Skip to content

Optetrak Logic 02-012-47-6011Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak Logic (408)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885862174659), received between Jan 8, 2025 and Jan 8, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Naturally Worn” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2021Class IIOpen, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0021-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn1

Patient problems

  • Osteolysis1
  • Pain1
  • Swelling/ Edema1
  • Synovitis1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 8, 2025Naturally Worn; Naturally Worn

    APPROXIMATELY 5 YEARS AND 8 MONTHS AFTER INITIAL IMPLANTATION OF LEFT TOTAL KNEE ARTHROPLASTY IN (B)(6) 2016, IT WAS REPORTED VIA LEGAL NOTIFICATION THAT A PATIENT EXPERIENCED PROGRESSIVE/INCREASED PAIN AND SWELLING ASSOCIATED WITH ASEPTIC LOOSENING. THE PATIENT WAS SURGICALLY REVISED IN (B)(6) 2022, ADDITIONALLY NOTING RADIOGRAPHS SHOWED LOOSENING ABOUT THE FEMORAL COMPONENT. THERE WAS AN ABUNDANT SYNOVIAL INFLAMMATION PRESENT. INSPECTION OF THE COMPONENTS SHOWED THAT THERE WAS EVIDENCE OF WEAR DEBRIS ON THE MEDIAL FEMUR POLYETHYLENE POSTERO-MEDIALLY. THERE WAS OBVIOUS LOOSING OF THE FEMORAL COMPONENT. THE TIBIAL COMPONENT WAS SECURELY FIXED. THE PATELLA WAS IN APPROPRIATE POSITION. THE KNEE WAS STABLE AFTER REVISION. NO COMPLICATIONS DURING THE SURGICAL PROCEDURE. THERE ARE NO RADIOGRAPHIC IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Osteolysis; Pain; Swelling/ Edema; Synovitis

    MAUDE report 21106690 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page