Skip to content

EXACTECHGPS I00020Recalls & reported problems

Exactech, Inc. · STATION · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10885862252821), received between May 21, 2019 and Nov 15, 2023 — classified as 2 injuries.

The most frequently reported specific device problem is “Computer Operating System Problem” (1 report).

1 other recall cover devices cleared under the same FDA submission (K100742, K152764); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Computer Operating System Problem1
  • Contamination /Decontamination Problem1
  • Display or Visual Feedback Problem1

Patient problems

  • Bone Fracture(s)1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 15, 2023Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT DURING A SHOULDER PROCEDURE, A ¿STRANGE¿ PHENOMENON WITH GPS UNIT SERIAL (B)(6) OCCURRED WHERE NEITHER THE REPRESENTATIVE NOR THE SALES MANAGER COULD FIND AN EXPLANATION. THE SURGEON FOUND A DISCREPANCY BETWEEN THE PLANNED CENTRAL HOLE / PEG POSITION, AND WHAT WAS DETERMINED INTRA-OPERATIVELY. THE GPS UNIT SEEMS TO HAVE HAD AN ISSUE WITH ACCURACY. THEY WENT BACK TO THE ACQUISITION SCREEN AFTER THE SURGEON SAID HE PERFORATED THE FAULT AND WAS QUESTIONING THE ACCURACY OF THE PILOT TIP AND THAT IT CAUSED THE ERROR, SO THE REP HAD THE SURGEON GO BACK AND PICK A FEW POINTS TO SHOW THEY WERE GOOD. THE SURGEON WAS STILL CONVINCED THEY WERE OFF AND HE RE-DID THE CENTRAL PEG WHERE HE FELT WAS BETTER. FROM THE REPS MEMORY, THE SURGEON USED THE SCANS AND VISUALIZATION TO DETERMINE A REAMING POSITION THAT BETTER FIT THE IMPLANT IN THE GLENOID TO AVOID PERFORATION OF THE PEG. AFTER REPOSITIONING THE REAMER TO A POSITION, HE FELT BETTER MATCHED WHAT WAS REQUIRED, THE BASEPLATE WAS INSERTED, AND THE SURGEON WAS HAPPY WITH THE FINAL OUTCOME. THE PERFORATION OF THE BONE WAS RESOLVED BY REPOSITIONING THE K-WIRE/IMPLANT TO A BETTER POSITION, TO POSITION THE IMPLANT INSIDE THE GLE

    Patient problems: Bone Fracture(s)

    MAUDE report 18144001 ↗
  • InjuryMay 21, 2019Contamination /Decontamination Problem; Computer Operating System Problem

    DURING THE SECOND GPS+ CASE OF THE DAY, DR. (B)(6) EXPERIENCED THE FOLLOWING: TRACKER ACQUISITION WAS A PROBLEM. INTERESTINGLY THE F TRACKER. WOULD NOT COMPLETE ACQUISITION THAT LED TO DECISION TO REBOOT BECAUSE IT WOULD NOT RE-ACQUIRE. ENDED UP REBOOTING COMPUTER AND REMOVING BATTERIES, RE-ACQUIRING TRACKERS. HIP CENTER - ACQUIRED WITHOUT DIFFICULTY. IM GUIDED - PINNING WITHOUT DIFFICULTY. G - SLOT GUIDE - READ 6 DEGREES VARUS. HAD MORE DIFFICULTY THAT USUALLY ACQUIRING ACCURACY. THE PATIENT IS SHORT - 4¿ 11 ¿ AND DEPTH OF RESECTION SEEMED REASONABLE. THE SPACER BLOCKS CONFIRMED THAT THE CUT WAS WAY OFF. CONVERTED TO CC FEMUR AND OLD SCHOOL - ALIGNMENT ROD AND BOVIE CORD - FIXED THE SOFT TISSUE BALANCE AND PROBLEM. AFTER DISCUSSION WITH THE DOCTOR, IT WAS DETERMINED THAT THE LIKELY CAUSE OF THE EVENT WAS THE "SMUDGE" ON THE CAMERA OF THE GPS UNIT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8627240 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K100742, K152764. One submission often covers many models, so check the product and lot details in each record.

  • Z-1149-2020 · Sep 22, 2019 · Terminated
    Blue Ortho TKA Pro v2.1 software installed on the Exactech GPS system -indicated for stereotaxic surgery in Knee Arthroplasty Catalog Reference: L00002

Other Neurological Stereotaxic Instrument devices

All Neurological Stereotaxic Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page