Equinoxe 320-02-42Recalls & reported problems
Exactech, Inc. · Class 2 · Implant
All models of Equinoxe (305)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10885862186768), received between May 22, 2019 and Oct 15, 2025 — classified as 10 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem4
- Device Dislodged or Dislocated3
- Naturally Worn3
Patient problems
- Joint Dislocation4
- Metal Related Pathology2
- Calcium Deposits/Calcification1
- Failure of Implant1
- Muscle/Tendon Damage1
- No Information1
- Osteolysis1
- Pain1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryOct 15, 2025Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL REVERSE TOTAL SHOULDER REPLACEMENT ON THE RIGHT SIDE. SUBSEQUENTLY, THE PATIENT EXPERIENCED THE DISASSEMBLY AND MIGRATION OF THE POLYETHYLENE COMPONENT. AS A RESULT, APPROXIMATELY 2 YEARS AFTER INITIAL REPLACEMENT, THE PATIENT UNDERWENT A RIGHT SHOULDER REVISION. NO FURTHER IMPACT TO THE PATIENT WAS REPORTED. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: Muscle/Tendon Damage
MAUDE report 23291268 ↗InjuryJun 9, 2025Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL REVERSE SHOULDER REPLACEMENT ON THE RIGHT SIDE. SUBSEQUENTLY, THE PATIENT NOTICED THE SHOULDER FELT DIFFERENT IN THE PAST MONTH. IT IS UNKNOWN IF THE PATIENT EXPERIENCED AN INJURY OR FALL PRIOR. IMAGING REVEALED THE HUMERAL LINER HAD DISSOCIATED AND WAS NO LONGER SITTING IN THE TRAY. THE SHOULDER WAS STILL REDUCED, WITH THE TRAY ARTICULATING WITH THE GLENOSPHERE. IT IS UNKNOWN HOW LONG THE HUMERAL LINER HAD BEEN DISSOCIATED. AS A RESULT, APPROXIMATELY 5 YEARS AND 6 MONTHS POST-SURGERY, THE PATIENT UNDERWENT A REVISION. THE HUMERAL STEM WAS FOUND TO BE WELL FIXED AND WAS RETAINED. THE HUMERAL LINER HAD MIGRATED POSTERIOR TO THE GLENOSPHERE, AS DISCOVERED FOLLOWING REMOVAL OF THE GLENOSPHERE. THE AREA WAS EXTENSIVELY DEBRIDED AND IRRIGATED WITH ANTIMICROBIAL WOUND SOLUTION AND SALINE. TISSUE CULTURES WERE TAKEN DUE TO OBSERVATION OF METALLOSIS, THERE WAS NO EVIDENCE OF INFECTION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO FURTHER IMPACT TO THE PATIENT WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THIS IS REPORT 3 OF 3 FOR THIS EVENT/PATIENT.
Patient problems: Metal Related Pathology
MAUDE report 22169301 ↗InjuryFeb 10, 2025Naturally Worn; Naturally Worn
AS REPORTED, APPROXIMATELY 5 YEARS AND 9 MONTHS POST INITIAL LEFT TOTAL SHOULDER ARTHROPLASTY (TSA) FOR SECONDARY ARTHRITIS AND A MASSIVE ROTATOR CUFF TEAR AND APPROXIMATELY 2 YEARS AND 5 MONTHS POST REVISION FOR PROSTHETIC DISLOCATION CAUSED BY DISASSEMBLY OF THE POLYETHYLENE INSERT, A SECOND REVISION WAS PERFORMED DUE TO PAIN REPORTED BY THE PATIENT AT WHICH TIME THE GLENOSPHERE, HUMERAL LINER, AND HUMERAL ADAPTER TRAY WERE REPLACED. THE HUMERAL LINER WAS CHANGED TO A CONSTRAINED HUMERAL LINER. DESPITE THE REVISIONS, THE INSERT CONTINUED TO DISASSEMBLE, AND THE INNER JOINT EXHIBITED AN UNUSUAL BROWNISH DISCOLORATION INDICATING METALLOSIS. EVENT AND PATIENT OUTCOME WERE NOT PROVIDED. PRODUCTS RETURNED.
