Optetrak Logic 02-012-47-2013Recalls & reported problems
Exactech, Inc. · Class 2 · Implant
All models of Optetrak Logic (408)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10885862174260), received between Sep 15, 2022 and Jul 13, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Naturally Worn” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K111400); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Aug 30, 2021Class IIOpen, Classified
Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.
Z-0021-2022 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Naturally Worn2
- Fracture1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Discomfort1
- Failure of Implant1
- Insufficient Information1
- Joint Laxity1
- Osteolysis1
- Pain1
- Synovitis1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJul 13, 2023Naturally Worn; Naturally Worn
AS REPORTED VIA LEGAL DOCUMENTATION, A PATIENT HAD LEFT KNEE REPLACEMENT ON (B)(6) 2018. THEY SUBSEQUENTLY UNDERWENT LEFT KNEE REVISION ON (B)(6) 2022, APPROXIMATELY 7 YEARS 2 MONTHS AFTER THEIR INITIAL IMPLANTATION. ON (B)(6) 2022 OP REPORT NOTED SIGNIFICANT SEROUS FLUID IDENTIFIED IN THE KNEE AND SIGNIFICANT SYNOVIAL HYPERTROPHY AND INFLAMMATION. THE POLYETHYLENE HAD SIGNIFICANT EROSION AND DELAMINATION AFFECTING THE MEDIAL AND THE POSTERIOR QUARTER OF THE POLYETHYLENE LINER. ALL COMPONENTS WERE SOLIDLY FIXED. THERE WAS MINIMAL EROSION MEDIALLY ON THE TIBIAL TRAY BUT WAS EXTREMELY STABLE. THE PATIENT TOLERATED THE PROCEDURE QUITE NICELY. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Insufficient Information; Synovitis
MAUDE report 17321038 ↗InjurySep 15, 2022Fracture; Naturally Worn; Fracture; Naturally Worn
AS REPORTED, THE PATIENT HAD AN INITIAL RIGHT TKA ON (B)(6) 2019. THE PATIENT WAS REVISED DUE TO WEAR IN THE LINER/TIBIAL COMPONENTS. THE TIBIA, LINER AND PATELLA WERE REVISED TO EXACTECH PRODUCTS. THERE WAS NO SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO OTHER PATIENT INFORMATION/MEDICAL HISTORY PROVIDED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Discomfort; Failure of Implant; Joint Laxity; Osteolysis; Pain
MAUDE report 15426451 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K111400. One submission often covers many models, so check the product and lot details in each record.
- Z-0583-2024 · Nov 21, 2023 · Open, ClassifiedExactech Optetrak Logic Knee System, Catalog Numbers: a) 02-012-47-3011, b) 02-012-47-5009, c) 02-012-47-5011, d) 02-012-65-3013
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page