1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10885862305138), received between Jan 5, 2024 and Jul 26, 2024 — classified as 6 injuries.

The most frequently reported specific device problem is “Naturally Worn” (6 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2021Class IIOpen, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0023-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn6
  • Insufficient Device Problem Information2
  • Unstable1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4
  • Discomfort4
  • Insufficient Information4
  • Pain4
  • Ambulation or Postural Difficulties1
  • Failure of Implant1
  • Joint Laxity1
  • Swelling/ Edema1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 26, 2024Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information

    LEGAL CASE ¿ USA BULK LEGAL EMAIL (440). CASE: (B)(4); LK REV # 1 CASE: (B)(4); LK REV #2 IS ASSOCIATED WITH THIS CASE. IT WAS REPORTED THAT APPROXIMATELY 29 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE FOR UNKNOWN ISSUES. THE DEVICE WAS NOT RETURNED FOR EVALUATION. NO PHOTOGRAPHS OR X-RAYS WERE PROVIDED FOR REVIEW; THEREFORE THE REPORTED EVENT CANNOT BE CONFIRMED THROUGH ANALYSIS. NO OPERATIVE NOTES WERE PROVIDED; THEREFORE, A REVIEW OF DEVICE USAGE AND TECHNIQUE COULD NOT BE PERFORMED. NO INFORMATION CONCERNING PATIENT CONDITIONS, CO-MORBIDITIES, OR OTHER HEALTH OR ANATOMICAL CONCERNS WAS PROVIDED AND IT IS THEREFORE UNKNOWN IF PATIENT-RELATED FACTORS MAY HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE CAUSE OF THE REPORTED EVENT CANNOT BE DETERMINED BASED ON THE INFORMATION MADE AVAILABLE. SHOULD ADDITIONAL, RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

    Patient problems: Insufficient Information

    MAUDE report 19844500 ↗
  • InjuryJun 4, 2024Unstable; Naturally Worn; Unstable; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 81 MONTHS AFTER A RIGHT KNEE REPLACEMENT PROCEDURE, THE PATIENT MAY REQUIRE A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, POPPING, CRACKING, OCCASIONAL WEAKNESS, AND DISCOMFORT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Ambulation or Postural Difficulties; Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Discomfort; Insufficient Information; Joint Laxity; Pain; Swelling/ Edema

    MAUDE report 19463752 ↗
  • InjuryApr 17, 2024Naturally Worn; Naturally Worn

    IT IS REPORTED VIA LEGAL DOCUMENTATION THAT A PATIENT HAD A RIGHT TOTAL KNEE ARTHROPLASTY SECOND REVISION ON (B)(6) 2022, WITH NO REVISION SURGERY REPORTED. NO RADIOGRAPHIC IMAGES OF THE DEVICE ARE PROVIDED.THERE IS NO OTHER INFORMATION AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Discomfort; Pain

    MAUDE report 19124886 ↗
  • InjuryFeb 28, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT A PATIENT HAD AN INITIAL LEFT TOTAL KNEE ARTHROPLASTY ON (B)(6) 2022, WITH NO REVISION SURGERY REPORTED. THERE WAS NO OTHER PATIENT/MEDICAL INFORMATION PROVIDED. NO X-RAYS OR IMAGES WERE PROVIDED. THERE IS NO OTHER INFORMATION AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Discomfort; Pain

    MAUDE report 18798313 ↗
  • InjuryJan 11, 2024Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A LEFT KNEE REVISION ON (B)(6) 2022. THIS DEVICE HAS NOT BEEN EXPLANTED. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE PATIENT HAS FILED A SHORT-FORM COMPLAINT IN A MULTIDISTRICT LITIGATION TITLED IN RE: EXACTECH POLYETHYLENE ORTHOPEDIC PRODUCTS LIABILITY LITIGATION, MDL NO. 3044 (NGG) (MMH), AND PENDING IN THE EASTERN DISTRICT OF NEW YORK. PER THE TRANSFER ORDER CREATING THIS MULTIDISTRICT LITIGATION, ¿PLAINTIFFS¿ALLEGE THAT THEIR KNEE OR HIP REPLACEMENT DEVICES¿FAILED PREMATURELY BECAUSE OF DEGRADATION OF THE DEVICE¿S POLYETHYLENE COMPONENT, WHICH RESULTED IN THE PREMATURE REMOVAL (OR PLANNED REMOVAL) OF THE PROSTHESIS AT ISSUE.¿ BECAUSE THE PATIENT HAS FILED A SUIT IN THIS MULTIDISTRICT LITIGATION, THE PATIENT APPEARS TO ALLEGE THAT THE PATIENT WAS INJURED AS A RESULT OF PREMATURE WEAR OF AN EXACTECH POLYETHYLENE DEVICE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Discomfort; Insufficient Information; Pain

    MAUDE report 18494158 ↗
  • InjuryJan 5, 2024Naturally Worn; Naturally Worn

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A LEFT KNEE REPLACEMENT ON (B)(6)2020. APPROXIMATELY 2 YEARS AND 11 MONTHS AFTER THE INITIAL PROCEDURE THE PATIENT HAD A LEFT KNEE REVISION ON (B)(6) 2023. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. REVISION OP REPORT ON (B)(6) 2023 DIAGNOSIS: FAILURE OF TOTAL KNEE REPLACEMENT THERE WAS SIGNIFICANT EFFUSION AND SYNOVITIS. NO PURULENCE. THERE WAS DELAMINATION OF THE POLYETHYLENE BUT EVIDENCE OF PROTHESIS LOOSENING. THE POLYETHYLENE WAS REMOVED. THERE WAS OSTEOLYSIS SURROUNDING THE FEMORAL COMPONENT ANTERIOR AND POSTERIOR AS WELL AS THE MEDIAL TIBIAL PLATEAU. PATIENT TOLERATED THE PROCEDURE WELL AND WAS TRANSFERRED TO THE RECOVERY ROOM IN STABLE CONDITION.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 18452257 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →