Novation 130-40-54Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Novation (184)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
2

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10885862207104), received between Nov 7, 2022 and Jun 3, 2024 — classified as 5 injuries.

The most frequently reported specific device problem is “Naturally Worn” (5 reports).

2 FDA recall records name this device's identifier, 2 still open.

5 other recalls cover devices cleared under the same FDA submission (K121392); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 11, 2022Class IIOpen, Classified

    Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.

    Z-1732-2022 · Root cause: Process control · Recall details · FDA record ↗

  2. Jun 29, 2021Class IIOpen, Classified

    Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.

    Z-2119-2021 · Root cause: Unknown/Undetermined by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn5
  • Insufficient Device Problem Information2
  • Adverse Event Without Identified Device or Use Problem1
  • Device Dislodged or Dislocated1
  • Unstable1

Patient problems

  • Failure of Implant3
  • Insufficient Information2
  • Osteolysis2
  • Pain2
  • Ambulation or Postural Difficulties1
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Bone Fracture(s)1
  • Foreign Body Reaction1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 3, 2024Unstable; Device Dislodged or Dislocated; Naturally Worn; Unstable; Device Dislodged or Dislocated; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 45 MONTHS AFTER A LEFT TOTAL HIP REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PAIN, POPPING, INSTABILITY, DISLOCATING AND COMPLETE NUMBNESS. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Implant Pain; Joint Dislocation; Joint Laxity; Numbness; Pain

    MAUDE report 19453454 ↗
  • InjuryJan 10, 2024Naturally Worn; Naturally Worn

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A LEFT HIP REVISION ON (B)(6) 2018. APPROXIMATELY 4 YEARS AND 11 MONTHS AFTER THE FIRST REVISION THE PATIENT HAD A SECOND LEFT HIP REVISION ON (B)(6) 2023. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. REVISION OP REPORT ON (B)(6) 2023. DIAGNOSIS: FAILED LEFT TOTAL HIP REPLACEMENT. THE PATIENT HAD EXTENSIVE SCARRING. THE PATIENT ENCOUNTERED A HUGE ADVERSE LOCAL TISSUE REACTION AND A LARGE FLUID COLLECTION AND A DARK GRAY TROCHANTERIC BURSITIS THAT HAD TO BE DEBRIDED. THERE WAS AN OPENING BREAKING THROUGH TO HIS GREATER TROCHANTER THAT THE SURGEON THOUGHT CONTRIBUTED TO TROCHANTER WAS FRACTURED. THIS HAD TO ALL BE CURETTAGED AND BONE GRAFTED. THE ACETABULAR COMPONENT ALSO HAD OSTEOLYSIS THAT HAD REQUIRED EXTENSIVE CURRATAGE AND BONE GRAFTING.

    Patient problems: Bone Fracture(s); Failure of Implant; Foreign Body Reaction; Joint effusion; Osteolysis; Scar Tissue

    MAUDE report 18482904 ↗
  • InjuryNov 7, 2023Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A RIGHT HIP REPLACEMENT ON (B)(6) 2016. APPROXIMATELY 6 YEARS AND 9 MONTHS AFTER THE INITIAL PROCEDURE THE PATIENT HAD A RIGHT HIP REVISION ON (B)(6) 2023. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE PATIENT HAS FILED A SHORT-FORM COMPLAINT IN A COORDINATED ACTION IN (B)(6) WITH MASTER CASE NO. (B)(4). THE CONSOLIDATED LONG FORM COMPLAINT THAT APPLIES TO CASES FILED IN THIS COORDINATED ACTION ALLEGES THAT PATIENTS FILING SUITS IN THIS COORDINATED ACTION WERE REQUIRED ¿TO UNDERGO REVISION SURGERIES DUE TO SEVERE, PAIN, SWELLING, AND INSTABILITY¿ DUE TO ¿WEAR OF THE POLYETHYLENE COMPONENTS AND RESULTING COMPONENT LOOSENING AND/OR OTHER FAILURE FAILURES CAUSING SERIOUS COMPLICATIONS INCLUDING TISSUE DAMAGE, OSTEOLYSIS, PERMANENT BONE LOSS, AND OTHER INJURIES.¿ BECAUSE THE PATIENT HAS FILED A SHORT-FORM COMPLAINT IN THIS COORDINATED ACTION, THE PATIENT APPEARS TO ALLEGE THAT THE PATIENT WAS INJURED AS A RESULT OF

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 18082751 ↗
  • InjuryAug 4, 2023Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information

    AS REPORTED VIA LEGAL DOCUMENTATION, A PATIENT HAD RIGHT HIP REPLACEMENT ON (B)(6) 2018. THEY UNDERWENT RIGHT HIP REVISION SURGERY ON (B)(6) 2023, APPROXIMATELY 4 YEARS 7 MONTHS POST PRIMARY PROCEDURE. THE PATIENT CLAIMS TO HAVE REQUIRED REVISION SURGERY FOR ISSUES INCLUDING BUT NOT LIMITED TO POLYETHYLENE WEAR, BONE LOSS, OSTEOLYSIS, INSTABILITY AND/OR COMPONENT LOOSENING. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 17458455 ↗
  • InjuryNov 7, 2022Naturally Worn; Adverse Event Without Identified Device or Use Problem; Naturally Worn; Adverse Event Without Identified Device or Use Problem

    AS REPORTED BY THE LEGAL DEPARTMENT, APPROXIMATELY 4 YEARS POST OP THE INITIAL RIGHT THA, THIS MALE PATIENT WAS REVISED. A BILATERAL HIP X-RAY REVEALED THE SOCKET AND SPACER ON THE RIGHT HIP HAD DISINTEGRATED THE BONE AROUND THE SCREW HOLDING HIS RIGHT HIP TOGETHER. THE X-RAY OF THE LEFT HIP WAS FINE.

    Patient problems: Ambulation or Postural Difficulties; Osteolysis; Pain

    MAUDE report 15742763 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K121392. One submission often covers many models, so check the product and lot details in each record.

  • Z-2118-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-36-XX, Acumatch GXL 0 Degree Liner, 36mm; and Novation GXL Liner, Neutral,
  • Z-2133-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-32-XX, Acumatch GXL 15 Degree Liner, 32mm; Novation GXL Liner, Lipped Ant,
  • Z-2129-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-40-XX, Novation GXL Liner, +5mm Lateralized, 40mm. Orthopedic hip implant
  • Z-2123-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-40-XX, Novation GXL Liner, Lipped Ant, 40mm; Novation GXL Liner, Lipped, 40
  • Z-2117-2021 · Jun 29, 2021 · Open, Classified
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-32-XX, Acumatch GXL 0 Degree Liner, 32mm; and Novation GXL Liner, Neutral,

Other Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate devices

All Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate devices →