIAL PROCEDURE THE PATIENT HAD A RIGHT KNEE REVISION ON (B)(6) 2022. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Foreign Body Reaction; Insufficient Information; Joint Laxity; Pain; Swelling/ Edema; Synovitis

    MAUDE report 16884761 ↗
  • InjuryOct 18, 2022Naturally Worn; Naturally Worn

    AS REPORTED, THE PATIENT HAD AN INITIAL RIGHT TKA ON (B)(6) 2018. THE PATIENT WAS REVISED DUE TO WEAR AND LOOSENING AND THE TIBIAL INSERT AND BASEPLATE WERE REPLACED. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO OTHER PATIENT INFORMATION/MEDICAL HISTORY PROVIDED.

    Patient problems: Aseptic loosening; Failure of Implant

    MAUDE report 15623874 ↗
  • InjuryFeb 21, 2022Naturally Worn; Naturally Worn

    AS REPORTED, APPROXIMATELY SEVEN YEARS POST INITIAL LEFT SIDE TKA, THIS (B)(6) PATIENT PRESENTED TO SURGEON WITH SIGNIFICANT POLY WEAR WITH SIGNIFICANT OSTEOLYSIS. SURGEON PERFORMED A TOTAL REVISION KNEE. THERE WAS NO BREAKAGE OF THE DEVICE. THERE WAS NO SURGICAL DELAY/PROLONGATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. PHOTOS AND X-RAYS RECEIVED. THE DEVICE WILL BE RETURN.

    Patient problems: Failure of Implant

    MAUDE report 13569775 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page