Skip to content

Optetrak Logic 02-012-49-3009Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak Logic (408)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
11
Deaths reported
0
Injuries reported
11
Recalls naming it
1

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10885862174413), received between Jun 30, 2020 and Sep 26, 2024 — classified as 11 injuries.

The most frequently reported specific device problem is “Naturally Worn” (11 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K111400); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2021Class IIOpen, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0021-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn11
  • Fracture3
  • Unstable2
  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Device Problem Information1

Patient problems

  • Failure of Implant7
  • Swelling/ Edema5
  • Insufficient Information4
  • Pain4
  • Joint Laxity3
  • Osteolysis3
  • Ambulation or Postural Difficulties2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjurySep 26, 2024Naturally Worn; Naturally Worn

    LEGAL CASE ¿ (B)(4) IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 27 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PAIN, SWELLING, INSTABILITY, OSTEOLYSIS. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information; Joint Laxity; Osteolysis; Pain; Swelling/ Edema

    MAUDE report 20310422 ↗
  • InjurySep 5, 2024Naturally Worn; Naturally Worn

    LEGAL CASE ¿ USA (MDL NO. 3044) (NGG) (MMH). IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 76 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT HAS EXPERIENCED PROSTHESIS WEAR, PAIN, MENTAL AND EMOTIONAL DISTRESS. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE. SERIAL NUMBER (B)(6) IS CONFIRMED TO HAVE BEEN PACKAGED IN A VACUUM BAG THAT DOES NOT CONTAIN EVOH. 510(K): K111400.

    Patient problems: Emotional Changes; Pain

    MAUDE report 20153208 ↗
  • InjuryAug 6, 2024Unstable; Naturally Worn; Unstable; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 113 MONTHS AFTER A RIGHT KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, SIGNIFICANT PAIN AND DISCOMFORT; GAIT IMPAIRMENT; POOR BALANCE; DIFFICULTY WALKING; COMPONENT PART LOOSENING; SOFT TISSUE DAMAGE; BONE LOSS. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Ambulation or Postural Difficulties; Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Balance Problems; Discomfort; Joint Laxity; Osteolysis; Pain; Swelling/ Edema

    MAUDE report 19925040 ↗
  • InjuryJul 20, 2023Naturally Worn; Naturally Worn

    AS REPORTED, APPROXIMATELY FOUR YEARS POST INITIAL LEFT SIDE TKA, THE FEMALE PATIENT HAD A REVISION DUE TO POLY WEAR. PATIENT WAS REVISED INTO A LOGIC PS TKR. THERE WAS NO BREAKAGE OF DEVICE OR SURGICAL DELAY/PROLONGATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. IMAGES RECEIVED. SALES REP WAS UNABLE TO OBTAIN X-RAYS. THE DEVICES ARE NOT AVAILABLE FOR EVALUATION AS THEY WERE DISPOSED AT THE HOSPITAL, NO OTHER PATIENT INFORMATION/MEDICAL HISTORY REPORTED.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 17361108 ↗
  • InjuryMay 8, 2023Fracture; Unstable; Naturally Worn; Insufficient Device Problem Information; Fracture; Unstable; Naturally Worn; Insufficient Device Problem Information

    AS REPORTED VIA LEGAL DOCUMENTATION, THIS PATIENT RIGHT KNEE REPLACEMENT ON (B)(6) 2012. HE HAD RIGHT KNEE REVISION ON (B)(6) 2017. APPROXIMATELY, 4 YEARS 9 MONTHS AFTER THEIR INITIAL REPLACEMENT SURGERY. POSTOPERATIVE DIAGNOSIS: FRACTURED POLYETHYLENE FAILURE WITH POLYETHYLENE DEBRIS RESULTING IN SEVERE SYNOVITIS. LOOSE FEMORAL COMPONENT WAS PRESENT. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Joint Laxity; Synovitis

    MAUDE report 16886616 ↗
  • InjuryApr 3, 2023Fracture; Naturally Worn; Adverse Event Without Identified Device or Use Problem; Fracture; Naturally Worn; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A 55 YO FEMALE, INITIAL LEFT KNEE IMPLANTED ON (B)(6) 2018, UNDERWENT A REVISION PROCEDURE ON (B)(6) 2023, APPROXIMATELY 4 YEARS 4 MONTHS POST THE INITIAL PROCEDURE. REASON FOR THE REVISION WAS NOT REPORTED. RECALLED POLY INDICATED. THERE WERE NO SURGICAL DELAYS. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO DEVICE RETURN ANTICIPATED AS THE HOSPITAL WOULD NOT ALLOW IT TO BE RETURNED.

