Optetrak Logic 02-010-01-0315Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak Logic (408)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10885862175854), received between Dec 23, 2021 and Aug 16, 2023 — classified as 2 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K093360); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth2
  • Insufficient Device Problem Information1

Patient problems

  • Failure of Implant2
  • Ambulation or Postural Difficulties1
  • Aseptic loosening1
  • Joint effusion1
  • Necrosis1
  • Pain1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 16, 2023Insufficient Device Problem Information; Loosening of Implant Not Related to Bone-Ingrowth; Insufficient Device Problem Information; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A RIGHT KNEE REPLACEMENT ON (B)(6) 2013. APPROXIMATELY 6 YEARS AND 6 MONTHS AFTER THE INITIAL PROCEDURE THE PATIENT HAD A RIGHT KNEE REVISION ON (B)(6) 2019. THE PATIENT HAS SUFFERED THE FOLLOWING INJURIES AND COMPLICATIONS: PAIN, DIFFICULTY WITH EVERYDAY ACTIVITIES, TROUBLE WALKING. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE PATIENT HAS FILED A SHORT-FORM COMPLAINT IN A MULTIDISTRICT LITIGATION TITLED IN RE: EXACTECH POLYETHYLENE ORTHOPEDIC PRODUCTS LIABILITY LITIGATION, MDL NO. 3044, AND PENDING IN THE EASTERN DISTRICT OF NEW YORK. PER THE TRANSFER ORDER CREATING THIS MULTIDISTRICT LITIGATION, ¿PLAINTIFFS¿ALLEGE THAT THEIR KNEE OR HIP REPLACEMENT DEVICES¿FAILED PREMATURELY BECAUSE OF DEGRADATION OF THE DEVICE¿S POLYETHYLENE COMPONENT, WHICH RESULTED IN THE PREMATURE REMOVAL (OR PLANNED REMOVAL) OF THE PROSTHESIS AT ISSUE.¿ BECAUSE THE PATIENT HAS FILED A SUIT IN THIS MULTIDISTRICT LITIGATION, THE PATIENT AP

    Patient problems: Ambulation or Postural Difficulties; Aseptic loosening; Failure of Implant; Joint effusion; Necrosis; Pain

    MAUDE report 17556464 ↗
  • InjuryDec 23, 2021Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED, A (B)(6) FEMALE PATIENT¿S REASON FOR SURGERY WAS THE FAILURE OF THE IMPLANTS. SURGEON THOUGHT THE SURGERY WAS JUST GOING TO BE A POLY EXCHANGE, HOWEVER ONCE THE KNEE WAS OPENED UP, IT WAS DISCOVERED THE FEMUR WAS LOOSE AS WELL AND THAT THERE WAS SIGNIFICANT BONE LOSE. IT WAS REVISED TO A SIZE CC FEMUR, 4MM OFFSET, 10X120 STEM, 10 AND 5 POST AUG BLOCK, TWO 5 DISTAL AUG BLOCKS AND A 1.5 10MM PSC POLY, THE TIBIAL TRAY WAS NOT TOUCHED AS IT WAS WELL FIXED. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. THE DEVICE WILL BE RETURN.

    Patient problems: Failure of Implant

    MAUDE report 13081381 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K093360. One submission often covers many models, so check the product and lot details in each record.

  • Z-0021-2022 · Aug 30, 2021 · Open, Classified
    OPTETRAK LOGIC Polyethylene Tibial Inserts, labeled as the following: a. OPTETRAK Logic PS Tibial Inserts, Product Line 02-012-35-XXXX; b. OPTETRAK Logic PSC Tibial Inserts,
  • Z-2310-2012 · Aug 19, 2011 · Terminated
    OPTETRAK LOGIC TIBIAL INSERT POSTERIOR STABILIZED***Match Component Colors 6 YELLOW, 11mm, Use with LOGIC Femoral and Tibial Components***REF 02-012-35-6011. There are three produ

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices