Skip to content

Optetrak Logic 02-012-49-5013Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak Logic (408)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10885862174635), received between Nov 10, 2021 and May 17, 2024 — classified as 5 injuries.

The most frequently reported specific device problem is “Naturally Worn” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K111400); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2021Open, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0021-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn5
  • Unstable3
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Failure of Implant5
  • Joint Laxity2
  • Pain2
  • Swelling/ Edema2
  • Implant Pain1
  • Joint effusion1
  • Osteolysis1
  • Synovitis1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 17, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 49 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 19346536 ↗
  • InjuryJan 16, 2024Unstable; Naturally Worn; Unstable; Naturally Worn

    AS REPORTED, APPROXIMATELY SIX YEARS POST INITIAL RIGHT TKA, THE 67 Y/O PATIENT RETURNED TO SURGEON'S OFFICE WITH COMPLAINTS OF PAIN, SWELLING, INSTABILITY, AND DISSATISFACTION WITH THEIR TKA. THE PATIENT HAS A RECALLED POLY IMPLANT. UPON EXAMINATION, THE PATIENT WAS SCHEDULED TO HAVE A REVISION TKA POSSIBLE POLY SWAP VS FULL REVISION. THE PATIENT UNDERWENT A POLY INSERT SWAP AND PATELLA IMPLANT SWAP. IT WAS NOTED THAT THE PATIENT HAD EXTENSIVE SYNOVITIS DUE TO POLY WEAR DEBRIS THAT REQUIRED A SYNOVECTOMY. BONE LOSS BEHIND THE LATERAL POSTERIOR/DISTAL CONDYLE REQUIRED BONE GRAFT SUBSTITUTE. PATIENT WAS REVISED TO REVISED TRULIANT CRC TIBIAL INSERT SZ 5 17MM AND OPTETRAK 3 PEG PATELLA 35MM. THE PATIENT LEFT THE OR STABLE. ALSO, THE INITIAL TIBIAL POLY INSERT OPENED FOR REPLACEMENT (17MM CRC) HAD DIFFICULTY INSERTING AND LOCKING PROPERLY. THIS IMPLANT WAS REMOVED AND A NEW 17MM CRC POLY WAS OPENED THAT DID INSERT AND LOCK APPROPRIATELY. THERE WAS NO BREAKAGE OF DEVICE OF SURGICAL DELAY/PROLONGATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. IMAGES RECEIVED. SALES REP WAS UNABLE TO OBTAIN X-RAYS. THE DEVICES ARE NOT AVAILABLE FOR EVALUATION DUE TO THE IMPLA

    Patient problems: Failure of Implant; Joint Laxity; Osteolysis; Pain; Swelling/ Edema; Synovitis

    MAUDE report 18522267 ↗
  • InjuryNov 30, 2023Unstable; Naturally Worn; Adverse Event Without Identified Device or Use Problem; Unstable; Naturally Worn; Adverse Event Without Identified Device or Use Problem

    AS REPORTED, APPROXIMATELY SIX YEARS POST INITIAL TKA, THE 72 Y/O MALE PATIENT RETURNED TO SURGEON'S OFFICE WITH COMPLAINTS OF PAIN, SWELLING, INSTABILITY, AND DISSATISFACTION WITH THEIR TKA. THE PATIENT HAS A RECALLED POLY IMPLANT. UPON EXAMINATION, THE PATIENT WAS SCHEDULED TO HAVE A REVISION TKA POSSIBLE POLY SWAP VS FULL REVISION. THE PATIENT UNDERWENT A FULL KNEE IMPLANT REMOVAL AND REVISED TO COMPETITOR'S DEVICES. THERE WAS NO BREAKAGE OF DEVICE. THERE WAS A 30-45 MINUTES SURGICAL DELAY/PROLONGATION. THERE WAS NO ADVERSE EVENTS AS A RESULT. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. PHOTOS RECEIVED. SALES REP WAS UNABLE TO OBTAIN X-RAYS. THE DEVICES ARE NOT AVAILABLE FOR EVALUATION DUE TO THE IMPLANT(S) WERE SENT TO THE LAB AND LEGAL AT THE HOSPITAL.

    Patient problems: Failure of Implant; Joint Laxity; Pain; Swelling/ Edema

    MAUDE report 18241079 ↗
  • InjuryJul 17, 2023Unstable; Naturally Worn; Unstable; Naturally Worn

    AS REPORTED VIA LEGAL DOCUMENTATION THE 59 Y/O MALE PATIENT HAD A RIGHT KNEE REPLACEMENT ON (B)(6)2015. APPROXIMATELY 7 YEARS AND 2 MONTHS AFTER THE INITIAL PROCEDURE THE PATIENT HAD A RIGHT KNEE REVISION ON (B)(6)2022. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. REVISION OP REPORT ON (B)(6)2022 DIAGNOSIS: FAILED RIGHT KNEE ARTHROPLASTY DUE TO POLYETHYLENE WEAR AND FLEXION INSTABILITY, FEMORAL COMPONENT FOUND TO BE THE DEBONDED FROM THE CEMENT. "COPIOUS STRAW-COLORED SYNOVIAL FLUID WAS OBTAINED. I THEN SENT SYNOVIAL SPECIMENS FOR CULTURE AS WELL AS FROZEN PATHOLOGY. A GENEROUS MEDIAL PEEL WAS PERFORMED TO ASSIST WITH EXPOSURE. A PARTIAL SYNOVECTOMY WAS PERFORMED TO REESTABLISH THE MEDIAL AND LATERAL GUTTERS. NEXT, THE POLYETHYLENE WAS REMOVED WITH AN OSTEOTOME. THERE WAS NO GROSS EVIDENCE OF POLYETHYLENE WEAR. THE FEMUR WAS THEN EASILY ABLE TO BE DISIMPACTED. THE IMPLANT WAS FOUND TO BE COMPLETELY DE BONDED FROM THE UNDERLYING CEMENT. THERE IS NO CEMENT ATTACHED TO THE FEMORAL COMPONENT. THE REMAINING CEMENT WAS REMOVED PIECEMEAL WIT

    Patient problems: Failure of Implant; Implant Pain; Joint effusion

    MAUDE report 17334502 ↗
  • InjuryNov 10, 2021Naturally Worn; Naturally Worn

    AS REPORTED, THIS MALE PATIENT'S LEFT KNEE WAS REVISED DUE TO POLY WEAR. FAILURE TO BALANCE FLEXION GAP. POLY IS AGED GREATER THAN 4 YEARS. TIBIAL CUT IS IN SIGNIFICANT VARUS. NO OTHER INFORMATION AVAILABLE AT THIS TIME.

    Patient problems: Failure of Implant

    MAUDE report 12788530 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K111400. One submission often covers many models, so check the product and lot details in each record.

  • Z-0583-2024 · Nov 21, 2023 · Open, Classified
    Exactech Optetrak Logic Knee System, Catalog Numbers: a) 02-012-47-3011, b) 02-012-47-5009, c) 02-012-47-5011, d) 02-012-65-3013

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page