Truliant 02-022-35-5011Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Truliant (414)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
19
Deaths reported
0
Injuries reported
19
Recalls naming it
1

Summary

The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 10885862304636), received between Jul 1, 2020 and Aug 10, 2026 — classified as 19 injuries.

The most frequently reported specific device problem is “Naturally Worn” (15 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 5 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2021Class IIOpen, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0023-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn15
  • Insufficient Device Problem Information7
  • Adverse Event Without Identified Device or Use Problem2
  • Unstable2
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Mechanical Problem1

Patient problems

  • Insufficient Information10
  • Pain7
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3
  • Failure of Implant3
  • Joint Laxity3
  • Swelling/ Edema3
  • Synovitis3
  • Discomfort2

Recent adverse event reports

The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 10, 2026Unstable; Naturally Worn

    Patient problems: Ambulation or Postural Difficulties; Discomfort; Joint Laxity; Pain; Scar Tissue; Swelling/ Edema; Synovitis

    MAUDE report 50189478 ↗
  • InjuryOct 30, 2024Naturally Worn; Naturally Worn

    LEGAL CASE ¿ USA CASE-(B)(4) IS ASSOCIATED WITH THIS CASE. IT WAS REPORTED THAT APPROXIMATELY 88 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT HAS EXPERIENCED PROSTHESIS WEAR. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE SERIAL NUMBER (B)(6) IS CONFIRMED TO BE A PART OF RECALL NUMBER: Z-0023-2022.

    Patient problems: Insufficient Information

    MAUDE report 20569522 ↗
  • InjuryAug 9, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT A MALE PATIENT, INITIAL RIGHT KNEE IMPLANTED IN (B)(6) 2018, UNDERWENT A REVISION PROCEDURE IN (B)(6) 2024, APPROXIMATELY 5 YEARS 11 MONTHS POST THE INITIAL PROCEDURE. THE PATIENT PRESENTED WITH COMPLAINTS OF PAIN. THEY WERE REVISED TO A COMPETITOR¿S DEVICES DUE TO THE RECALL. RECALLED POLY INDICATED. THERE WERE NO DEVICE BREAKAGES OR SURGICAL DELAYS DURING THE PROCEDURE. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO X-RAYS WERE ABLE TO BE OBTAINED. THE EXPLANTED DEVICES ARE NOT AVAILABLE DUE THE RECALL, THE HOSPITAL WILL NOT RELEASE THEM. AN IMAGE OF THE EXPLANTED DEVICES WAS PROVIDED. NO FURTHER INFORMATION.

    Patient problems: Pain

    MAUDE report 19956187 ↗
  • InjuryJul 24, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 75 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT HAS EXPERIENCED PROSTHESIS WEARDEFENDANT'S DEFECTIVE IMPLANT NECCESSITATED OR WILL "NECESSITATE" SURGICAL REVISION, "PERMANENT" IMPAIRMENT AND HEALTH CARE EXPENSES DUE TO THE REVISION. "PLAINTIFF" HAD AND WILL CONTINUE TO HAVE PAIN AND SUFFERING, LOST ENJOYMENT OF LIFE, DISFIGUREMENT FROM SCARS AND LOST EARNING CAPACITY.. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Pain; Scar Tissue

    MAUDE report 19818536 ↗
  • InjuryJul 16, 2024Naturally Worn; Naturally Worn

    AS REPORTED, THE PATIENT HAD AN INITIAL RIGHT TKA ON (B)(6) 2017. THE PATIENT COMPLAINED OF PAIN AND WAS REVISED ON (B)(6) 2024 DUE TO A RECALLED POLY AND LOOSE FEMUR. THERE WAS NO REPORTED BREAKAGE OF A DEVICE OR SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO OTHER PATIENT INFORMATION/MEDICAL HISTORY REPORTED.

    Patient problems: Joint effusion; Osteolysis; Pain

    MAUDE report 19750485 ↗
  • InjuryJul 11, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 20 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT HAS EXPERIENCED PROSTHESIS WEAR. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information; Medical device site infection

    MAUDE report 19719964 ↗
  • InjuryJul 11, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 54 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT HAS EXPERIENCED PROSTHESIS WEAR. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 19719383 ↗
  • InjuryJul 10, 2024Mechanical Problem; Naturally Worn; Mechanical Problem; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 15 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, JOINT PAIN, JOINT SWELLING AND STIFFNESS, LIMITED RANGE OF MOTION, LOSS OF MOBILITY, TISSUE DAMAGE, REVISION SURGERY, POLYETHYLENE WEAR, AND SYNOVITIS. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Implant Pain; Loss of Range of Motion; Swelling/ Edema; Synovitis

    MAUDE report 19716717 ↗
  • InjuryApr 10, 2024Naturally Worn; Naturally Worn

    AS REPORTED, THE PATIENT HAD AN INITIAL RIGHT TKA ON (B)(6) 2018. THE PATIENT WAS REVISED ON (B)(6) 2024 DUE TO POLY WEAR OVER TIME. THERE WAS NO REPORTED BREAKAGE OF A DEVICE OR SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.

    Patient problems: Failure of Implant

    MAUDE report 19079519 ↗
  • InjuryMar 19, 2024Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A LEFT KNEE REPLACEMENT ON (B)(5) 2017. APPROXIMATELY 5 YEARS AFTER THE INITIAL PROCEDURE THE PATIENT HAD A LEFT KNEE REVISION ON (B)(6) 2022. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE PATIENT HAS FILED A SHORT-FORM COMPLAINT IN A MULTIDISTRICT LITIGATION TITLED IN RE: EXACTECH POLYETHYLENE ORTHOPEDIC PRODUCTS LIABILITY LITIGATION, MDL NO. 3044 (NGG) (MMH), AND PENDING IN THE EASTERN DISTRICT OF NEW YORK. PER THE TRANSFER ORDER CREATING THIS MULTIDISTRICT LITIGATION, ¿PLAINTIFFS¿ALLEGE THAT THEIR KNEE OR HIP REPLACEMENT DEVICES¿FAILED PREMATURELY BECAUSE OF DEGRADATION OF THE DEVICE¿S POLYETHYLENE COMPONENT, WHICH RESULTED IN THE PREMATURE REMOVAL (OR PLANNED REMOVAL) OF THE PROSTHESIS AT ISSUE.¿ BECAUSE THE PATIENT HAS FILED A SUIT IN THIS MULTIDISTRICT LITIGATION, THE PATIENT APPEARS TO ALLEGE THAT THE PATIENT WAS INJURED AS A RESULT OF PREMATURE WEAR OF AN EXACTECH POLYETHYLENE DEVICE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Discomfort; Insufficient Information; Pain

    MAUDE report 18928728 ↗
  • InjurySep 21, 2023Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A RIGHT KNEE REPLACEMENT ON (B)(6) 2019. THE DEVICE HAS NOT BEEN EXPLANTED. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE PATIENT HAS FILED A SHORT-FORM COMPLAINT IN A MULTIDISTRICT LITIGATION TITLED IN RE: EXACTECH POLYETHYLENE ORTHOPEDIC PRODUCTS LIABILITY LITIGATION, MDL NO. 3044, AND PENDING IN THE EASTERN DISTRICT OF NEW YORK. PER THE TRANSFER ORDER CREATING THIS MULTIDISTRICT LITIGATION, ¿PLAINTIFFS¿ALLEGE THAT THEIR KNEE OR HIP REPLACEMENT DEVICES¿FAILED PREMATURELY BECAUSE OF DEGRADATION OF THE DEVICE¿S POLYETHYLENE COMPONENT, WHICH RESULTED IN THE PREMATURE REMOVAL (OR PLANNED REMOVAL) OF THE PROSTHESIS AT ISSUE.¿ BECAUSE THE PATIENT HAS FILED A SUIT IN THIS MULTIDISTRICT LITIGATION, THE PATIENT APPEARS TO ALLEGE THAT THE PATIENT WAS INJURED AS A RESULT OF PREMATURE WEAR OF AN EXACTECH POLYETHYLENE DEVICE.

    Patient problems: Insufficient Information

    MAUDE report 17789543 ↗
  • InjurySep 14, 2023Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A LEFT KNEE REVISION ON (B)(6) 2020. APPROXIMATELY 2 YEARS AND 11 MONTHS AFTER THE FIRST REVISION THE PATIENT HAD A SECOND LEFT KNEE REVISION ON (B)(6) 2023. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE PATIENT HAS FILED A SHORT-FORM COMPLAINT IN A MULTIDISTRICT LITIGATION TITLED IN RE: EXACTECH POLYETHYLENE ORTHOPEDIC PRODUCTS LIABILITY LITIGATION, MDL NO. 3044, AND PENDING IN THE EASTERN DISTRICT OF NEW YORK. PER THE TRANSFER ORDER CREATING THIS MULTIDISTRICT LITIGATION, ¿PLAINTIFFS¿ALLEGE THAT THEIR KNEE OR HIP REPLACEMENT DEVICES¿FAILED PREMATURELY BECAUSE OF DEGRADATION OF THE DEVICE¿S POLYETHYLENE COMPONENT, WHICH RESULTED IN THE PREMATURE REMOVAL (OR PLANNED REMOVAL) OF THE PROSTHESIS AT ISSUE.¿ BECAUSE THE PATIENT HAS FILED A SUIT IN THIS MULTIDISTRICT LITIGATION, THE PATIENT APPEARS TO ALLEGE THAT THE PATIENT WAS INJURED AS A RESULT OF PREMATURE WEAR OF AN EXACTECH POLYETHYLENE DEVICE.

    Patient problems: Insufficient Information; Joint Laxity; Osteolysis; Synovitis

    MAUDE report 17746431 ↗
  • InjurySep 7, 2023Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information

    AS REPORTED, THE PATIENT HAD AN INITIAL RIGHT TKA ON (B)(6) 2018. THE PATIENT WAS REVISED AND THE POLY EXCHANGED ON (B)(6) 2023 DUE TO THE RECALLED POLY. THERE WAS NO REPORTED BREAKAGE OF A DEVICE OR SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO OTHER PATIENT INFORMATION/MEDICAL HISTORY REPORTED.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 17706744 ↗
  • InjuryJun 2, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    AS REPORTED VIA LEGAL DOCUMENTATION, A PATIENT UNDERWENT LEFT KNEE REPLACEMENT ON (B)(6) 2017. THEY THEN HAD LEFT KNEE REVISION ON (B)(6) 2023, APPROXIMATELY 5 YEARS 5 MONTHS AFTER THEIR INITIAL PROCEDURE. THE REASON FOR REVISION HAS NOT BEEN PROVIDED. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 17054315 ↗
  • InjuryMay 19, 2023Unstable; Naturally Worn; Insufficient Device Problem Information; Unstable; Naturally Worn; Insufficient Device Problem Information

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A RIGHT KNEE REPLACEMENT ON 1 MAR 2021. THE PATIENT HAS SUFFERED THE FOLLOWING INJURIES AND COMPLICATIONS AS A RESULT OF THIS EXACTECH DEVICE: SWELLING, LOOSENESS, AND KNOCKING. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. CONCOMITANTS: 6693719, 02-012-60-1425 - LOGIC STEM EXT 14MM X 25MM. 6608185, 02-020-11-0350 - RULIANT PS CEM FEM PS CEM RIGHT SZ 5. 6666469, 02-022-45-5050 - TRULIANT TIB FIT TRAY CEM SZ 5F / 5T. 6606281, 200-07-35 - ADVANCED PATELLA 35MM 3 PEG IMPLANT. 6619824, 204-70-00 - TIBIAL STEM EXT. SCREW.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Insufficient Information; Joint Laxity; Swelling/ Edema

    MAUDE report 16964182 ↗
  • InjuryMay 3, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    AS REPORTED VIA LEGAL DOCUMENTATION, THIS PATIENT HAD KNEE IMPLANTATION DATE OF (B)(6) 2019. THEY HAD REVISION SURGERY (B)(6) 2023, APPROXIMATELY 3 YEARS 3 MONTHS AFTER THEIR INITIAL PROCEDURE. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 16860109 ↗
  • InjuryMar 21, 2023Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED, APPROXIMATELY 3.5 YEARS POST OP THE INITIAL LEFT TKA, THIS 60 Y/O MALE PATIENT WAS REVISED DUE TO PAIN. PATIENT HAD A LOOSE FEMUR AND RECALLED POLY. NO BREAKAGE OF DEVICE OR SURGICAL DELAY. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO OTHER PATIENT INFORMATION/MEDICAL HISTORY REPORTED. X-RAY & PHOTO ARE ATTACHED. PRODUCT IS NOT RETURNING, DUE TO RECALL HOSPITAL WILL NOT RELEASE IMPLANTS.

    Patient problems: Failure of Implant; Joint effusion; Pain

    MAUDE report 16584618 ↗
  • InjuryFeb 1, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED, APPROXIMATELY EIGHT MONTHS POST INITIAL RIGHT SIDE TKA, THIS MALE PATIENT UNDERWENT REVISION DUE TO INFECTION. NO ADDITIONAL INFORMATION IS AVAILABLE OR FORTHCOMING. THE DEVICE IS NOT AVAILABLE FOR EVALUATION.

    Patient problems: Bacterial Infection

    MAUDE report 13419087 ↗
  • InjuryJul 1, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    APPROXIMATELY 10 YEARS POST OP THE INITIAL IMPLANT, THIS (B)(6) Y/O MALE PATIENT PRESENTED COMPLAINING OF RIGHT KNEE PAIN. THE PATIENT WAS SCHEDULED FOR I & D AND POLY SWAP PROCEDURE. THIS PROCEDURE WAS A MAJOR DEBRIDEMENT OF KNEE WITH RESULTED IN THE LOOSENING OF THE EXTENSION GAP (PRE OP FLEXION CONTRACTURE) USING THE SAME SIZE AND CONSTRAINT INSERT WE WERE ABLE TO PROPERLY BALANCE THE JOINT. THE SURGEON REPLACED THE SAME SIZE LINER (SIZE 5 11MM PS INSERT). THE REMOVED INSERT WAS IN GOOD SHAPE AND NO ISSUES REPLACING THE INSERT WITH THE TRULIANT. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. DEVICE NOT RETURNING: ALL HSS EXPLANTS SENT TO BIOMEDICAL.

    Patient problems: Pain

    MAUDE report 10219721 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →