Optetrak Logic 02-010-06-0340Recalls & reported problems
Exactech, Inc. · Class 2 · Implant
All models of Optetrak Logic (408)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10885862265722), received between Mar 2, 2020 and Nov 7, 2024 — classified as 5 injuries.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (4 reports).
4 other recalls cover devices cleared under the same FDA submission (K150890); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth4
- Adverse Event Without Identified Device or Use Problem1
- Unstable1
Patient problems
- Failure of Implant4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Inadequate Osseointegration1
- Joint Laxity1
- Joint effusion1
- Osteolysis1
- Pain1
- Synovitis1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryNov 7, 2024Unstable; Loosening of Implant Not Related to Bone-Ingrowth; Unstable; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT APPROXIMATELY 75 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT HAS EXPERIENCED PROSTHESIS WEAR, EFFUSION WITH SIGNIFICANT LAXITY, SEVERE OSTEOLYSIS, INSTABILITY, AND SYNOVITIS. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Joint Laxity; Joint effusion; Osteolysis; Synovitis
MAUDE report 20631267 ↗InjuryApr 12, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
LEGAL CASE - (B)(6) (B)(4). RELATED CASE (B)(4) (B)(4). EMAIL FROM LEGAL: PER A REPORT FROM THE LEGAL DEPARTMENT, THE PATIENT HAD AN INITIAL RIGHT KNEE IMPLANTED ON (B)(6) 2013. THE PATIENT HAD TWO REVISIONS OF THE RIGHT KNEE (B)(6) 21 AND (B)(6) 22 ***ALL 3 OP REPORTS ARE ATTACHED. OP REPORT (B)(6) 2022. PRE & POST OP DIAGNOSIS: FAILED TOTAL RIGHT KNEE REPLACEMENT INDICATIONS: PATIENT HAD SEVERE PAIN DUE TO FAILED ASEPTIC LOOSENING OF TOTAL RIGHT KNEE REPLACEMENT. PT HAS FAILED CONSERVATIVE TREATMENT AND WISHED TO UNDERGO OPERATIVE INTERVENTION. RISKS, BENEFITS, AND ALTERNATIVES DISCUSSED WITH PATIENT. DESPITE POTENTIAL RISKS, THEY WISHED TO PROCEED WITH THE OPERATIVE INTERVENTION. INTRAOPERATIVE FINDINGS: LOOSE FEMORAL IMPLANT - REMOVED EASILY WITH MINIMAL BONE LOSS BUT CONCERNED FOR BONE QUALITY DISTAL FEMUR. I ELECTED TO SUE FEMORAL CONE. TIBIAL IMPLANT WAS VERY SOLID AND WELL FIXED - BACKSLAP DEVICE AND OSTEOTOMES WERE USED BUT STEM WOULD NOT BUDGE - CORTICAL WINDOW DISTAL TO THE STEM ALLOWED ME TO HIT THE STEM TIP AND KNOCK IT OUT RETROGRADE. NO DEVICE RETURN ANTICIPATED DUE TO THIS BEING A LEGAL CASE. EBI INITIAL SURGERY ATTACHED. HISTORICAL RECORD & SMARTSOLVE SEARCHED FOR
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant; Pain
MAUDE report 16725637 ↗InjuryNov 17, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AS REPORTED, SURGEON REVISED THE FEMORAL COMPONENTS OF A LOGIC CC KNEE THAT HE IMPLANTED ON (B)(6) 2017. DURING THE CASE SURGEON NOTED INCOMPLETE FIXATION OF THE FEMORAL DEVICE. THERE WAS NO BREAKAGE OF A DEVICE OR SURGICAL DELAY/PROLONGATION. PATIENT WAS REVISED TO EXACTECH COMPONENTS. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. IMAGES RECEIVED. THE DEVICES ARE NOT AVAILABLE FOR EVALUATION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 15820614 ↗InjuryOct 21, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED BY THE EXACTECH KNEE REPLACEMENT STUDY, APPROXIMATELY EIGHT MONTHS POST TKA, THE PATIENT EXPERIENCED PAIN AND INSTABILITY. ON (B)(6) 2017, PATIENT WAS SQUATTING DOWN AND FELT RT. KNEE POP FORWARD AND DISLOCATED. PATIENT'S KNEE WAS MANIPULATED WHILE HE WAS IN (B)(6) AND WAS REDUCED. ON (B)(6) 2018, PATIENT CONTINUED TO HAVE INSTABILITY AND IN (B)(6) 2018 THE KNEE WAS GETTING WORSE BY GIVING AWAY AND PAINFUL. DR. (B)(6) SCHEDULE A REVISION (B)(6) 2018. THE EVENT IS POSSIBLY RELATED TO THE DEVICE BUT UNLIKELY RELATED TO THE PROCEDURE. NO FURTHER INFORMATION AVAILABLE, IT APPEARS THE PATIENT LEFT THE STUDY.
Patient problems: Failure of Implant
MAUDE report 12674149 ↗InjuryMar 2, 2020Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
AS REPORTED BY THE EXACTECH KNEE REPLACEMENT STUDY, APPROXIMATELY 3.5 YEARS POSTOP IMPLANTATION, THIS (B)(6) Y/O MALE PATIENT, WHO WEIGHS (B)(6) LBS., WAS REVISED DUE TO ¿FAILED RIGHT TKA, ASEPTIC LOOSENING¿. NO OTHER INFORMATION AVAILABLE. TREATMENT IS NOTED AS RESOLVED. THE EVENT IS DEFINITELY RELATED TO THE DEVICE AND UNLIKELY RELATED TO THE PROCEDURE. THIS EVENT REPORT WAS RECEIVED THROUGH CLINICAL DATA COLLECTION ACTIVITIES.
Patient problems: Failure of Implant
MAUDE report 9775498 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K150890. One submission often covers many models, so check the product and lot details in each record.
- Z-0583-2024 · Nov 21, 2023 · Open, ClassifiedExactech Optetrak Logic Knee System, Catalog Numbers: a) 02-012-47-3011, b) 02-012-47-5009, c) 02-012-47-5011, d) 02-012-65-3013
- Z-1364-2020 · Jan 27, 2020 · Open, ClassifiedTruliant Splined Stem Extension, 18mm x 120mm - Product Usage: The devices are used in a cemented total knee prosthesis system, which provides femoral components, tibial inserts, s
- Z-3209-2018 · May 8, 2018 · Open, ClassifiedOptetrak Logic CC Stem Extension, 22mm x 160mm The OPTETRAK Logic CC Total Knee System is indicated for use in skeletally mature individuals undergoing primary surgery for total k
- Z-2704-2017 · Apr 7, 2017 · TerminatedExactech Optetrak Logic CC 15mm Tibial Augment Screws
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page