Skip to content

Truliant 02-022-47-4011Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Truliant (414)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10885862306487), received between Aug 8, 2023 and Apr 23, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Fracture” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2021Class IIOpen, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0023-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Fracture1
  • Insufficient Device Problem Information1
  • Mechanical Problem1
  • Naturally Worn1
  • Unstable1

Patient problems

  • Failure of Implant2
  • Anxiety1
  • Bone Fracture(s)1
  • Implant Pain1
  • Insufficient Information1
  • Joint Laxity1
  • Loss of Range of Motion1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 23, 2024Mechanical Problem; Unstable; Naturally Worn; Adverse Event Without Identified Device or Use Problem; Mechanical Problem; Unstable; Naturally Worn; Adverse Event Without Identified Device or Use Problem

    AS REPORTED, THE SURGEON HAD A FEMALE PATIENT THAT CAME IN WITH A SMALL NON DISPLACED DISTAL FEMUR FRACTURE ON THE MEDIAL SIDE OF THE FEMUR THAT HAD A TRULIANT TOTAL RIGHT KNEE IMPLANTED ON (B)(6) 2021. THE SURGEON SAID THE PATIENT KNEW SHE WAS PART OF THE POLY RECALL AND STATED SHE FELL BECAUSE HER KNEE WAS UNSTABLE "BECAUSE OF THE RECALL". DURING SURGERY A COMPETITOR'S RETROGRADE NAIL WAS USED TO REPAIR THE FRACTURE. THE TOTAL KNEE COMPONENTS WERE WILL FIXED AND THE POLY INSERT WAS REMOVED. THERE WERE NO SIGNS OF WEAR ON THE POLY AND IT HAD NO NOTICEABLE OXIDATION OR DELAMINATION. A NEW ACTIVIT-E SZ 4, 12MM POLY WAS INSERTED AND IMPLANTED. THERE WAS NO BREAKAGE OF DEVICE OR SURGICAL DELAY/PROLONGATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. IMAGES OF THE LINER WERE TAKEN. PHOTOS/X-RAYS RECEIVED. THE DEVICE IS NOT AVAILABLE FOR EVALUATION DUE TO POLY WAS KEPT AFTER CASE CONCLUDED.

    Patient problems: Anxiety; Bone Fracture(s); Failure of Implant; Implant Pain; Joint Laxity; Loss of Range of Motion

    MAUDE report 19169045 ↗
  • InjuryAug 8, 2023Fracture; Insufficient Device Problem Information; Fracture; Insufficient Device Problem Information

    IT WAS REPORTED THAT THIS MALE PATIENT'S KNEE WAS REVISED DUE TO THE RECALL. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO FURTHER INFORMATION WAS PROVIDED. NO X-RAYS OR PHOTOS AVAILABLE. DEVICE IS NOT RETURNING - HOSPITAL KEPT.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 17478846 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page