Skip to content

Optetrak Logic 02-012-49-3013Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak Logic (408)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
7
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10885862174437), received between Feb 8, 2022 and May 20, 2024 — classified as 7 injuries.

The most frequently reported specific device problem is “Naturally Worn” (7 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K111400); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2021Class IIOpen, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0021-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn7
  • Unstable4
  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • Failure of Implant6
  • Joint Laxity4
  • Pain4
  • Swelling/ Edema4
  • Osteolysis3
  • Synovitis3
  • Implant Pain1
  • Joint effusion1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 20, 2024Unstable; Naturally Worn; Unstable; Naturally Worn

    AS REPORTED, APPROXIMATELY 5 YEARS AND 5 MONTHS POST THE INITIAL TKA, THE 60 YEAR OLD FEMALE PATIENT RETURNED TO THE SURGEON'S OFFICE WITH COMPLAINTS OF PAIN, SWELLING, INSTABILITY, AND DISSATISFACTION WITH THEIR TKA. THE PATIENT HAS A RECALLED POLY IMPLANT. UPON EXAMINATION, THE PATIENT WAS SCHEDULED TO HAVE A REVISION TKA POSSIBLE POLY SWAP VS FULL REVISION. THE PATIENT HAD REVISION SURGERY ON (B)(6) 2024. THERE WAS EXTENSIVE SYNOVITIS THAT WAS DEBRIDED. THERE WAS OSTEOLYSIS BEHIND THE LATERAL FEMORAL CONDYLE AND UNDER THE MEDIAL POSTERIOR TIBIA IMPLANT. BOTH THE FEMORAL COMPONENT AND THE TIBIAL COMPONENT WERE TESTED FOR LOOSENING AND SHOWED TO BE STABLE/NOT LOOSE. THE PATIENT UNDERWENT A POLY TIBIAL INSERT SWAP AND PATELLA SWAP. THERE WAS NO REPORTED BREAKAGE OF A DEVICE OR SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.

    Patient problems: Failure of Implant; Joint Laxity; Osteolysis; Pain; Swelling/ Edema; Synovitis

    MAUDE report 19350238 ↗
  • InjuryNov 16, 2023Unstable; Naturally Worn; Adverse Event Without Identified Device or Use Problem; Unstable; Naturally Worn; Adverse Event Without Identified Device or Use Problem

    AS REPORTED, THE 73 Y/O FEMALE PATIENT HAD AN ORIGINAL EXACTECH TKA DONE ON (B)(6) 2018. THE PATIENT RETURNED TO SURGEON'S OFFICE WITH COMPLAINTS OF PAIN, SWELLING, INSTABILITY, AND DISSATISFACTION WITH THEIR TKA. THE PATIENT HAS A RECALLED POLY IMPLANT. UPON EXAMINATION, THE PATIENT WAS SCHEDULED TO HAVE A REVISION. THE PATIENT HAD REVISION SURGERY ON (B)(6) 2023. THE PATIENT UNDERWENT A POLY INSERT SWAP AND PATELLA IMPLANT SWAP. PATIENT WAS REVISED TO A TRULIANT CRC TIBIAL INSERT CRC SZ 3, 15MM AND OPTETRAK 3 PEG PATELLA CEMENTED 29MM. THERE WAS NO BREAKAGE OF DEVICE OR SURGICAL DELAY/PROLONGATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. IMAGES AND X-RAYS ATTACHED . THE DEVICES ARE NOT AVAILABLE FOR EVALUATION DUE TO THE IMPLANT(S) WERE SENT TO THE LAB AND LEGAL AT THE HOSPITAL.

    Patient problems: Implant Pain; Joint Laxity; Pain; Swelling/ Edema; Synovitis

    MAUDE report 18148975 ↗
  • InjuryJul 5, 2023Unstable; Naturally Worn; Unstable; Naturally Worn

    AS REPORTED, APPROXIMATELY 6 YEARS POST OP INITIAL RIGHT TKA, THIS 76 Y/O MALE PATIENT WAS REVISED. THE PATIENT RETURNED TO SURGEONS OFFICE WITH COMPLAINTS OF PAIN, SWELLING, INSTABILITY, AND DISSATISFACTION WITH THEIR TKA. THE PATIENT HAS A RECALLED IMPLANT. UPON EXAMINATION, THE PATIENT WAS SCHEDULED TO HAVE A REVISION TKA POSSIBLE POLY SWAP VS FULL REVISION. THE PATIENT HAD REVISION SURGERY ON (B)(6) 2023. THE PATIENT UNDERWENT A TKA FULL REVISION TO ANOTHER COMPANY REVISION KNEE DUE TO A DAMAGED POLY LOCKING MECHANISM IN THE EXACTECH KNEE DUE TO POLY WEAR. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. THERE IS NO KNOWN MEDICAL HISTORY FOR THIS PATIENT. TWO X RAYS WERE PROVIDED. THE PRODUCT IS NOT BEING RETURNED DUE TO THE IMPLANT WAS SENT TO THE LAB AND LEGAL AT THE HOSPITAL.

    Patient problems: Failure of Implant; Joint Laxity; Joint effusion; Osteolysis; Pain; Swelling/ Edema; Synovitis

    MAUDE report 17262465 ↗
  • InjuryApr 10, 2023Unstable; Naturally Worn; Unstable; Naturally Worn

    AS REPORTED BY THE LEGAL TEAM, ON (B)(6) 2018, PATIENT UNDERWENT SURGERY UTILIZING RECALLED EXACTECH COMPONENT PARTS. NO PATIENT REVISION SCHEDULED. NO OTHER PATIENT INFORMATION/MEDICAL HISTORY REPORTED. ADDITIONAL INFORMATION RECEIVED INDICATES THAT PATIENT PRESENTED BACK TO SURGEON'S OFFICE WITH COMPLAINTS OF PAIN, SWELLING, INSTABILITY, AND DISSATISFACTION WITH THEIR RIGHT TKA. THE PATIENT HAS A RECALLED IMPLANT. UPON EXAMINATION, THE PATIENT WAS SCHEDULED TO HAVE A REVISION TKA POSSIBLE POLY SWAP VS FULL REVISION. THE PATIENT HAD REVISION SURGERY ON (B)(6) 2023. THE PATIENT UNDERWENT A RIGHT TKA TOTAL IMPLANT REVISION TO COMPETITORS DEVICE DUE TO A BENT/MALFUNCTIONING LOCKING MECHANISM OF THE EXACTECH TIBIAL BASEPLATE, A POLY EXCHANGE WAS NOT POSSIBLE. THE LOCKING MECHANISM WAS DAMAGED DUE TO THE POLY WEAR AND METAL-ON-METAL IMPLANT CONTACT. THERE WAS NO BREAKAGE OF A DEVICE OR SURGICAL DELAY/PROLONGATION. PATIENT LEFT THE OPERATING ROOM (OR) STABLE FOLLOWING FULL REVISION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. IMAGES RECEIVED. THE DEVICES ARE NOT AVAILABLE FOR EVALUATION DUE TO THE RECALLED POLY INSERT WAS REQUIRED TO BE SENT TO THE LAB AND LEGA

    Patient problems: Failure of Implant; Joint Laxity; Pain; Swelling/ Edema

    MAUDE report 16710475 ↗
  • InjuryOct 5, 2022Naturally Worn; Adverse Event Without Identified Device or Use Problem; Naturally Worn; Adverse Event Without Identified Device or Use Problem

    THIS FEMALE PATIENT HAS REPORTED PAIN AND SUFFERING. NO OTHER INFORMATION AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 15546865 ↗
  • InjuryApr 20, 2022Naturally Worn; Naturally Worn

    AS REPORTED, APPROXIMATELY 3.5 YEARS POST INITIAL RIGHT TKA, THIS (B)(6) MALE PATIENT HAD A TOTAL KNEE REPLACEMENT WHICH FAILED. PATIENT REASON FOR THE REVISION WAS POLY FAILURE. THERE IS LOTS OF POLY WEAR AND IT HAD A SANDPAPER TEXTURE TO IT. IT WAS DISCOLORED AND YELLOW IN SPOTS. THERE WAS OSTEOLYSIS. SURGEON SAW THE PATIENT IN CLINIC AND RECOGNIZED A FAILING COMPONENT. PATIENT REVISED TO COMPETITOR'S DEVICE. THERE WAS NO PARTS OR PIECES FELL INTO THE PATIENT WOUND SITE. THERE WAS NO SURGICAL DELAY/PROLONGATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. SALES REP WAS UNABLE TO OBTAIN PHOTOS/X-RAYS. THE DEVICE WILL BE RETURN.

    Patient problems: Failure of Implant; Osteolysis

    MAUDE report 14161293 ↗
  • InjuryFeb 8, 2022Naturally Worn; Naturally Worn

    AS REPORTED, APPROXIMATELY 5 YEARS POST INITIAL LEFT SIDE TKA, THIS (B)(6) FEMALE PATIENT EXPERIENCED PROGRESSIVE INSTABILITY OVER TIME THAT BROUGHT HER BACK TO THE CLINIC. POLY WEAR OVER TIME OF BOTH COMPARTMENTS. POSTERIOR WEAR BILATERALLY BUT WORSE MEDIALLY. IT WAS NOT PUT IN TOO TIGHT POSTERIORLY AND THE KNEE WAS WELL BALANCED. WHEN SURGEON TOOK POLY OUT AND REPLACED IT THERE WAS NO POSTERIOR TIGHTNESS. THERE WAS NO PARTS OR PIECES FELL INTO THE PATIENT WOUND SITE. THERE WAS NO SURGICAL DELAY/PROLONGATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. THE DEVICE WILL BE RETURN.

    Patient problems: Failure of Implant

    MAUDE report 13485882 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K111400. One submission often covers many models, so check the product and lot details in each record.

  • Z-0583-2024 · Nov 21, 2023 · Open, Classified
    Exactech Optetrak Logic Knee System, Catalog Numbers: a) 02-012-47-3011, b) 02-012-47-5009, c) 02-012-47-5011, d) 02-012-65-3013

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page