Skip to content

Equinoxe 300-30-15Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Equinoxe (305)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885862515834), received between Oct 2, 2024 and Oct 2, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Naturally Worn” (1 report).

1 other recall cover devices cleared under the same FDA submission (K162726); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Naturally Worn1

Patient problems

  • Implant Pain1
  • Loss of Range of Motion1
  • Metal Related Pathology1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 2, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THIS PATIENT HAD A RIGHT REVERSE SHOULDER PUT IN 2020. REVISED AGAIN IN 2021 BY A LOCAL ORTHOPEDIC SURGEON. THEN PATIENT SAW A DIFFERENT SURGEON WHO REVISED THE PATIENT AND TOOK OUT HUMERAL COMPONENTS AND GLENOSPHERE AND REPLACED. THERE WAS A LOT OF METALLOSIS AND THE TRAY WAS COMPLETELY SHEERED OFF THE STEM. THE STEM HAD OVERSTUFFED THE JOINT, A 15 PRESERVE STEM WAS USED. THE SURGEON COULD NOT FIND THE LINER POLY FROM THE ORIGINAL SURGERY. SURGEON CHECKED AROUND THE JOINT FOR IT AS WELL AS HIS PA AND ASSIST. NO ONE COULD FEEL FOR THE 46 +2.5MM LINER IN THE JOINT. PATIENT WAS REVISED TO A FX STEM AND A NEW GLENOSPHERE AND TRAY AND LINER. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT

    Patient problems: Implant Pain; Loss of Range of Motion; Metal Related Pathology

    MAUDE report 20351970 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K162726. One submission often covers many models, so check the product and lot details in each record.

  • Z-0958-2019 · Dec 5, 2018 · Terminated
    Equinoxe Preserve Humeral Stem, 8mm Product Usage: The Equinoxe Preserve Stems are indicated for use in skeletally mature individuals with degenerative diseases of the glenohum