Skip to content

Truliant 02-022-45-4040Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Truliant (414)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10885862305886), received between Oct 18, 2024 and Jan 23, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth2
  • Unstable2

Patient problems

  • Discomfort2
  • Joint Laxity2
  • Swelling/ Edema2
  • Ambulation or Postural Difficulties1
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Balance Problems1
  • Implant Pain1
  • Joint effusion1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 23, 2025Unstable; Loosening of Implant Not Related to Bone-Ingrowth; Unstable; Loosening of Implant Not Related to Bone-Ingrowth

    ADDITIONAL INFORMATION RECEIVED FROM LEGAL ON 07AUG2023. AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A RIGHT KNEE REPLACEMENT ON (B)(6)2020. APPROXIMATELY 2 YEARS AND 2 MONTHS AFTER THE INITIAL PROCEDURE THE PATIENT HAD A RIGHT KNEE REVISION ON (B)(6) 2022. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. REVISION OP REPORT ON 8 NOV 2022. DIAGNOSIS: FAILED RIGHT TKA; RIGHT KNEE INSTABILITY WITH VARUS STRESS FINDINGS: THE QUAD WAS ADHERENT ANTERIORLY AND A QUADRICEPSPLASTY WAS REQUIRED TO FREE THE EXTENSOR MECHANISM. THE FEMORAL COMPONENT WAS WELL FIXED BUT INTERNALLY ROTATED. THE TIBIAL TRAY WAS IN SLIGHT VARUS, BUT THERE WAS A NOTICEABLE GAP AT THE CEMENT BONE INTERFACE CONCERNING FOR LOOSENING. ORIGINAL DESCRIPTION SUMMARY: LEGAL CASE ¿ (B)(6). IMPLANT DATE: (B)(6) 2020. EXPLANT DATE: (B)(6) 2022. NO DEVICE RETURN ANTICIPATED DUE TO THIS BEING A LEGAL CASE. UNABLE TO LOCATE SURGEON/PATIENT/SERIAL NUMBER IN EBI. NO SERIAL NUMBERS AVAILABLE. HISTORICAL RECORD AND SMARTSOLVE SEARCHED FOR SN/PATIENT NAME WITH NO RESULTS. NO FURTHER

    Patient problems: Discomfort; Implant Pain; Joint Laxity; Joint effusion; Scar Tissue; Swelling/ Edema

    MAUDE report 21218077 ↗
  • InjuryOct 18, 2024Unstable; Loosening of Implant Not Related to Bone-Ingrowth; Unstable; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A LEFT KNEE REVISION ON (B)(6) 2020. APPROXIMATELY 2 YEARS AFTER THE FIRST REVISION THE PATIENT HAD A SECOND LEFT KNEE REVISION ON (B)(6) 2022 DUE TO ACCELERATED AND PREVENTABLE WEAR. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Ambulation or Postural Difficulties; Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Balance Problems; Discomfort; Joint Laxity; Pain; Swelling/ Edema

    MAUDE report 20481895 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page