    Patient problems: Ambulation or Postural Difficulties; Balance Problems; Discomfort; Implant Pain; Insufficient Information; Joint effusion; Osteolysis; Swelling/ Edema

    MAUDE report 16668552 ↗
  • InjuryFeb 22, 2023Naturally Worn; Naturally Worn

    AS REPORTED, THE PATIENT HAD AN INITIAL RIGHT TKA ON (B)(6) 2018. THE PATIENT WAS REVISED ON (B)(6) 2023 AS THE POLY HAD DELAMINATED. THERE WAS NO REPORTED BREAKAGE OF DEVICES OR SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO OTHER PATIENT INFORMATION/MEDICAL HISTORY REPORTED.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 16419980 ↗
  • InjuryFeb 17, 2023Fracture; Naturally Worn; Fracture; Naturally Worn

    AS REPORTED, APPROXIMATELY 5 YEARS POST OP THE INITIAL RIGHT TKA, THIS 62 Y/O FEMALE PATIENT WAS REVISED DUE TO WEAR ON IMAGING. THE ORIGINAL POLY IMPLANT IS INVOLVED IN THE RECALL. A POLY SWAP WAS PERFORMED. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO OTHER PATIENT INFORMATION/MEDICAL HISTORY REPORTED. PRODUCT NOT RETURNING - THE POLY IMPLANT HAD TO BE TAKEN TO THE LEGAL DEPARTMENT AT THE HOSPITAL FOR HOLD AND EXAMINATION. THE CORPORATE CAN FOLLOW UP WITH THE HOSPITAL ON THE STATUS OF POLY. THE PATIENT REQUESTED TO TAKE THE POLY AFTER IT WAS RELEASED FROM THE HOSPITAL.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant; Implant Pain; Swelling/ Edema

    MAUDE report 16395844 ↗
  • InjurySep 29, 2022Naturally Worn; Naturally Worn

    APPROXIMATELY 5 YRS POSTOP THE INITIAL IMPLANT, THE 77 Y/O FEMALE PATIENT PRESENTED TO THE SURGEON'S OFFICE WITH COMPLAINTS OF PAIN, STIFFNESS, AND SWELLING IN LEFT KNEE. AFTER EXAMINATION AND X-RAY OF THE LEFT TKA, THERE IS EVIDENCE OF POLY WEAR AND POOR ALIGNMENT ON THE RADIOGRAPH. THE POLY IMPLANT THAT IS CURRENTLY IN WAS SEARCHED IN THE RECALL DATABASE AND IS A PART OF THE KNEE POLY IMPLANT RECALL. THE ORIGINAL POLY IS ALSO THE SLOPE ++ STYLE. THE PATIENT WAS SCHEDULED FOR A REVISION OF LEFT TKA, A POSSIBLE POLY SWAP FOR (B)(6) 2022. THE SURGEON OPENED THE KNEE, PERFORMED A SYNOVECTOMY, EXAMINED THE IMPLANTS AND HAD QUESTIONS ON IF THE PATELLA OR ANY OF THE EXACTECH PATELLAE ARE A PART OF THE RECALL. AS THE REPRESENTATIVE, I REPLIED NO, THE KNEE, ANKLE, AD OLD HIP POLYS ARE INVOLVED IN THE RECALL. THE SURGEON EXPLAINED THAT THE PATELLA IMPLANT LOOKED POOR OR WEIRD AND ABNORMALLY WORN FOR THE LENGTH OF TIME THAT IT HAS BEEN IMPLANTED. THE SURGEON SAID HE IS NOT CONFIDENT THAT IT IS ONLY A PACKAGING ISSUE FOR THE CAUSE OF POLY FAILURES, BUT POTENTIALLY AN ISSUE WITH THE MANUFACTURING PROCESS ITSELF. THE SURGEON THEN PAID ATTENTION TO THE TIBIAL INSERT POLYETHYLENE. IT WAS REMOVED

    Patient problems: Failure of Implant; Pain; Scar Tissue; Swelling/ Edema

    MAUDE report 15503203 ↗
  • InjuryNov 8, 2021Naturally Worn; Naturally Worn

    AS REPORTED, THIS FEMALE PATIENT'S RIGHT KNEE WAS REVISED DUE TO POLY WEAR. THE POLY IS GREATER THAN 5 YEARS OLD AT THE TIME OF IMPLANTATION. NO OTHER INFORMATION AVAILABLE AT THIS TIME.

    Patient problems: Failure of Implant

    MAUDE report 12774181 ↗
  • InjuryJun 30, 2020Naturally Worn; Naturally Worn

    APPROXIMATELY 4 YEARS POSTOP THE INITIAL LEFT TKA, THIS (B)(6) Y/O FEMALE PATIENT WITH PATIENT IS 5¿4¿. (B)(6) POUNDS. BMI 37.25, WAS REVISED TO A BADLY WORN POLY INSERT. THE DEVICE WILL BE RETURNED. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.

    Patient problems: Failure of Implant

    MAUDE report 10214723 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K111400. One submission often covers many models, so check the product and lot details in each record.

  • Z-0583-2024 · Nov 21, 2023 · Open, Classified
    Exactech Optetrak Logic Knee System, Catalog Numbers: a) 02-012-47-3011, b) 02-012-47-5009, c) 02-012-47-5011, d) 02-012-65-3013

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page