Optetrak Logic 02-012-49-1009Recalls & reported problems
Exactech, Inc. · Class 2 · Implant
All models of Optetrak Logic (408)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10885862174208), received between Jan 13, 2021 and Nov 1, 2022 — classified as 2 injuries.
The most frequently reported specific device problem is “Fracture” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K111400); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Aug 30, 2021Class IIOpen, Classified
Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.
Z-0021-2022 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Fracture1
- Naturally Worn1
Patient problems
- Failure of Implant2
- Pain1
- Swelling/ Edema1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryNov 1, 2022Naturally Worn; Naturally Worn
AS REPORTED, APPROXIMATELY 3 YEARS POST OP THE INITIAL LEFT TKA, THIS 71 Y/O FEMALE PATIENT WAS REVISED DUE TO WEAR OF THE TIBIAL INSERT. THE PATIENT WAS EXPERIENCING KNEE SWELLING AND PAIN. BREAKAGE AND WEAR WERE OBSERVED ON THE RETRIEVED TIBIAL INSERT. THE SURGEON DECIDED THAT THERE WAS NO PROBLEM FOUND ON THE LOCKING MECHANISM OF THE TIBIAL TRAY, AND AFTER REMOVING ALL PROLIFERATIVE TISSUES, ONLY THE INSERT WAS REPLACED. DEVICE IS NOT RETUNING, DEVICE IS GIVEN TO PATIENT.
Patient problems: Failure of Implant; Pain; Swelling/ Edema
MAUDE report 15708061 ↗InjuryJan 13, 2021Fracture; Fracture
AS REPORTED, APPROXIMATELY THIRTY MONTHS POSTOP THE INITIAL RIGHT TKA, THE (B)(6) Y/O FEMALE PATIENT COMPLAINED KNEE SWELLING AND PAIN, THE SURGEON DIAGNOSED THE WEAR OF THE TIBIAL INSERT, THEN REVISION SURGERY FOR THE REPLACEMENT OF THE TIBIAL INSERT WAS PERFORMED. BREAKAGE AND WEAR WERE OBSERVED IN THE RETRIEVED TIBIAL INSERT. REMOVING ALL PROLIFERATIVE ISSUES, THE INSERT WAS REPLACED TO NEW ONE. THE SURGEON JUDGED THAT THERE IS NO PROBLEM WITH THE STABILITY OF THE TIBIAL TRAY AND THE LOCKING MECHANISM OF THE TIBIAL TRAY FOR TIBIAL INSERT FIXATION, THE INSERT WAS REPLACED TO 9MM THICKNESS OF SIZE 1 CR NEURAL TIBIAL INSERT.
Patient problems: Failure of Implant
MAUDE report 11163060 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K111400. One submission often covers many models, so check the product and lot details in each record.
- Z-0583-2024 · Nov 21, 2023 · Open, ClassifiedExactech Optetrak Logic Knee System, Catalog Numbers: a) 02-012-47-3011, b) 02-012-47-5009, c) 02-012-47-5011, d) 02-012-65-3013
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page