Equinoxe 300-30-09Recalls & reported problems
Exactech, Inc. · Class 2 · Implant
All models of Equinoxe (305)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10885862515773), received between Oct 28, 2024 and Jan 6, 2026 — classified as 3 injuries.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
1 other recall cover devices cleared under the same FDA submission (K162726); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Detachment of Device or Device Component1
- Lack of Effect1
- Loosening of Implant Not Related to Bone-Ingrowth1
Patient problems
- Insufficient Information2
- Bone Fracture(s)1
- Limb Fracture1
- Loss of Range of Motion1
- No Clinical Signs, Symptoms or Conditions1
- Pain1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJan 6, 2026Detachment of Device or Device Component; Lack of Effect
IT WAS REPORTED THAT THE PATIENT UNDERWENT A RIGHT REVERSE TOTAL SHOULDER REVISION. SUBSEQUENTLY, THE PATIENT'S TORQUE DEFINING SCREW SEPARATED FROM THE HUMERAL STEM RESULTING IN DISASSOCIATION OF THE HUMERAL TRAY. AS A RESULT, THE PATIENT UNDERWENT AN ADDITIONAL SHOULDER REVISION SURGERY APPROXIMATELY 6 MONTHS AFTER THEIR FIRST REVISION SURGERY AND 6 YEARS AFTER THEIR INITIAL IMPLANTATION. NO FURTHER PATIENT IMPACT REPORTED. NO FURTHER INFORMATION AVAILABLE. THIS IS REPORT 2 OF 2 FOR THIS EVENT.
Patient problems: Insufficient Information; Loss of Range of Motion; Pain
MAUDE report 23981470 ↗InjuryDec 8, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL TSA (TOTAL SHOULDER ARTHROPLASTY) ON THE RIGHT SIDE. SUBSEQUENTLY, THE PATIENT WAS REVISED DUE TO THE INITIAL GLENOSPHERE BEING TOO BIG. THE SURGEON INITIALLY THOUGHT THE STEM WAS LOOSE. AS A RESULT, THE SURGEON REMOVED THE 42MM GLENOSPHERE AND REPLACED WITH 38MM GLENOSPHERE. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23746255 ↗InjuryOct 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INTRA-OPERATIVELY OF A RIGHT RTSA, IT WAS REPORTED VIA CLINICAL STUDY THAT THE PATIENT EXPERIENCED A HUMERAL FRACTURE. IT WAS FURTHER REPORTED THAT A SMALL MEDIAL CALCAR CRACK OCCURRED DURING TRIALING. THE ACTION TAKEN DURING THE PROCEDURE WAS CEMENTED STEM AND THE OUTCOME OF THIS EVENT REMAINS CONTINUING. THE CLINICAL REPORT INDICATES THAT THIS EVENT IS UNLIKELY RELATED TO THE DEVICE(S) AND POSSIBLY RELATED TO THE PROCEDURE.
Patient problems: Bone Fracture(s); Limb Fracture
MAUDE report 20550404 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K162726. One submission often covers many models, so check the product and lot details in each record.
- Z-0958-2019 · Dec 5, 2018 · TerminatedEquinoxe Preserve Humeral Stem, 8mm Product Usage: The Equinoxe Preserve Stems are indicated for use in skeletally mature individuals with degenerative diseases of the glenohum
Other Shoulder Prosthesis, Reverse Configuration devices
- DJO SURGICAL 508-36-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration508 reports
- DJO SURGICAL 508-32-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration433 reports
- DJO SURGICAL 508-32-103Encore Medical, L.P. · shoulder prosthesis, reverse configuration421 reports
- DJO SURGICAL 509-00-036Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall259 reports
- DJO SURGICAL 508-32-104Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall203 reports
- DJO SURGICAL 509-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration233 reports
- DELTA XTEND 130760000Depuy (Ireland) · shoulder prosthesis, reverse configuration223 reports
- DJO SURGICAL 508-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration122 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page