Patient problems: Joint Dislocation; Metal Related Pathology; Pain
MAUDE report 21344621 ↗InjuryNov 27, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
AS REPORTED, APPROXIMATELY 28 WEEKS POST INITIAL LEFT REVERSE TOTAL SHOULDER ARTHROPLASTY (TSA), THE SHOULDER DISLOCATED WHILE THE PATIENT WAS ON HOLIDAYS AND AS A RESULT BECAME INFECTED. A REVISION WAS PERFORMED AT WHICH TIME THE SURGEON PUT A LARGER GLENOSPHERE AND PERFORMED A WASH OUT. THE EVENT WAS NOT RELATED TO THE BREAKAGE OF A DEVICE AND DID NOT LEAD TO A SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.
Patient problems: Joint Dislocation; Unspecified Infection
MAUDE report 20799371 ↗InjuryJan 24, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
AS REPORTED, THIS (B)(6) MALE PATIENT REQUIRED A REVISION ABOUT 5 MONTHS POST OP DUE TO DISLOCATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. DEVICES WILL NOT BE RETURNING.
Patient problems: Joint Dislocation
MAUDE report 13333699 ↗InjuryJun 26, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED, APPROXIMATELY 26 MONTHS AFTER THE INITIAL IMPLANT, THIS (B)(6) Y/O MALE PATIENT PRESENTED TO SURGEON WITH SOME SHOULDER INSTABILITY IN THEIR REVERSE SHOULDER AFTER OVER A YEAR OF GOOD FUNCTION AND STABILITY. SURGEON WAS ABLE TO DISLOCATE IN OFFICE AND ELECTED TO UPSIDE COMPONENTS SURGICALLY TO ACHIEVE STABILITY. SURGEON UPSIZED THE 42MM GLENOSPHERE TO A 46MM. ALSO REMOVED A 42 X 2.5MM LINER AND REPLACED WITH A 46 X 2.5MM CONSTRAINED LINER. SURGEON EXPECTS A GREAT OUTCOME AND PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. DEVICES WILL NOT RETURN DUE TO HOSPITAL POLICY.
Patient problems: Failure of Implant
MAUDE report 10199548 ↗InjuryNov 19, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED, APPROXIMATELY 1.5 YEARS POSTOP THE INITIAL IMPLANT, THIS 66 Y/O MALE PATIENT WAS REVISED IN (B)(6) 2019. REASON FOR REVISION WAS NOT REPORTED. EVENT DISCOVERED IN A RECENT COMPLAINT. DEVICES ARE NOT RETURNING. NO OTHER INFORMATION IS PATIENT CONDITIONS.
Patient problems: No Information
MAUDE report 9340392 ↗InjuryNov 15, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THIS PATIENT HAD HIS PRIMARY SURGERY IN OF (B)(6) 2019, THEN HE WAS DISLOCATING SO THE DOCTOR REVISED HIM. HE PULLED OUT THE STEM, TRAY, AND LINER. HE ALSO REPLACED THE GLENOSPHERE. THE +10 TRAY DISASSOCIATED FROM THE STEM. THE SCREW ON THE DISASSOCIATED TRAY WAS NOT BROKEN OR STRIPPED.
Patient problems: Joint Dislocation
MAUDE report 9328136 ↗InjuryJul 8, 2019Adverse Event Without Identified Device or Use Problem
DOCTOR REVISED THE SHOULDER BUT BASED ON THE BONE LOSS, ORIGINAL POSITIONING OF THE IMPLANT, AND POTENTIAL INFECTION HE WAS UNABLE TO REPAIR THE DAMAGE. NO IMPLANTS WERE PLACED IN THE PATIENT.
Patient problems: Osteolysis
MAUDE report 8767837 ↗InjuryMay 22, 2019Device Dislodged or Dislocated
REVISION PERFORMED ON (B)(6) 2019. THIS SHOULDER WAS DISLOCATING DUE TO CALCIFICATION OF THE DELTOID WHICH KEPT PUSHING THE HUMERUS ANTERIOR AND DISLOCATING THE SHOULDER.
Patient problems: Calcium Deposits/Calcification
MAUDE report 8633198 ↗
Other Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented devices
- Equinoxe 320-42-00Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented3 recalls107 reports
- Equinoxe 320-38-00Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented3 recalls107 reports
- Equinoxe 320-42-03Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls87 reports
- Equinoxe 320-38-03Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls43 reports
- Equinoxe 320-42-13Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls41 reports
- Equinoxe 320-10-00Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented132 reports
- Equinoxe 320-36-00Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls27 reports
- Equinoxe 320-42-10Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls14 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page