Skip to content

Truliant 02-022-44-3511Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Truliant (414)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
7
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10885862305190), received between Apr 27, 2023 and Oct 16, 2024 — classified as 7 injuries.

The most frequently reported specific device problem is “Naturally Worn” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2021Class IIOpen, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0023-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn4
  • Adverse Event Without Identified Device or Use Problem3
  • Insufficient Device Problem Information3
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Unstable1

Patient problems

  • Failure of Implant5
  • Insufficient Information4
  • Implant Pain2
  • Scar Tissue2
  • Joint Laxity1
  • Pain1
  • Rupture1
  • Swelling/ Edema1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 16, 2024Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    DURING REVIEW OF THE PATIENT'S PREVIOUS REVISION (REFER TO REPORT NUMBER: 1038671-2024-03993), IT WAS DISCOVERED THE PATIENT UNDERWENT A SECOND KNEE REVISION SURGERY; REASON UNKNOWN. NO FURTHER INFORMATION AT THIS TIME.

    Patient problems: Insufficient Information

    MAUDE report 20464975 ↗
  • InjuryJul 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT APPROXIMATELY 2 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS A QUAD RUPTURE AND ALSO AN INCIDENTAL POLY EXCHANGE. THERE WAS NO REPORTED BREAKAGE OF A DEVICE OR SURGICAL DELAY/PROLONGATION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Rupture

    MAUDE report 19727197 ↗
  • InjuryApr 15, 2024Naturally Worn; Adverse Event Without Identified Device or Use Problem; Naturally Worn; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT A PATIENT HAD A KNEE ARTHROPLASTY ON AN UNKNOWN DATE IN (B)(6) 2020, THEN APPROXIMATELY 2 YEARS LATER HAD A KNEE REVISION SURGERY ON (B)(6) 2022. THE DATE OF IMPLANT IS NOT REPORTED/ THE DEVICES ARE NOT REPORTED/ THERE IS NO INFORMATION ON THE WHICH KNEE WAS REPLACED OR REVISED. THERE IS NO OTHER INFORMATION PROVIDED.

    Patient problems: Failure of Implant; Implant Pain; Insufficient Information; Scar Tissue

    MAUDE report 19104755 ↗
  • InjuryFeb 22, 2024Unstable; Naturally Worn; Unstable; Naturally Worn

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A LEFT KNEE REPLACEMENT ON (B)(6) 2020. APPROXIMATELY 2 YEARS AND 9 MONTH AFTER THE INITIAL PROCEDURE THE PATIENT HAD A LEFT KNEE REVISION ON (B)(6) 2023. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. REVISION OP REPORT ON (B)(6) 2023. DIAGNOSIS: FAILED LEFT KNEE SECONDARY TO PREMATURE POLYETHYLENE WEAR AS WELL AS ASEPTIC LOOSENING OF THE FEMORAL COMPONENT. THERE WAS EARLY POLYETHYLENE SYNOVITIS WITH PARTICULATE DEBRIS IN THE GUTTERS. THERE WAS FLATTENING AND EARLY WEAR ON THE LATERAL FACET. THE PATIENT WAS TRANSFERRED TO THE RECOVERY ROOM IN STABLE CONDITION.

    Patient problems: Failure of Implant; Joint Laxity; Pain; Swelling/ Edema; Synovitis

    MAUDE report 18758135 ↗
  • InjuryJan 11, 2024Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information

    IT WAS REPORTED THAT 1.5 YEARS POST OP INITIAL RIGHT TKA, THIS PATIENT HAD A REVISION. THE 11MM PSC POLY WAS EXCHANGED FOR A 17MM PSC POLY DURING A POLY SWAP AND QUAD TENDON REPAIR PROCEDURE. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO X-RAYS OR PHOTOS PROVIDED, UNABLE TO OBTAIN. NO PRODUCT RETURN; AGAINST POLICY.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 18498250 ↗
  • InjuryJun 20, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THIS PATIENT'S LEFT KNEE WAS REVISED. SURGEON PREFORMED A POLY SWAP, POLY WAS SWAPPED TO TRULIANT PSC TIBIAL INSERT SZ 3.5 X 15MM. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO X-RAYS OR PHOTOS PROVIDED, UNABLE TO OBTAIN. NO PRODUCT RETURN. NO REASON REPORTED FOR PRODUCT NOT BEING RETURNED.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 17165327 ↗
  • InjuryApr 27, 2023Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED VIA LEGAL NOTIFICATION: A MALE PATIENT HAD AN INITIAL RIGHT KNEE ARTHROPLASTY ON (B)(6) 2020, AND THEN UNDERWENT REVISION SURGERY ON (B)(6) 2022, APPROXIMATELY 2 YEARS AFTER IMPLANT PROCEDURE. REVISION SURGERY OPERATIVE REPORT OF (B)(6) 2022-FAILED RIGHT KNEE ARTHROPLASTY SECONDARY TO ASEPTIC LOOSENING AND POLY WEAR. OBESE BMI 37-THE DURATION OF THE PROCEDURE WAS INCREASED BY APPROXIMATELY 40%, DUE TO PATIENT SIZE, THIS INCREASED COMPLEXITY AND DIFFICULTY OF SURGICAL PROCEDURE. FINDINGS: INTRAOPERATIVE FINDINGS CONFIRMED THE RADIOGRAPHIC AND CLINICAL FINDINGS OF FAILURE OF THE KNEE ARTHROPLASTY SECONDARY TO ASEPTIC LOOSENING (FEMORAL COMPONENT, TIBIAL COMPONENT WELL FIXED) CHARACTERIZED BY NO BONY INGROWTH ON THE BACKSIDE OF THE FEMORAL COMPONENT WITH EASY COMPONENT REMOVAL. ABUNDANT POLYETHYLENE PARTICLES THROUGHOUT THE KNEE. POLY WAS REMOVED AND INSPECTED, THERE WAS NOTABLE WEAR ON THE TIBIAL POST. THE PATELLA WAS FOUND TO BE WELL-FIXED AND WITHOUT SIGNIFICANT SURFACE DAMAGE, DELAMINATION OR PITTING; IT WAS DECIDED TO PLACE A NEW BUTTON. A DRESSING AND WRAP WERE APPLIED, THE PATIENT WAS TRANSFERRED TO THE RECOVERY ROOM. THE PATIENT TOLERATED THE PROCEDURE WELL WITHOUT

    Patient problems: Failure of Implant; Implant Pain; Scar Tissue

    MAUDE report 16824514 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page