Skip to content

Optetrak Logic 02-012-47-6009Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak Logic (408)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10885862174642), received between Feb 11, 2021 and Dec 30, 2025 — classified as 3 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2021Class IIOpen, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0021-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Device Problem Information1
  • Naturally Worn1

Patient problems

  • Failure of Implant1
  • Insufficient Information1
  • Osteolysis1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 30, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT A MALE PATIENT WHO HAD A LEFT TKA, UNDERWENT A REVISION PROCEDURE APPROXIMATELY 6 YEARS 4 MONTHS POST THE INITIAL PROCEDURE. PATIENT PRESENTED FOR A PATELLA RESURFACING PROCEDURE. DURING THE PROCEDURE THE SURGEON ALSO DECIDED TO EXTRACT AND REPLACE THE TIBIAL POLY LINER AS IT WAS PART OF THE TKR POLY LINER RECALL. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO X-RAYS WERE PROVIDED. THE EXPLANTED DEVICES ARE NOT AVAILABLE FOR RETURN. THEY WERE DISCARDED. NO DEVICE IMAGES WERE PROVIDED. NO FURTHER INFORMATION.

    Patient problems: Insufficient Information

    MAUDE report 23929905 ↗
  • InjurySep 6, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THIS MALE PATIENT'S LEFT KNEE WAS REVISED DUE TO SUSPECTED IMPLANT LOOSENING IN FEMUR AND POLY WEAR/ LYSIS. EVERYTHING WAS REMOVED. PATIENT WAS NOT REVISED TO EXACTECH DEVICES. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. PRODUCT NOT RETURNING: DISCARDED AT THE HOSPITAL.

    Patient problems: Osteolysis

    MAUDE report 20157408 ↗
  • InjuryFeb 11, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED, APPROXIMATELY 1.5 YEARS POSTOP THE INITIAL IMPLANT, THIS (B)(6) Y/O MALE PATIENT'S RIGHT PATELLA FACETECTOMY WAS COMPLETED. A REVISION FOR LINER EXCHANGE PERFORMED DUE TO KNEE BEING IN SLIGHT VALGUS STILL TO RELIEVE AP INSTABILITY. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. THE ORIGINAL LINER WAS DISPOSED OF AT HOSPITAL.

    Patient problems: Failure of Implant

    MAUDE report 11315150 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page