Skip to content

Truliant 02-022-47-4009Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Truliant (414)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10885862306470), received between Sep 13, 2022 and Nov 4, 2024 — classified as 4 injuries.

The most frequently reported specific device problem is “Naturally Worn” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2021Class IIOpen, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0023-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn3
  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Device Problem Information1

Patient problems

  • Pain2
  • Bacterial Infection1
  • Failure of Implant1
  • Insufficient Information1
  • Joint Laxity1
  • No Clinical Signs, Symptoms or Conditions1
  • Osteolysis1
  • Synovitis1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 4, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THIS PATIENT'S KNEE WAS REVISED APPROXIMATELY 3 YEARS 8 MONTHS POST OP. PATIENT HAS HAD CONSISTENT PAIN AND EXPERIENCED KNEE BUCKLING. FOLLOWING THE SURGERY, SHE HAD TO COMPLETED PHYSICAL THERAPY, WATER THERAPY, AND MUSCLE STIM THERAPY BEFORE GETTING THE IMPLANTS REMOVED. SURGEON MENTIONED BONE LOSS AND OVERLAPPING DEVICE TO BONE. PATIENT WAS REVISED TO COMPETITORS DEVICE.

    Patient problems: Joint Laxity; Osteolysis; Pain

    MAUDE report 20605835 ↗
  • InjurySep 22, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED, THE PATIENT HAD AN INITIAL TKA ON (B)(6) 2019. THE PATIENT PRESENTED ON (B)(6) 2019 AND PRESENTED TO THE ER WITH SEVERE INFECTION, THE LEG WITH THE HARDWARE WAS THEN AMPUTATED. OUTCOME: RESOLVED ON (B)(6) 2019.

    Patient problems: Bacterial Infection; Unspecified Infection

    MAUDE report 17800454 ↗
  • InjuryApr 13, 2023Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information

    REVISION OPERATIVE REPORT OF (B)(6) 2022. INDICATION FOR SURGERY, THIS 61 YO FEMALE HAS A RECALLED POLYETHYLENE LINER. THE PATIENT WAS HAVING POSTOPERATIVE PAIN AND HAD A NEGATIVE INFECTION WORK-UP. POLY EXCHANGE WAS DONE REPLACING THE 9MM TO A 10MM POLY. AT THE TIME OF SURGERY THERE WAS NOTED TO BE SOME OVERGROWTH OF BONE ALONG THE MEDIAL TIBIAL TRAY AS WELL AS AN OSSIFIED FRAGMENT IN THE POSTERIOR MEDIAL JOINT LINE THAT WAS REMOVED. THERE IS SOME MILD WEAR ON THE POSTERIOR MEDIAL CORNER OF THE POLYETHYLENE AND SOME MILD WEAR ALONG THE SURFACE OF THE POLYS THERE WAS NO SIGNIFICANT OSTEOLYSIS SEEN AND MINIMAL SYNOVITIS. STERILE DRESSINGS WERE APPLIED, AND THE PATIENT WAS TAKEN TO RECOVERY ROOM IN STABLE CONDITION, THERE WERE NO SURGICAL COMPLICATION. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Pain; Synovitis

    MAUDE report 16731135 ↗
  • InjurySep 13, 2022Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT A MALE PATIENT, WHO HAD INITIAL RIGHT KNEE IMPLANTED WITH RECALLED POLY ON (B)(6) 2019, WAS REVISED TO A LOGIC CC SYSTEM ON (B)(6) 2022, APPROXIMATELY 3 YEARS 2 MONTHS POST THE INITIAL IMPLANT PROCEDURE DUE TO WEAR. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. THE DEVICE IS AVAILABLE FOR RETURN. NO OTHER PATIENT INFORMATION/MEDICAL HISTORY PROVIDED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15413610 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page